Lazar Fleysher1, Mounir El Mendili2, Guilaume Madelin3, Fernando Boada 3, and Matilde Inglese2
1Mount Sinai Medical Center, New York, NY, United States, 2Icahn School of Medicine at Mount Sinai, New York, NY, United States, 3NYU School of Medicine, New York, NY, United States
Synopsis
We report results of a comparison of three methods of tissue sodium concentration (TSC) quantification which use GRE, FLORET and TPI acquisition techniques at different field strengths. Results demonstrate a good inter-method agreement (better than 10%) in TCS quantification.
Introduction
We report results of a comparison
of three methods of tissue sodium concentration (TSC) quantification
which use GRE, FLORET and TPI acquisition techniques at different
field strengths.Methods
Acquisition. Nine healthy volunteers (four females) were studied. Images were
acquired with 7T Siemens Magntom and 3T Siemens Prisma MRI scanners
(Siemens, Erlangen, Germany) equipped with a custom-made dual-tuned
Na/H head coils. Acquisition protocol included 1H-MPRAGE
and sodium images using a TPI-based[1], FLORET-based[2] and
GRE-based[3] sequences. TPI-based acquisition was conducted at 3T,
while the other two were done at 7T. For the purposes of tissue
sodium concentration (TSC) quantification, 150mmol/L and 100mmol/L
reference tubes were used at 7T, while at 3T the vitreous body of the
eye was used as a signal reference. It was not possible to place
reference tubes in the FOV at 3T due to a small size of the head
coil. To evaluate the pair-wise differences between the methods,
Bland-Altman plots[4] were used. To avoid partial-volume effects with
CSF, signal from the white matter was used in the comparison.
Systematic errors between the methods. Vertical
axis in Bland-Altman plots
depicts
the measurement of the difference between the two methods with 95%
confidence interval marked
as dashed-dotted lines.
Consequently, a non-zero
average (solid line)
represents a combination of
systematic methodological
and instrumental differences
between the two compared methods. The
systematic error could be caused by un-accounted B0 and B1
inhomogeneities, B0 and B1-offsets, image SNR, signal calibration
technique and (in case of GRE acquisition) signal decay correction.
The scatter of values along the vertical axis represents a
combination of intra-subject methodological and instrumental
systematic errors.
Biological spread. The
absolute values on the horizontal axis represent the average of the
two compared TSC concentrations and the horizontal scatter represents
a combination of subject-to-subject biological variations (mostly
due to age differences) and
intra-subject methodological/instrumental systematic errors.Results
The obtained measurements show that FLORET vs TPI systematic
difference is about 3.2%, GRE vs FLORET one is 6.5% and GRE vs TPI
one is 9.7%. On average, GRE-based method produces the lowest TSC
values, while TPI method produces the highest ones. The GRE-FLORET
95% confidence interval is the shortest probably due the fact that
the two images were acquired in a single session at 7T which
increases SNR and minimizes differences due to B1 and B0
inhomogeneities. TPI images were acquired in a separate session on a
3T scanner therefore contribution of noise, B1 and B0 effects into
the quantification could be very different increasing the
random-appearing scatter in the values.Conclusions
In this work we demonstrate a good inter-method agreement (better
than 10%) in TCS quantification.Acknowledgements
This study was supported in
part by NS099527.References
1.Qian et al. Magnetic
Resonance Imaging 27:656
(2009)
2. Madelin et al. Scientific
Reports
4:4763
(2014)
3. Fleysher et al. NMR
Biomed 26:9
(2013)
4.
Bland et al. Statistical
Methods in Medical Research
8:135
(1999)