Qi Peng1, Can Wu2, Xiaojuan Li3, Michael L Lipton1, and Craig Branch1
1Gruss Magnetic Resonance Research Center, Department of Radiology, Albert Einstein College of Medicine and Montefiore Medical Center, Bronx, NY, United States, 2Philips Healthcare, Andover, MA, United States, 3Program of Advanced Musculoskeletal Imaging (PAMI), Cleveland Clinic, Cleveland, OH, United States
Synopsis
Quantitative T1rho (T1ρ) mapping MRI has recently
gained wider clinical/research application in human brain imaging. However, high
resolution 3D brain T1rho mapping on clinical scanners is still time consuming with
compromised quantitative accuracy due to image blurring or artifacts. We propose a 3D
FLAIR MAPSS T1rho imaging scheme, which combines CSF-suppression with 3D MAPSS sequence
for improved quantitative accuracy in T1rho mapping. Phantom, volunteer, and
patient studies validated its sensitivity and reliability for isotropic high resolution
of voxel size (1.3mm)3 quantitative mapping using continuous-wave
or adiabatic RF pulse T1rho preparation, acquired within clinically acceptable
scan duration at 3T.
INTRODUCTION
Quantitative
T1rho (T1ρ) mapping MRI has been shown to be sensitive to pH, macromolecule
density, and metabolic dynamics in the brain and have gained wide applications
in human brain conditions such as multiple sclerosis,1 bipolar,2
tumor,3
and stroke.4 Ideally,
quantitative T1rho mapping for the brain has isotropic high spatial resolution 3D
whole brain coverage and high measurement accuracy and reliability, all acquired
within reasonable scan duration. However, the scan duration for high resolution
3D T1rho mapping is particularly lengthened due to high SAR induced by long
spin-lock RF pulses needed by relatively long T1rho of brain GM/WM tissues, and
multiple spin-lock times needed for curve fitting. In addition, the accuracy
and reliability of the T1rho measurements are greatly compromised by the image
blurring/ghosting due to transient or relaxation decay effects by multiple shared
acquisitions per preparation in segmented 3D GRE or FSE sequences. Artifacts
and partial volume effects originated from the CSF (which has more than 10-fold
higher T1rho than those of the brain tissues) further contaminate the
quantitative maps. The purpose of this work is to present an isotropic
high-resolution 3D
FLAIR MAPSS T1rho sequence, re-optimized for brain based on a
synovial fluid suppressed 3D T1rho sequence recently developed to achieve fast
and accurate quantitative mapping for human knee cartilage.5 This
3D FLAIR MAPSS
technique was combined with either traditional continuous-wave T1rho
(cw-T1rho), or more recent adiabatic T1rho (adiab-T1rho) preparation modules and
was tested in phantom, volunteer, and patient studies to validate its
effectiveness for high resolution and accurate quantitative mapping of the
brain. METHODS
The
3D FLAIR MAPSS sequence scheme is shown in Fig.1. Bloch simulation was
performed to optimize sequence timing and shot duration of 5000ms was chosen to
achieve maximum SNR efficiency for brain WM. In each shot, a T2-preparation
module (with TE=140ms) followed by an inversion recovery RF pulse was employed to
selectively invert CSF and saturate brain tissues.6 After
optimized TI delay of 1600ms to null CSF, a MAPSS sequence scheme was used to achieve fast and accurate T1rho
imaging.7 All imaging was
performed on a 3T Philips Ingenia Elition MR scanner with a 32-channel head
coil. The cw-T1rho preparation module had 90°-TSL/2-180°-TSL/2-90°
RF pulse train, with spin lock frequency=500Hz and TSL=0, 50, 60ms. The
adiab-T1rho preparation module had adiabatic HS-8 pulse trains with TSL=4,
52, 68ms. 3D sagittal volumetric imaging was performed with the following scan
parameters: FOV=230/230/180mm, acquisition voxel size=(1.3mm)3, TR/TE=5.7/2.7ms,
echo train length of 180 with centric profile ordering, readout bandwidth =48kHz,
and compressed SENSE factor=9. As previously proposed,7
phase-cycling, and variable flip angle train as described in were employed to
minimize imaging blurring/ghosting along the phase encode directions to
maximize quantitative accuracy. The scan duration was about 6:45 min for both the
cw-T1rho and the adiab-T1rho scans. In the phantom study, the proposed cw-T1rho
3D FLAIR MAPSS was compared to an earlier validated 3D MAPSS cw-T1rho sequence
(with TSL=0, 20, 40, 60) using the same phantom of the report.8RESULTS
Phantom
regions of interest measurement results were highly consistent between the
proposed 3D FLAIR MAPSS sequence and the reference 3D MAPSS sequence (see Table
1). The mean difference of the T1rho measurements were all under 1.2% with average
mean difference -0.4%. In human studies, both cw- and adib-T1rho 3D FLAIR MAPSS
sequences achieved effective CSF suppression, with WM-to-CSF contrast-to-noise
ratio consistently larger than 10. No CSF-related artifacts could be identified.
Fig.2 shows representative 3D FLAIR MAPSS T1rho imaging results from a 66-yo
volunteer without prior known WM conditions. It is shown that both cw- and
adiab-T1rho sequences had high resolution and sensitivity to detect small
demyelinating lesions. Fig.3 shows representative T1rho imaging results
from an MS patient, compared with the corresponding 2D FLAIR image. All hyperintense
lesions shown on FLAIR had elevated T1rho values, validating the sensitivity of
the tested sequence. DISCUSSION
Commonly
used segmented GRE or FSE brain T1rho mapping sequences are flawed since quantitative
accuracy is compromised by ghosting/blurring originated from transient or
relaxation decay effects. It is more problematic when high spatial resolution
is needed. 3D MAPSS T1rho technique is a robust imaging technique specifically
proposed to mitigate this problem.7 The
CSF signal, if not suppressed, can be another major source of uncertainty in brain
T1rho mapping. High CSF signal can compromise quantitative accuracy through partial
volume effects in addition to severe blurring/ghosting. Furthermore, CSF-induced
artifacts can spread to non-neighboring brain voxels when k-space
under-sampling techniques (e.g., parallel imaging and compressed sensing) are
employed. These fast imaging techniques are indispensable to reduce total scan
duration in high resolution 3D T1rho mapping. It is shown in this work that the
proposed 3D FLAIR MAPSS scheme can be used as a general solution for 3D cw-T1rho
and adiab-T1rho mapping techniques for high resolution quantitative brain
imaging. With this, high-resolution 3D T1rho mapping without CSF signal contamination
can be obtained on clinical 3T scanners within acceptable scan duration. CONCLUSION
High resolution, CSF-suppressed, whole brain T1rho mapping with
isotropic voxel size can be achieved within clinically acceptable scan duration
using 3D FLAIR MAPSS sequences. This will help improve its quantitative accuracy,
expand its research applications, and facilitate its wider clinical adoption.Acknowledgements
No acknowledgement found.References
1. Borthakur
A, Sochor M, Davatzikos C, et al. T1rho MRI of Alzheimer's disease. Neuroimage.
2008;41(4):1199-1205.
2. Johnson
CP, Follmer RL, Oguz I, et al. Brain abnormalities in bipolar disorder detected
by quantitative T1rho mapping. Mol Psychiatry. 2015;20(2):201-206.
3. Villanueva-Meyer
JE, Barajas RF, Jr., Mabray MC, et al. Differentiation of brain tumor-related
edema based on 3D T1rho imaging. Eur J Radiol. 2017;91:88-92.
4. Tan
Y, Xu J, Chen R, et al. Use of T1 relaxation time in rotating frame (T1 rho)
and apparent diffusion coefficient to estimate cerebral stroke evolution. J
Magn Reson Imaging. 2018;48(5):1247-1254.
5. Peng
Q, Wu C, Li X, et al. Improved 3D T1rho and T2 Mapping with Synovial Fluid Suppression
for the Knee Cartilage on 3T. Proc Intl Soc Mag Reson Med 24. 2018:5056.
6. Wong
EC, Liu TT, Luh WM, et al. T(1) and T(2) selective method for improved SNR in
CSF-attenuated imaging: T(2)-FLAIR. Magn Reson Med. 2001;45(3):529-532.
7. Li
X, Han ET, Busse RF, et al. In vivo T(1rho) mapping in cartilage using 3D
magnetization-prepared angle-modulated partitioned k-space spoiled gradient
echo snapshots (3D MAPSS). Magn Reson Med. 2008;59(2):298-307.
8. Kim
J, Mamoto K, Lartey R, et al. Multi-vendor multi-site T1ρ and T2 quantification
of knee cartilage. Proc Intl Soc Mag Reson Med 25. 2019:0127.