Fatiha Andoh1, Marion Tardieu2, Claire Barakat Pellot3, and Xavier Maître1
1Univ. Paris Saclay, IR4M Laboratory, Orsay, France, 2Institut du Cancer de Montpellier, ICM, Montpellier, France, 3INSERM, IMIV Laboratory, Orsay, France
Synopsis
Mechanical parameters may be underestimated by MR-Elastography if data
quality is not properly monitored to discarded inaccurate and unprecised biased
data. Optimal condition defined by proper spatial data sampling upon λ/a≈[6;9] can
alleviate the quality requirements by lowering Qthreshold, which is defined on each dataset at the drop
of correlation between the quality factor and the targeted mechanical parameters.
Introduction
Brain Magnetic Resonance
Elastography is usually performed in humans with 3 mm isotropic voxels at
low mechanical excitation frequency, between 40 Hz and 60 Hz, to ease
wave propagation deep inside the organ. Reported overall mean shear elasticity
modulus expectedly increased from 2 kPa at 40 Hz to 2.5 kPa at
60 Hz [1] [2]. Yet, MRE
quantitation is highly conditioned by data quality and sampling as 6 to 9 voxels per wavelength were found optimal in isotropic and
homogeneous media to produce valid mechanical parameters [3]. Here, comparative low and high frequency whole brain MRE was investigated voxel-by-voxel to analyse brain MRE optimal conditions
by evaluating the required data quality factor Q and the optimal data sampling
λ/a.Methods
Shear waves were generated in the
brain at 45 Hz and 113 Hz by pressure waves guided into the buccal
cavity [4].
Displacement fields were acquired for the same volunteer in a 1.5 T
Achieva system (Philips, Best, Netherlands) with parameters reported in Table 1.
Maps of shear dynamic and loss moduli, Gd and Gl,
were extracted from the three motion-encoding directions by inversion of the
Helmoltz equation of the rotational displacement field, q. T1-weighted image was used as anatomical reference
and was segmented with SPM12 (UCL,
United-Kingdom) to infer masks of grey and white matters after removal of
CSF, bone and background tissue. Computed maps were spatially-normalized. Data
quality maps, Q=q/Δq, were calculated according to [4]. Correlation
coefficients, R45 Hz and R113 Hz, between
Q and Gd, were calculated to quantify the alteration of the
mechanical outcomes with Q, first over
the whole organ then for Q-bins of n voxels (n=1000). The correlation
coefficients were plotted versus Q and exponentially fitted for 1000 to
determine a quality threshold, Qthreshold,
beyond which the correlation approaches zero. Finally, the
shear
dynamic and loss moduli, Gd and Gl
were evaluated only for voxels which quality was above Qthreshold.Results
MRE SNR are 7 at 45 Hz and 20 at 113 Hz. Q, Gd, and Gl maps are represented for
one of the slices along the three directions on Fig. 1. Mean Q is higher at 113 Hz than at 45 Hz
(Fig.1 and Table 2). At
both frequencies, maximal values were found at the 2 mm thick tentorial
surface of the cerebellum (where Q
values also reach maximal values) [5].
Elsewhere, Q is rather homogenous at
45 Hz whereas at 113, despite much higher values at the periphery of the
cerebral hemispheres and cerebellum, Q
falls down deep inside the brain. Mean
V, Gd and Gl increase with the
frequency excitation according to the values reported in Table 2.Discussions
The mean
coefficient correlation between Q and Gd shows that Gd is highly correlated to Q at 45 Hz (R45Hz=0.70),
whereas, at 113 Hz, it remains
rather immune to Q inhomogeneity (R113Hz=0.26).
The correlation curve between Gd
and Q shows that it is possible to
define a Qthreshold of 28
for data at 45 Hz which discriminates data with a mean correlation of 0.21 below the Qthreshold and 0.03 above. A 113 Hz, Qthreshold=15 and the mean correlation is 0.11 below Qthreshold
and 0.003 above (Table 2). Finally, at 45 Hz the mean value of Gd for voxels above Qthreshold is (0.75±0.30) kPa and this value is underestimated for voxels below Qthreshold
with (0.39±0.09) kPa. Gd
and R follow the same trend at 113 Hz
but the discrepancy below and above Qthreshold is less pronounced (see Table 2).Conclusions
Mechanical parameters may be underestimated by MR-Elastography if data
quality is not properly monitored to discarded inaccurate and unprecised biased
data. Optimal condition defined by proper spatial data sampling upon λ/a [6;9] can
alleviate the quality requirements by lowering Qthreshold, which is defined on each dataset at the drop
of correlation between the quality factor and the targeted mechanical parameters.Acknowledgements
MRE experiments
were performed on the 1.5 T MRI platform of CEA/SHFJ, affiliated to the France
Life Imaging network (grant ANR-11-INBS-0006).References
[1] I. Sack, K.-J. Streitberger, D.
Krefting, F. Paul, et J. Braun, « The Influence of Physiological Aging and
Atrophy on Brain Viscoelastic Properties in Humans », PLoS ONE,
vol. 6, no 9, p. e23451, sept. 2011.
[2] K.
M. Pepin et al., « Magnetic resonance elastography analysis of
glioma stiffness and IDH1 mutation status », AJNR Am J Neuroradiol,
vol. 39, no 1, p. 31‑36, janv. 2018.
[3] J.
L. Yue et al., « Acquisition and reconstruction conditions in
silico for accurate and precise magnetic resonance elastography », Phys.
Med. Biol., oct. 2017.
[4] X. Maitre, E. Lamain, R. Sinkus, B. Louis, L. Darrasse (2011) Whole
brain MRE with guided pressure waves. in 19th Annual Meeting of the
International Society for Magnetic Resonance in Medicine (ISMRM) (Montréal
(Canada)).
[5] Castel Jean-Pierre, « L’enveloppe
ostéo-durale », Campus Numérique de Neurochirurgie, févr-2006.
[6] J.
Diedrichsen, « A spatially unbiased atlas template of the human
cerebellum », NeuroImage, vol. 33, no 1, p. 127‑138,
oct. 2006.