Kosuke Morita1, Masami Yoneyama2, and Takeshi Nakaura3
1Kumamoto University Hospital, Kumamoto, Japan, 2Philips Japan, Tokyo, Japan, 3Graduate School of Medical Sciences, Kumamoto University, Kumamoto, Japan
Synopsis
Multi-slice single-shot 2D
echo-planar imaging (EPI) readout can be used effectively for pseudo-continuous
arterial spin labeling (pCASL) perfusion imaging. In this study we tried to utilize the Compressed SENSE
reconstruction for pCASL-EPI without further optimization of EPI
sampling scheme. pCASL-EPICS
clearly reduces noise-like artifacts and significantly improves the robustness
of perfusion images in a short scan time less than 3 minutes compared with
conventional pCASL SENSE, without any penalty for scan parameters. This
technique may be helpful to further assess the many brain diseases.
PURPOSE
Multi-slice single-shot 2D
echo-planar imaging (EPI) readout can be used effectively for pseudo-continuous
arterial spin labeling (pCASL) perfusion imaging and it should be considered a
viable alternative to segmented 3D sequences1, because they are available on
all systems and are insensitive to image artifacts from motion. However, 2D
imaging basically requires many signal averages to gain signal-to-noise ratio, which leads to longer scan time. In addition, typical clinical brain EPI-based
ASL imaging has limited spatial resolution due to its high sensitivity to B0
inhomogeneities as well as diffusion-weighted imaging (DWI). EPI with sensitivity
encoding (SENSE) helps to reduce the voxel size without increasing of image
distortion, but it often suffers from increased noise-like artifacts on the
center of the images due to the high geometry factor2,3. Recently, a
combination of parallel imaging and compressed sensing technique (Compressed
SENSE, C-SENSE) has been developed to accelerate the acquisition time without
increasing the image artifacts4,5. Although C-SENSE is basically
applied for non-EPI scans, it has been demonstrated that the C-SENSE
reconstruction could clearly reduce noise-like artifacts and significantly
improve the image quality of EPI based DWI without further optimization of EPI
sampling scheme6,7.
We hypothesized that EPI-based
ASL can also be applied the C-SENSE reconstruction framework as well as DW-EPI.
The purpose of this study was to demonstrate the feasibility of a combination
of EPI-based ASL and C-SENSE (pCASL-EPICS) for improving multi-slice perfusion
imaging.MATERIALS AND METHODS
Experimental
data was collected from 6 healthy volunteers. Written informed consent was
obtained from each volunteer and the protocol was approved by the ethics
committee. All studies were performed with clinical 3.0T MR scanner (Philips,
Ingenia 3.0T CX) and 32-channel dS-head coil.
EPICS is based single-shot EPI.
We did not modify its sampling pattern of single-shot EPI and we applied it
into the C-SENSE framework. Scan parameters of pCASL-EPICS were as follows: Transverse plane acquisition, TR/TE = 4550/7.9ms, slices
thickness = 6mm, voxel size = 3.75 × 3.93 × 6.0mm (similar setting with
commercially available 3D pCASL Gradient spin-echo (GraSE) protocol), number of
slices = 14, 18 dynamic scans, EPI factor = 21, CS factor or SENSE factor = 3.0,
pCASL label duration =1800ms, post label delay =2000ms, two-inversion pulses
background suppression and acquisition time = 2:52.
pCASL-EPICS were compared to conventional 2D SENSE-EPI
images and 3D GraSE-pCASL images for image quality, especially for the
reduction of image noise. ROIs were placed on left and right gray matter (GM). We
calculated the Coefficient of Variation (CV) of signal intensities (SI) of GM
on perfusion image in respective volunteers as follows: CV = SDSI /
AverageSI, to compare the consistency of perfusion related signal intensities
among pCASL-EPICS, SENSE 2D EPI and 3D GraSE-pCASL. RESULTS
Figure 1 shows the results of CV
values on the perfusion images b among pCASL-EPICS, SENSE 2D EPI and 3D
GraSE-pCASL. CV values of pCASL-EPICS (0.23±0.07) was significantly lower than
that of SENSE 2D EPI (0.37±0.08) (p < 0.01). The cause of high CV values of
SENSE was most likely due to the presence of severe noise across the images. On
the other hand, there were no significant differences between CV values of pCASL-EPICS
and 3D GraSE-pCASL(0.24±0.10) (p=0.47).
Figure
2 shows representative perfusion images using pCASL-EPICS, SENSE 2D EPI and 3D
GraSE-pCASL in two volunteers. pCASL-EPICS clearly reduced the noise which
exists in the center of the SENSE images and almost similar SNR compared with
3D GraSE-pCASL. It indicated that pCASL-EPICS can provide more accurate perfusion
information with high reproducibility and robustness. Figure 3 also shows
representative perfusion images using pCASL-EPICS, SENSE 2D EPI and 3D
GraSE-pCASL with increased voxel size (2.5mm2) from same setting of
Figure 2 images. Although further optimization to obtain more sufficient SNR is
needed, pCASL-EPICS still maintained higher SNR compared with other two methods.
Figure 4 shows the representative multi-slice perfusion images using optimized
pCASL-EPICS.CONCLUSION
pCASL-EPICS
clearly reduces noise-like artifacts and significantly improves the robustness
of perfusion images in a short scan time less than 3 minutes compared with
conventional pCASL SENSE, without any penalty for scan parameters. This
technique may be helpful to further assess the many brain diseases.Acknowledgements
No acknowledgement found.References
1.
Alsop DC, et al. Recommended implementation of arterial spin-labeled perfusion
MRI for clinical applications: A consensus of the ISMRM perfusion study group
and the European consortium for ASL in dementia. Magn Reson Med.
2015;73(1):102-16.
2. Patricia N,
et al. Parallel Imaging Artifacts in Body Magnetic Resonance Imaging. Can Assoc
Radiol J. 2009;60: 91–98.
3. Yanasak NE,
et al. MR imaging artifacts and parallel imaging techniques with calibration
scanning: a new twist on old problems. Radiographics. 2014;34:532-48.
4.
Geerts-Ossevoort L, et al. Compressed SENSE Speed done right. Every time. The
Netherlands: Philips Healthcare; 2018 Jan. Report No: 4522 991 31821.
https://www.philips.de/content/dam/b2bhc/de/resourcecatalog/landingpages/ingeniaelition/White_Paper_Compressed_SENSE-opt.pdf
5. Mönch S, et
al. Magnetic Resonance Imaging of the Brain Using Compressed Sensing - Quality
Assessment in Daily Clinical Routine. Clin Neuroradiol. 2019 May 16. doi:
10.1007/s00062-019-00789-x. [Epub ahead of print]
6. Morita, et
al. Pseudo-3D Diffusion-Weighted
Imaging of the Brain using Echo Planar Imaging with Compressed SENSE (EPICS). Proc.
ISMRM. 2019:3355.
7. Yoneyama, et
al. Noise Reduction in Prostate Single-Shot DW-EPI utilizing Compressed SENSE Framework. Proc. ISMRM. 2019:1634