Lukas M. Gottwald1, Joao Tourais2,3, Eva S. Peper1, Jouke Smink2, Bram F. Coolen4, Gustav J. Strijkers4, Pim van Ooij1, and Aart J. Nederveen1
1Radiology and Nuclear Medicine, Amsterdam UMC, Amsterdam, Netherlands, 2MR R&D – Clinical Science, Philips Healthcare, Best, Netherlands, 3Department of Biomedical Engineering, University of Technology, Eindhoven, Netherlands, 4Biomedical Engineering and Physics, Amsterdam UMC, Amsterdam, Netherlands
Synopsis
This study aimed to compare the performance of the novel camera-based respiratory navigation sensor (VitalEye) in retrospective respiratory binned Cartesian 4D flow MRI to conventional liver navigator and self-gating. Analyzed were the cross-correlation of the respiratory signals, peak flow rate error compared to 2D flow and the image quality in terms of edge sharpness of the liver/diaphragm border and signal-to-noise ratio. The novel camera-based respiratory navigation sensor VitalEye performed as good as conventional liver navigator and self-gating. Respiratory signal, flow rate error, and image quality showed no significant difference, but VitalEye has the advantage of a 10-times higher sampling frequency.
Introduction
Respiratory
correction is important in cardiac MRI to avoid blurring and motion artifacts.
Several techniques have been implemented, mainly by restricting the acquisition
to the same short time window during the end-expiratory phase. Recently, a
novel contact-less camera-based respiratory navigation sensor has been
introduced1 and shown improved image quality in abdominal MRI compared to
respiratory belt-based triggering2. This
may also apply to 4D flow MRI in which the recommended method is a
liver/diaphragm MR navigator3 or self-gating4.
This study aimed
to compare the performance of the novel camera-based respiratory navigation sensor
(VitalEye) in retrospective respiratory binned Cartesian 4D flow MRI to conventional liver
navigator (RNAV) and self-gating (SG).Methods
Pseudo-spiral
compressed sensing (CS) accelerated 4D flow MRI
5,6 of the heart or the aorta were scanned in N
total=14
subjects on a 3T MR system (Philips Ingenia ElitionX; Philips Medical Systems,
Best, The Netherlands) which was equipped with a built-in-the-bore camera of
type IDS uEye (IDS, Obersulm, Germany). The subject cohort N
total
was divided into four sub-groups N
1A=14, N
1B=7, N
2=6,
and N
3=6 depending on the performed scan-exam and analysis used. The
scanner software was modified to both continuously sample the k-space in a
pseudo-spiral Cartesian fashion, which also enabled SG due to repetitive
k-space center sampling (k
0/0), and to acquire the RNAV with a
predefined frequency without ECG-synchronization, which led to different RNAV
sample points over the cardiac cycle. 4D flow parameters: TE/TR/flip angle=
2.1ms/3.9ms/8°, spatial resolution= 2.5x2.5x2.5mm
3, cardiac frames=
30, VENC= 150cm/s, scan time= 5:18min (aorta) or 6:58min (heart). 4D flow
scans of the heart were acquired two times, one with RNAV sampling (4D flow
[1]RNAV/VitalEye)
and one without (4D flow
[2]VitalEye), but always with VitalEye. As a reference, 2D flow scans were acquired at three ROI in the ascending aorta (AAo) and two in the descending aorta (DAo
1, DAo
2).
4D flow data were processed offline using ReconFrame (Gyrotools, Zurich, Switzerland) in MATLAB (The
MathWorks Inc., Natick, MA, USA) together with the Berkeley Advanced
Reconstruction Toolbox
7 (BART) for CS reconstruction with a sparsifying total variation
transform in time. The SG signal was extracted in four steps
8: Fourier transforms of k
0/0 along the readout direction,
singular value decomposition per coil, bandpass filtering, and coil clustering.
Each respiratory signal was extracted from the raw data. VitalEye was sampled
with a frequency of $$$f$$$
VitalEye=20.0 Hz, RNAV with $$$f$$$
RNAV=2.0 Hz, and SG with $$$f$$$
SG=2.3±0.2 Hz. After phase-binning with
expiratory-phase defined by a 60% acceptance rate, the final
undersampling/acceleration was R = 9.26±0.02 (aorta) to 10.45±0.01 (heart).
Three different
analyses were performed in MATLAB to compare the respiratory navigation
performance:
- Cross-correlation between VitalEye vs RNAV (N1A =14) and VitalEye vs RNAV vs SG (N1B =7). Cross-correlation $$$C$$$ of two respiratory signals or their phases $$$g$$$ and $$$h$$$ with delay $$$\tau$$$ is defined: $$C = (g\star h)(\tau)=\int_{-\infty}^{\infty} \overline{g(t)}h(t+\tau) dt$$
- Peak flow rate difference between 4D flowRNAV/VitalEye/SG vs 2D flow MRI (N2=6) in three ROI in the form
of Bland-Altman and correlation plots. All flow analyses were done with Segment9.
- Image
quality based on the signal-to-noise ratio $$$SNR=\frac{\overline{S}}{\sigma_{N}}$$$ as well as the
diaphragm sharpness $$$S$$$ defined by a sigmoid fit width over a line profile. Analyzed
were two heart scans 4D flow[1]RNAV/VitalEye binned with
either RNAV or VitalEye signal, and a 4D flow[2]VitalEye
binned with VitalEye signal to investigate if the RNAV sampling introduces
image distortions (N2=6).
Results
The results of
the cross-correlation (Figure 1),
the peak flow rate difference (Figure 2)
as well as the image quality measurements (Figure 3) are listed in Tables 1.1-1.3 (Figure 4).
It can be seen that RNAV and VitalEye show a high correlation, especially for the
respiratory phase or the eventual binning, respectively, are almost identical to $$$C$$$PhaseRNAV/VitalEye=0.97±0.01. A moderate correlation can
be seen when comparing SG to VitalEye or RNAV of $$$C$$$PhaseRNAV/SG=0.85±0.03
and $$$C$$$PhaseVitalEye/SG=0.85±0.03. The peak flow rate
comparison to 2D flow MRI showed the least error for 4D flowVitalEye
of 6.1% followed by 4D flowRNAV of 6.7% and 4D flowSG of
6.8%. The image quality measurements showed equal $$$SNR$$$ and $$$S$$$ values.Discussion
The comparison between
the three respiratory signals revealed that there is no significant difference
in terms of respiratory signal monitoring, flow rate error, and image quality. However,
each technique has its (dis-)advantages. The major advantage for the RNAV is
that the signal is a displacement measurement in millimeter in contrast to
VitalEye and SG, which signal units are arbitrary. Therefore, the RNAV binning
might improve if absolute binning is used instead of phase binning as in this
study. The major advantage of VitalEye is the high sampling frequency of 20.0
Hz, which allows sampling of any occurring respiration rate with Nyquist
frequency – even those for infants of up to 1.5 Hz. Additionally, the external
setup does not change or interfere with the MRI sequence. In contrast, scan
time is sacrificed for RNAV or the sampling pattern needs to be adopted for SG.Conclusion
The novel
camera-based respiratory navigation sensor (VitalEye) performed as good as conventional
liver navigator and self-gating in retrospective respiratory binned Cartesian 4D flow MRI.
Respiratory signal, flow rate error, and image quality showed no significant
difference, but VitalEye has the advantage of a 10-times higher sampling
frequency.Acknowledgements
This work has been supported by the Netherlands Organization for Scientific Research – NWO (HTSM2014).
We thank Martin Bührer for his quick replies and technical assistance whenever needed.
References
- Kruger MG,
Springer RPW, Kersten GM, Bril RJ. Towards Contact-less Vital Sign Monitoring
using a COTS Resource-Constrained Multi-Core System-an Experience report. In: 2019
IEEE 22nd International Symposium on Real-Time Distributed Computing (ISORC).
; 2019:77-78.
- Harder
F, Lohöfer FK, Kaissis GA, et al. Camera-based respiratory triggering
improves the image quality of 3D magnetic resonance cholangiopancreatography. Eur
J Radiol. 2019. doi:10.1016/j.ejrad.2019.108675
- Dyverfeldt
P, Bissell M, Barker AJ, et al. 4D flow cardiovascular magnetic resonance
consensus statement. J Cardiovasc Magn Reson. 2015:1-19.
doi:10.1186/s12968-015-0174-5
- Uribe S,
Beerbaum P, Sørensen TS, Rasmusson A, Razavi R, Schaeffter T. Four-dimensional
(4D) flow of the whole heart and great vessels using real-time respiratory
self-gating. Magn Reson Med. 2009. doi:10.1002/mrm.22090
- Gottwald LM,
Peper ES, Zang Q, et al. Compressed Sensing accelerated 4D flow MRI using a
pseudo spiral Cartesian sampling technique with random undersampling in time. Proc
Intl Soc Mag Reson Med 25. 2017:1263.
- Peper* ES,
Gottwald* LM, Zhang Q, et al. Highly accelerated carotid 4D flow MRI using
Pseudo-Spiral Cartesian acquisition and a Total Variation constrained
Compressed Sensing reconstruction. In: Journal of Cardiovascular Magnetic
Resonance. ; 2019:IN PRESS.
- Uecker M,
Ong F, Tamir JI, et al. Berkeley Advanced Reconstruction Toolbox. Proc Intl
Soc Mag Reson Med. 2015;23(1):2486.
- Walheim J,
Dillinger H, Kozerke S. Multipoint 5D flow cardiovascular magnetic resonance -
accelerated cardiac- and respiratory-motion resolved mapping of mean and
turbulent velocities. J Cardiovasc Magn Reson. 2019;21(1):42.
doi:10.1186/s12968-019-0549-0
- Heiberg E,
Sjögren J, Ugander M, Carlsson M, Engblom H, Arheden H. Design and validation
of Segment - freely available software for cardiovascular image analysis. BMC
Med Imaging. 2010. doi:10.1186/1471-2342-10-1