Hans Hoogduin1, Mark Gosselink1, Giel Mens1, Wim Prins2, Tijl van der Velden1, and Dennis Klomp1
1UMCU, Utrecht, Netherlands, 2Philips, Best, Netherlands
Synopsis
Directional couplers are used to measure reflected waves from an eight channel PTx coil to detect head motion at 7T. The method doesn't require any changes to pulse sequences and has no time penalty. A general linear model is used to predict head motion from the signals measured at the couplers.
Introduction
Subject motion remains one of the key sources of image
degradation in MRI. This is especially true for the detailed images acquired at
high field (7T and above) which require long scan times (10 min). Ideally motion
should be detected and corrected for in real time. In this work, we use the
method that is based on measuring reflected RF waves from an 8 channel parallel
transmit (PTx) coil using directional couplers (DICOs) [1]. The amount
of reflection depends on the loading of the coil which in turn depends on the
position and orientation of the head in the coil. Scattered RF waves are
measured for each RF pulse in the sequence without any time penalty. Unlike recent previous work [2], our implementation doesn’t require any changes to pulse
sequences and has no time penalty. Only the scattered waves from the coil are
used to detect and estimate changes in the position of the subjects head.Methods
All measurements were performed on an 8 channel PTx 7T
scanner (Philips) using the DICOs integrated into the eight 2kW amplifiers (CPC)
for motion detection together with an 8 channel transmit coil (NOVA medical).
Three single shot EPI sequences with 48 slices per volume (TR 2s) were acquired
on a single volunteer. During the first scan (10 volumes) the subject was
instructed to keep his head still. During the other two scans, consisting of 30
volumes with a manual start of each volume, the subject moved his head in
between the acquisition of the volumes. Both head rotations and translations
were performed. DICO data consisted of 8
times 48 forward and reflected RF pulses (asymmetric sinc gauss) per volume
with 43 samples per pulse. Only the reflected RF waves were used for motion
detection.
EPI data was realigned (rigid body) to first volume of
the first scan using the McFLIRT package from FSL. DICO signals were processed
in Matlab. The maximum of the volume averaged reflected RF
pulses (magnitude only) was used as the ‘motion detector’(MD) sample of a volume. The mean of the 10 MD samples
obtained from the scan without head motion was subtracted from each of the 70 MD
samples. The 40 mean corrected MD samples of the first and second scan (10
without motion and 30 with motion) were fitted to the corresponding rotations
and translations obtained from the FSL image
registration using a General Linear Model (GLM). The fitted GLM matrix was used
to calculate the head motion of the subject based on the MD samples from the DICOs
for all 3 scans. The correlation coefficient between the GLM and the FSL based realignment parameters was used to
evaluate the quality of the GLM fit (scan 1 no motion and scan 2 motion) and
the predicted motion (scan 3 motion).Results
Figure 1 shows the raw reflected signals from a DICO
of one of the RF channels. The sensitivity to motion is clearly visible in the raw
data. The additional small variation on top of the data is related to the slice
dependent change in RF frequency in combination with the high Q-factor of the
coil. Figure 2 shows the realignment parameters from FSL and the mean corrected
MD signals for scan 1 and 2 used to build the GLM. Figure 3 shows the
realignment parameters together with the fitted signal for scan 1 and 2
(correlation coefficient 0.98), and the motion predicted by the model for scan
3 (correlation coefficient 0.87). Discussion
Reflected signals, as measured by DICOs, scattered
from an 8 channel head coil can be used to detect head motion. In previous work [2] the full complex S-matrix was used to determine head motion. Here we only use the
magnitude of reflected signals. In addition, no changes to the pulse sequence are
needed and no additional scan time is required. Both gross rotations and
translations can be predicted with reasonable accuracy. By using both the real
and imaginary part of the signal the fits can probably be improved. In future
work, the method will be used to update the scan orientation based on GLM
predictions in real time. Conclusion
The proposed method for detection of subject head motion
opens the way to improve high field high resolution image quality without a
penalty in scan time or the need for sequence modifications.Acknowledgements
The first author wishes to thank Nico van den Berg for fruitful discussions on motion detection in MRI.References
[1] Buikman D, Helzel T, Roeschmann P. The RF coil as a sensitive motion detector for magnetic resonance imaging. Magn Reson Imaging. 1988;6:281–289.
[2] Papp D. Jaeschke S, Rieger S, Clare S, Hess A. Simultaneuous detection of cardiac, respiratory, and rigid body head motion using the scattering of a parallel transmit RF coil at 7T. Abstract 1168, ISMRM 2018.