J. Rock Hadley1, Henrik Odeen2, Robb Merrill2, Sam Adams2, Viola Rieke2, Allison Payne2, and Dennis Parker2
1Radiology and Imaging Sciences, University of Utah, Salt Lake City, UT, United States, 2University of Utah, Salt Lake City, UT, United States
Synopsis
This work uses an RF screen, placed over the top of a human
skull phantom, to reduce image banding artifacts that are common in
transcranial transducer MRI. The goals
of the study are to improve imaging homogeneity over the region of the brain by
changing RF field patterns that cause the artifacts, and to find a solution
that doesn’t attenuate or distort the ultrasound properties of the
transducer. Hydrophone and focused
ultrasound heating studies are performed to measure ultrasound screen transparency
and MRI studies are performed to evaluate the effects the screen has on homogeneity
and artifact reduction.
Introduction
This work investigates a solution to
the Radio Frequency (RF) signal banding artifacts that can occur with Magnetic
Resonance Imaging (MRI) of the head inside an Insightec Transcranial Transducer
(ITT), see Figure 1.
We
hypothesized that these artifacts are caused by the dielectric properties of
the large water bath and the waveguide nature of the boundary conditions of the
transducer ground plane. Furthermore, we
hypothesized that the banding artifacts could be shifted by the insertion of a
conductive screen in the space between the transducer and the patient’s head.
The two major
goals in this work were to: 1) modify the electromagnetic field behavior of the
B1-transmit and receive field inside the transducer to improve field
homogeneity and eliminate the banding artifacts in the MR images, and 2) Do so
in such a way that the ultrasound transmission into the human skull would not
be attenuated or otherwise negatively affected. Methods
A human skull,
with a Data Quality Assurance phantom mounted inside, was imaged in the ITT
(650 kHz) with a 2D Gradient Echo pulse sequence using a Skyra MRI with body
coil (Siemens Medical Solutions). Imaging was done with and without a 24-cm
diameter conductive screen positioned over the top of the head as shown in
Figure 2. Two different imaging studies were performed using the RF
screen. The first study was using the
screen in position #1, where it had slipped off the skull, but the images look
interesting so the study was continued. the
second imaging study was done with the screen in position #2, the originally
desired configuration (see Figure 3). The
copper screen was made from un-insulated bronze wire that was approximately
0.25-mm in diameter, and the wires formed small mesh squares that were 2mm in
diameter. The screen was held in position with string and small rubber spacers that
were used to isolate the screen from the skull.
Hydrophone measurements were performed
with the screen to assess ultrasound transparency. The screen was positioned in the hydrophone
tank perpendicular to the propagating beam at 2.5, 4.5 and 6.5-cm from the
focus and at 45° to the
beam propagation. The transducer was a
256 element semi-rectangular transducer with a 14.4x9.8-cm aperture ultrasound
transducer (Image Guided Therapy/Imasonic, France) with a 10-cm focal
length and operating at 940 kHz.
Heating studies in the ITT transducer
were performed with and without the screen in place. For each situation, five
different heating runs were performed and the standard deviation of the
temperatures and FWHM of the temperature distributions were measured.Results
The screen caused pressure attenuations
on the order of 6% for the different flat and angled configurations compared to
the water-only hydrophone measurement.
Imaging results (see Figure 3) show
improved signal homogeneity in the brain region with the screen in position.
The Signal-to-Noise Ratio (SNR) was improved on the order of 50% in regions of
no artifact and by factors of 5 and more in regions of artifact. Temperature accuracy was improved and more homogeneous
throughout the brain when using the screen.
The mean and
standard deviation of the temperatures and temperature profile widths from the
heating studies are shown in Figure 4. Discussion
Figure 4 demonstrates that there was negligible change in
the peak temperature with the screen although the temperature distribution
(FWHM) was reduced with the screen in place.
The increased SNR with the screen in place resulted in improved
precision of the peak and FWHM of the temperature distribution measurements.
Hydrophone and force balance studies indicate that the
heating through the screen should be reduced compared heating without the
screen, however, the hydrophone studies don’t account for the lower frequency
and different geometry of the transducer.
In the skull heating studies, not all of the transducer elements are
required to shoot through the screen.Conclusion
In conclusion, for the various
imaging parameters of this study, the copper screen: 1) provided significant
improvement in field homogeneity throughout the
region of the brain by eliminating the banding artifacts; 2) increased
SNR by a factor of 1.5 to 6 over the region of the brain, and 3) was
essentially transparent to the ultrasound beam. We hypothesize that these
improvements occurred because the copper screen changed
the field boundary conditions and modified the waveguide nature of the
transducer ground plane. Future work will more fully characterize and
optimize the RF screen functionality.Acknowledgements
The Mark H. Huntsman Endowed Chair and NIH grant 1R01
EB028316
References
No reference found.