Jerome L. Ackerman1,2, Shahin Pourrahimi3, Marcus Donaldson1,2, Nadder Pourrahimi3, Julien Rivoire4, Julien Muller4, Hizami Murad4, Ouri Cohen5, and Isabela Choi1
1Martinos Center, Dept of Radiology, Massachusetts General Hospital, Charlestown, MA, United States, 2Department of Radiology, Harvard Medical School, Boston, MA, United States, 3Superconducting Systems, Inc., Billerica, MA, United States, 4RS2D, Mundolsheim, France, 5Memorial Sloan Kettering Cancer Center, New York, NY, United States
Synopsis
We developed an MRI scanner
designed for limb scanning based on a novel conduction-cooled three-bore cryogen-free
1.5T magnet. The two outer bores accommodate the unscanned leg for enhanced
patient comfort. We describe the issues relevant to this unique magnet and
scanner, including our solution to the field instability problem common in
conduction-cooled MRI magnets. By recording the field variation as a function
of the phase of the cold head cycle, a compensating waveform may be played out via
the numerically-controlled oscillator of the scanner to reduce the periodic
field excursion from 25 Hz to about 1-2 Hz.
Introduction
We developed a compact, easy to
site prototype limb scanner for orthopedic and metabolic bone disease
applications. Using a prototype (Alpha) tilted, conduction-cooled, cryogen-free,
highly shielded 1.5T magnet with three bores allows the scanner to be installed
in an extremely compact space, eliminates the need for a quench duct or helium
refills, and offers a more comfortable patient experience. This presentation reports
our experience in recovering after a quench, the reduced cooling efficiency
operating the magnet at a tilt, and suppressing the field shift caused by the
periodic cycling of the cold head. Because conduction-cooled magnets may have a
small cyclic field variation arising from the periodic cold head regenerator
displacement or temperature cycling, or from mechanical vibrations, these
magnets may suffer “motion” artifacts caused by the field fluctuation rather
than actual motion with respect to the gradient. Results of testing of the
Alpha magnet have been used to design a Beta magnet with improved performance
and patient comfort.Methods
The magnet was designed and
produced by Superconducting Systems under NIH grant R44AR065903 and sited at
MGH to be incorporated into a scanner.1 The console is based on an RS2D
Chameleon 3 spectrometer (comprising a digital RF transceiver, pulse sequencer,
gradient drivers, shim supplies and RFPA), a Ningbo Yici (China) EC04 gradient
coil, three PCI (Performance Controls, Inc., Montgomeryville, PA, USA) GA-300
amplifiers powered with a Sorensen-Ametek (Berwyn, PA, USA) SGA330X15C-1EAA DC
power supply. The magnet was cooled with a Sumitomo (SHI Cryogenics, Allentown,
PA, USA) SRP-082B2 pulse tube cold-head with F70LP helium compressor. RF coils
were either designed and built in-house or were modifications of commercial devices
retuned to 1.5T. The magnet diode temperature sensor was read with a Lake Shore
Cryotronics (Westerville, OH, USA) 211A monitor. The scanner host computer
records magnet temperature, compressor diagnostics, gradient and water temperatures
via serial ports.Results
Tilting the magnet modestly
considerably enhances patient comfort for arm and leg scanning, but can subject
internal magnet structures to undue strains and displacements, and would be
particularly challenging were the coil to be cooled in a liquid helium bath.
Pulse tube refrigerators exhibit lower vibration than Gifford-McMahon refrigerators
but suffer reduced efficiency when operated at an angle from vertical. We
selected a modest tilt of 7 degrees, and found that the coil temperature
increased by only 0.2 K to about 4.8 K. Although a spontaneous quench of this
magnet has never occurred, quenches due to interrupted compressor electrical
power or cooling do occasionally occur. A quench will result within about 30
minutes of an interruption, but is uneventful as there is no helium release.
The magnet temperature rises to about 50 K, and requires about 12 hours of
cooling to resume operation at 5 K or less. The cold head-induced periodic
field fluctuation is about 25 Hz peak-to-peak, and with a period of about 2 Hz causes
severe “motion” artifacts. Hypothesizing that the fluctuation was due to a
cyclic magnetization of the regenerator, we tried placing a superconducting
shield around the cold head, but this failed to eliminate the fluctuation. External
actively powered compensation coils about the cold head also failed; in this
case the 2 Hz magnetic waveform did not penetrate the superconducting shield. The
spectrometer is equipped with a memory that reads out a modulation of the spectrometer’s
numerically controlled oscillator (NCO) which generates the base operating
frequency. It runs asynchronously with respect to the pulse sequence. The uncompensated
B0 waveform was measured as a function of the phase of the cold head
pressure cycle, synchronized via the compressor’s pressure sensor. This waveform
was used to modulate the NCO, reducing the field fluctuation to about 1-2 Hz
peak-to-peak. Preliminary tests of the field variation while the magnet was in
motion (tilted up and down by several degrees at about a 1 Hz rate, or pushed
horizontally back and forth by ~50 cm) showed excellent stability of the NMR
spectrum.Discussion
We found that the periodic field fluctuation
that may occur with conduction cooled MRI magnets, where the cryocooler must
have tight thermal (and therefore possibly significant mechanical) coupling to
the main coil, can be compensated with feedforward compensation using the measured
uncompensated field waveform. This worked for our magnet, which exhibited primarily
a spatially uniform periodic field shift. If the field spatial uniformity also
fluctuates, a more complex higher order correction will be necessary, and cannot
be accomplished purely by the means reported here (i.e., frequency offset).
The static tilt of the system as
it is currently configured is not a barrier to magnet operation, and the motion
tests suggest that it may be possible to use such a system for dynamic
biomechanical studies in which the entire magnet and subject move together, thereby
suppressing motion artifacts due to relative motion of the subject with respect
to the magnet and gradient coil.Conclusions
A prototype extremity MRI scanner
based on a highly compact, conduction-cooled, three-bore cryogen-free 1.5T tilted
superconducting magnet is feasible. By proper compensation of cold head induced
field fluctuations, a highly stable magnetic field may be achieved. It is possible to translate and tilt the magnet while in operation.Acknowledgements
Funding was provided by grant R44AR065903
from the National Institute of Arthritis and Musculoskeletal and Skin Diseases.
I. C. received funding from the Fulbright Foundation.References
-
Pourrrahimi
S, Ackerman JL, Williams JEC, Pourrahimi N, Kaplan A. A compact affordable three-bore
cryogen-free superconducting magnet for extremity imaging, ISMRM, Honolulu, HI,
April 22-27, 2017.