Thomas O'Reilly1, Wouter Teeuwisse1, Bart de Vos2, and Andrew Webb1
1C.J. Gorter Center for High Field MRI, Leiden University Medical Center, Leiden, Netherlands, 2Circuits and Systems, Delft University of Technology, Delft, Netherlands
Synopsis
We show the first 3D in-vivo images acquired from our custom-built 50 mT
low-cost Halbach based portable MRI scanner. 3D Images of a knee were acquired
with a ~2 mm isotropic resolution using a 3D turbo spin echo sequence in less
than 12 minutes. Gradient non-linearity induced image distortions are minimal
within the central ~10 cm of the magnet bore length, but require correction
beyond this point. These results represent the latest step towards our goal of
creating a fully portable MRI scanner targeting pediatric neuroimaging in the
developing world for less than 30 000 euros.
Introduction
MR imaging has become a vital tool in clinical
settings, yet the high hardware cost as well as both technical and resource
challenges in keeping MRI scanners fully operational has resulted in a distinct
lack of access to this technology for a significant fraction of the world’s
population [1]. In this work we present progress on developing a low-cost
permanent magnet based MRI scanner targeted at imaging paediatric hydrocephalus
at a 3 x 3 x 5 mm resolution in low-resource settings. We previously presented
the design of a highly homogenous Halbach array
for generating the required B0 field [2]. In this work we have
incorporated gradient coils designed using a modified target-field approach [3]
to acquire in-vivo images of the knee. Method
The B0 field is produced by a 50
cm long, Halbach array with a 27 cm inner diameter (figure 1), a B0
field strength of 50 mT and a homogeneity of 2400 ppm over a 20 cm sphere in
the isocentre of the magnet [2], see figure 2. Gradient coils were designed
using the Target Field Method adapted to produce their encoding fields commensurate
with the transverse B0 direction of the Halbach array, see figure 3.
Gradients coils were constructed using 1.5 mm diameter copper wire pressed in
to 3D printed formers. The efficiencies of the X, Y and Z gradient coils are
0.59, 0.95 and 1.02 mT/m/A, respectively. The inductance of the X gradient is
180 µH with a resistance of 0.37 Ω, the inductance of the Y and Z gradients are both 225 µH with a resistance of 0.4 Ω. The
gradients were driven with two AE Techron 7224 and one AE Techron 2105
amplifiers (Elkhart, IN, USA). A Magritek Kea2 spectrometer (Aachen, Germany)
was used to generate gradient wave forms and RF pulses as well as digitising
the generated echoes. The RF pulses generated by the spectrometer were
amplified by a custom built 1 kW RF amplifier (adapted from [4]) and
transmitted to a 15 cm long, 15 cm diameter solenoid used as an RF transceiver
through a T/R switch built in to the spectrometer. The magnet was placed inside
a 80 cm long faraday cage constructed using aluminium plates and extrusion
profiles. To eliminate noise introduced in to the system by the human body, the
part of the body that extended out of the Faraday shield was wrapped in a
conductive fabric (Holland Shielding Systems, Dordrecht, the Netherlands). A
phantom was constructed of 8 mm diameter, 30 mm long tubes filled with water,
evenly spaced 17 mm apart on a rectangular grid to map geometric distortions in
the images due to gradient non-linearity.
Phantom images were acquired using a 2D
spin echo sequence with the following parameters: FoV: 200 x 200 x 30 mm for
the transverse images and 256 x 256 x 30 mm for the coronal images, acquisition
matrix: 128 x 128, TR/TE: 6000 ms/100 ms, acquisition bandwidth: 20 KHz, no
signal averaging, acquisition time: 12 minutes 48 seconds. In-vivo knee images
were acquired from a healthy volunteer using a T1 weighted 3D RARE sequence
with the following parameters: FoV: 256 x 256 x 256 mm, acquisition matrix: 128
x 128 x 128, TR/TE: 130 ms/15 ms, acquisition bandwidth: 20 KHz, echo train
length: 3, no signal averaging, acquisition time: 11 minutes 50 seconds. The RF
pulses were non-selective. The acquired k-space data was filtered using a
squared sine bell filter and reconstructed using the Numpy v1.16 FFT library in
Python 3.7.Results
Figure 4 shows
images acquired of the geometric phantom. No significant geometric distortions
are visible in the central transverse plane, however, clear warping is visible
in the coronal plane as the distance from the isocentre increases, closely
matching simulations of the fields generated by the gradient coils, and with
relative magnitudes similar to those on conventional clinical scanners. Figure 5 shows
in-vivo images acquired of the knee. Major structures in the knee can be
differentiated with clear contrast between muscle, fat and cartilage. Image
distortions, particularly visible in the coronal image, closely match those
seen in Figure
4. Discussion
In-vivo images have been acquired with an
SNR of 26, sufficient to clearly delineate different structures in the knee
joint and show the feasibility of neuroimaging in paediatrics at, or exceeding,
the targeted 3 x 3 x 5 mm resolution. Images are shown without correction for
gradient non-linearities (the main cause of artefacts in these images). Simulations
indicate temperature induced B0 drift causes by heat radiated from
the body, typically less than 1 kHz an hour, only causes minimal blurring of
the reconstructed images with a full-width-half-maximum (FWHM) of the point
spread function (PSF) of around 1.4 pixels, comparable to the FWHM of the PSF
caused by T2 relaxation in the muscle during the RARE echo train. Future work
in this project will focus on implementing standard gradient distortion
algorithms based on local Jacobian scaling factors [5], as well as switching
the gradient amplifier and spectrometer to low-cost open source alternatives to
create a fully open-source imaging system for neuroimaging of infants in
low-resource settings for less than 30 000 euros.Acknowledgements
This work is supported by the following grants:Horizon 2020 ERC FET-OPEN 737180 Histo MRI, Horizon 2020 ERC Advanced NOMA-MRI 670629, Simon Stevin Meester Award and NWO WOTRO Joint SDG Research Programme W 07.303.101.References
[1] World Health Organization ‘Baseline country survey on medical
devices, 2014 update’, March 9, 2016, http://gamapserver.who.int/gho/interactive_charts/health_technologies/medical_equipment/atlas.html?&indicator=i1
Accessed November 5, 2019
[2] O’Reilly et al. “Three-dimensional MRI
in a homogenous 27 cm diameter bore Halbach array magnet”, J Magn Reson. 307 (2019).
[3] Turner “A target field approach to
optimal coil design”, J. Phys. D, 19(8), 1986
[4] Blücher et al "COSI Transmit: Open Source Soft-
and Hardware Transmission System for traditional and rotating MR", Proc
ISMRM, 2017, #184
[5] S.J. Doran,
L. Charles-Edwards, S.A. Reinsberg, M.O. Leach, "A complete distortion correction
for MR images: I. Gradient warp correction," Physics in medicine and
biology 50, 1343-1361 (2005).