Ole Geldschläger1, Dario Bosch1, Nikolai Avdievitch1, Klaus Scheffler1,2, and Anke Henning1,3
1High-field Magnetic Resonance, Max-Planck-Institut for biolog. Cybernetics, Tübingen, Germany, 2Institute for Biomedical Magnetic Resonance, University Hospital Tübingen, Tübingen, Germany, 3Advanced Imaging Research Center, University of Texas Southwestern Medical Center, Dallas, TX, United States
Synopsis
This study presents the first investigations with algorithmic
spinal cord-segmentation, as well as gray matter/white matter-segmentation
within the spinal cord, at the ultrahigh field strength of 9.4T. On multi-echo
gradient-echo acquisitions from three subjects, the tested algorithms perform
the segmentations correctly. Based on these multi-echo data, pixel-wise T2*-relaxation
time maps were calculated. By means of the segmentations, averaged T2*-times of
24.88ms +- 6.68ms for gray matter and 19.37ms +- 8.66ms for white matter, were
calculated.
Introduction
Nowadays, for clinical diagnostic and research
investigations of the human spinal cord (SC), magnetic-resonance-imaging (MRI)
is an indispensable method. Tissue changes in gray matter (GM) and white matter
(WM) within the SC have been associated to several neurological disorders such
as SC injury, neoplastic lesions1, or multiple
sclerosis2. Based on 3T MRI-acquisitions,
the segmentation of the entire SC is routinely done to measure SC atrophy3 or to support
surgical planning4. Recently, SC-detection
and GM-segmentation of 7T human SC images were performed manually5 as well as
automatically6 using the segmentation algorithms of the spinal cord toolbox7.
This work presents the first investigations of algorithmic
SC-detection, as well as GM/WM- segmentation of high-resolution human SC images
acquired at 9.4T. Furthermore, as tissue relaxation rates change with
increasing B0 field strength8, T2*-relaxation time
maps are calculated and T2*-times of GM and WM in the human SC at 9.4T are
presented.Methods
All experiments were
performed on a Siemens (Erlangen, Germany) Magnetom 9.4T whole-body MRI
scanner. Analog to Geldschläger et al.9, for RF-transmission
and reception, a 16-channel tight-fit array10, consisting of eight
transceiver surface loops and eight receive-only vertical loops, was used. The
coil was originally constructed for human brain scans with emphasis to enhance
the Signal-to-Noise-Ratio (SNR) in deep structures, but through subject
replacing (Figure
1)
it can be used to image the cervical SC, as well.
9.4T anatomical imaging data
was acquired using an axial T2*-weighted multi-echo gradient-echo (GRE)-sequence
(field of view: 140x140mm2, In-plane res.: 0.23x0.23mm2, Number of slices: 12, slice thickness: 3mm, Repetition time (TR):
500ms, FA: 50°, 4 echoes, Acquisition time: 5min:04sec, 2D). On three different
healthy volunteers, the sequence was applied twice, respectively: once with the
echo times (TEs) 4, 10, 16 and 22ms and once with the TEs 5.1, 9.4, 13.8 and
17.7ms (analog to the TEs from Massire et al.6).
The T2*-weighted anatomical imaging data acquired at a TE of 9.4ms
were segmented with the Spinal-Cord-Toolbox (SCT)7
(PropSeq-algorithm11 for SC-segmentation
and deep-learning-algorithm12
for GM-segmentation).
Pixel-wise T2*-time calculation was performed with a
mono-exponential fit, based on the eight sample points per pixel acquired with
the multi-echo GRE-sequence. The fitting was performed with a nonlinear least-squares algorithm using Matlab (The Mathworks, Natick ,MA, USA).
Using the beforehand obtained GM/WM-masks, the T2*-time of
GM and WM were averaged.Results
In Figure 2,
the anatomical imaging results from one subject, are depicted. It is visible, that with increasing TE, the signal
decays. For a TE between 9.4 ms and 13.8 ms a good GM/WM contrast is achieved
and a TE of 9.4 ms is the best compromise between SNR and contrast. At longer
TEs, signal-dropouts can be seen in slices located next to intervertebral disks.
Represented on the example of three slices acquired from
another subject, Figure 3
shows the segmentation results. Both, the SC-segmentation and the
GM-segmentation algorithm delivers the correct masks. For the three scanned
volunteers, the segmentation algorithms produced correct results for all
acquired slices when using images acquired with a TE of 9.4ms.
Figure 4
shows an example slice of a T2*-map. The SC is recognizable and the GM and WM
within the SC is distinguishable. T2*-slices in which these tissue types were
not recognizable at all, were excluded from the averaging. For the cervical SC
at 9.4T the calculated mean T2*-time for GM is 24.88ms +- 6.68ms and for WM
19.37ms +- 8.66ms.Discussion
It was shown, that algorithmic
SC-segmentation and GM/WM-segmentation is possible using high-resolution T2*-weighted
images acquired at 9.4T with the proposed brain coil. The employed algorithms
from the SCT7 produced correct SC
and GM/WM masks for all tested datasets. That leads to the assumption that the
algorithms work reliably for 9.4T human SC images and have a low probability to
produce wrong segmentations in general.
The presented averaged T2*-times
in the human SC at 9.4T are consistent with the literature, although an
uncertainty remains when using this pixel-by-pixel fit method (see noise in Figure
4 and the resulting relatively high standard deviation). Additionally, the subject database size was relatively
small (3 subjects).
As expected, the T2*-times
from the cervical SC at 7T (GM: 29.3ms +- 4.5ms, WM: 23.5ms +-5.7ms)6 are
slightly higher, while the T2*-times at 9.4T for the human brain (GM: 23.8ms +- 1.0ms, WM: 19.2ms +- 0.9ms)13, are very similar to the values calculated for the
SC at 9.4T.
Through the application of other sequences (for example
MP2RAGE14 or T2W) in the future, T1- and
T2-relaxation time measurements need to be performed, in order to complete the
set of tissue relaxation times for the SC at 9.4T.Conclusion
With this work the next step into SC research at a B0 field strength
higher than 7T was done. With a coil, originally optimized for maximal SNR in
deep brain structures, we were able to acquire high-resolution anatomical SC
data on which algorithmic SC-segmentation, as well as GM/WM-segmentation,
reliably works. The T2*-times of GM and WM in the human SC at 9.4T were
presented. This
might open new possibilities in the field of SC-research and clinical patient
care at ultra-high-field. Knowing the T2*-relaxation time allows optimization of imaging
parameters and potentially enables improvements in sensitivity and contrast.Acknowledgements
Funding by the
European Union (ERC Starting Grant, SYNAPLAST MR, Grant Number: 679927) is
gratefully acknowledged.References
1.
Amukotuwa SA, Cook
MJ. SPINAL DISEASE: NEOPLASTIC, DEGENERATIVE, AND INFECTIVE SPINAL CORD
DISEASES AND SPINAL CORD COMPRESSION. In Neurology and Clinical
Neuroscience.: Elsevier; 1999. p. 511-538.
2.
Schlaeger R, Papinutto N, Panara V, Bevan C, Lobach IV,
Bucci M, et al. Spinal cord gray matter
atrophy correlates with multiple sclerosis disability. Annals of Neurology.
2014 Aug; 76: 568-580.
3.
Lin X, Tench CR,
Evangelou N, Jaspan T, Constantinescu CS. Measurement of Spinal Cord Atrophy
in Multiple Sclerosis. Journal of Neuroimaging. 2004 Jul; 14: 20-26.
4.
Mukherjee DP, Cheng
I, Ray N, Mushahwar V, Lebel M, Basu A. Automatic Segmentation of Spinal Cord
MRI Using Symmetric Boundary Tracing. IEEE Transactions on Information
Technology in Biomedicine. 2010 Sep; 14: 1275-1278
5.
Sigmund EE, Suero GA,
Hu C, McGorty K, Sodickson DK, Wiggins GC, et al. High-resolution human
cervical spinal cord imaging at 70.167emT. NMR in Biomedicine. 2011 Dec; 25:
891-899.
6.
Massire A, Taso M,
Besson P, Guye M, Ranjeva JP, Callot V. High-resolution multi-parametric
quantitative magnetic resonance imaging of the human cervical spinal cord at
7T. NeuroImage. 2016 Dec; 143: 58-69.
7.
Leener BD, Lévy S, Dupont SM,
Fonov VS, Stikov N, Collins DL, et al. SCT: Spinal Cord Toolbox, an
open-source software for processing spinal cord MRI data. NeuroImage. 2017
Jan; 145: 24-43.
8.
Chavhan GB, Babyn PS,
Thomas B, Shroff MM, Haacke EM. Principles, Techniques, and Applications of
T2∗-based
MR Imaging and Its Special Applications. RadioGraphics. 2009 Sep; 29:
1433-1449.
9.
Geldschläger O, Manohar SM,
Wright A, Avdievitch N, Henning A. First MRI of the human spinal cord at
9.4T. In 27th ISMRM; 2019; Montreal, Canada.
10.
Avdievich NI,
Giapitzakis IA, Pfrommer A, Borbath T, Henning A. Combination of surface and
`vertical' loop elements improves receive performance of a human head
transceiver array at 9.4 T. NMR in Biomedicine. 2017 Dec; 31: e3878.
11.
Leener BD, Kadoury S,
Cohen-Adad J. Robust, accurate and fast automatic segmentation of the spinal
cord. NeuroImage. 2014 Sep; 98: 528-536.
12.
Perone CS, Calabrese
E, Cohen-Adad J. Spinal cord gray matter segmentation using deep dilated
convolutions. Scientific Reports. 2018 Apr; 8.
13.
Pohmann R, Speck O,
Scheffler K. Signal-to-noise ratio and MR tissue parameters in human brain
imaging at 3, 7, and 9.4 tesla using current receive coil arrays. Magnetic
Resonance in Medicine. 2015 Mar; 75: 801-809.
14.
Marques JP, Kober T,
Krueger G, Zwaag W, Moortele PFV, Gruetter R. MP2RAGE, a self bias-field
corrected sequence for improved segmentation and T1-mapping at high field.
NeuroImage. 2010 Jan; 49: 1271-1281.