Agazi Samuel Tesfai1, Andreas Vollmer2, Moritz Braig1, Johannes Fischer1, Ute Ludwig1, and Michael Bock1
1Dept. of Radiology, Medical Physics, Medical Center - University of Freiburg, Freiburg, Germany, 2Dept. of Oral and Craniomaxillofacial Surgery, Center for Dental Medicine, Medical Center - University of Freiburg, Freiburg, Germany
Synopsis
Detection of root canals is vital for dental diagnosis, however it is difficult to
locate these anatomies within sub-millimeter dimension. To determine the
ability of MRI to display such structures, a bovine tooth with different sized artificial cavities was prepared. It was evaluated with a
preclinical 7T system and a clinical 3T MR
system against cone beam CT. 7T measurements with UTE offer precise distinction of cavities
up to 200µm. 3T UTE allows only differentiation of 1000µm cavities due to blurring. Tooth immersed in contrast agent solution allows localization up to 200µm cavity with spin echo sequence and negative contrast.
Introduction
Dental root canals show large anatomical variabilities
which stimulates the interest in non-invasive imaging methods1. Since the prognosis of periapical treatments depends on the
preparation and disinfection of root canals, reliable imaging of the canals is
mandatory2,3. Current standard methods for apical imaging are periapical radiography
(PR) and cone beam computed tomography (CBCT)4 both of which only offer a very limited root canal contrast.
To distinguish structures in dentin and enamel (e.g., hard
substance fractures and gaps between restorations and healthy tooth substance),
a spatial resolution of 200 μm is required.
In MRI, dentin and enamel are thus difficult to visualize due to the extremely
short T2* values5, and the required high spatial resolution leads to
low Signal-to-Noise (SNR) which further limits root canal visualization in MRI.
To overcome the rapid signal decay, special sequences such as ultra-short echo
time (UTE)6 are required to minimize TE and maximize signal
intensity7. In this work we test UTE MRI methods to visualize small
structures in dental material at a preclinical high-field and a clinical MR
system, and imaging results are compared to CBCT.Materials and Methods
For all imaging experiments an extracted bovine tooth
was used which was placed in 0.1 % thymolic solution. Five different reference
cavities (1000µm, 600µm, 400µm, 300µm and 200µm) were milled into dentin and
enamel to emulate side root canals of different sizes as visible in Figure 1. MRI
was performed on a 7T Bruker Biospec System (Bruker bioSpin GmbH, Ettlingen,
Germany) with a cryogenically cooled (77 K) Tx/Rx RF probe. A 3D UTE sequence was
applied with TR of 8ms, TE=70µs, α=5°, BW=1875Hz/Pixel, 1 average,
isotropic resolution=100µm, FoV=32mm and TA=19 min. As a reference, a CBCT
(Orthophos SL, 943 mGycm2, TR=14.2 ms, tube voltage 85 kV, tube current 6 mA,
Dentsply Sirona GmbH Germany) and PR (VistaScan Mini VI, TR=0.06 s, tube
voltage 70 kV, tube current 7 mA, Dentsply Sirona GmbH Germany) was performed.
Furthermore, MRI data
were acquired at a 3T clinical MR system (PRISMAFit, Siemens Healthineers,
Erlangen, Germany) with a Siemens loop coil (4cm diameter) and with a 3D UTE sequence:
TR=5 ms, TE=70 µs, α=16 °, FoV=50 mm³, BW=2000Hz/Pixel, 2 averages, nominal isotropic
resolution=361µm, FoV=50mm, TA=35 min and 260,000 radial spokes. Additional measurements were acquired with
different TEs (70, 100, 200, 400, and 800µs). Afterwards, the tooth was
immersed in a contrast agent solution (1% Multihance), and images were acquired
with a 2D Turbo Spin Echo protocol: TR=500 ms,
TE=50 ms, α=90 °, BW=279 Hz/Pixel, 5 averages, base resolution 0.2mm*0.2mm, FoV=50
mm³, BW=279 Hz/Pixel, TA=10 min and slice thickness of 0.5mm.Results
The 3D UTE sequence allows clear depiction of hard
tissue at 7T (Figure 2). All cavities are clearly distinguishable at a
resolution of 100 µm at transversal (2a) and sagittal (2b) orientation along
with other tissues such as dentin, dental pulp, the root canal and cementum.
Figure 2c) shows a line profile along sagittal orientation with clear
differentiation of the cavities. With the UTE sequence at 3T, dental tissues
like the root canal, dentin, cementum and dental pulp are also visible in
Figure 3 and calculation from multiple TEs yield an average T2* relaxation time
for dentin of 250µs. The biggest cavity (1000µm) is discernable despite
blurring. Immersing bovine tooth in contrast solution allows visualization of
all five cavities (in enamel and dentin respectively) as shown in Figure 4 along with
corresponding signal intensity profile. CBCT measurement also shows all
cavities on dentin and enamel tissue as reference.Discussion
At 7T all cavities are clearly visible with high base resolution
(100 µm), because preclinical MRI systems offer higher slew rates and
amplitudes, so that a higher spatial resolution is feasible with virtually no
blurring artifacts and superior to CBCT images (Figure 1). At 3T, the nominal spatial
resolution is only 0.361 µm³, and a further blurring is expected due to very fast
T2* decay, which can be calculated from the readout bandwidth and the T2* to FWHM
= 2.21 pixel (i.e., 800 µm)8. Filling the drilled holes with a
contrast agent solution solves this problem, as now a conventional MRI sequence
with higher spatial resolution can be used. This, however, might not always be
possible with the root canals, even though in X-ray imaging also contrast
agents are applied for visualization.
The
high spatial resolution requirements and the short T2* are currently
challenging for clinical MRI systems with limited gradient amplitude (up to 80
mT/m) and slew rate (here: 200 T/m/s), which might be overcome with dedicated
high-power gradients (Connectome project). Adding contrast agents offers new
possibilities for dental MRI (Figure 4) as it allows imaging of hollow
structures with conventional acquisition times (10 min) and higher in plane
resolution.Conclusion
Preclinical 7T MRI systems can distinguish structures
down to 0.2 mm size, whereas at clinical 3T MRI systems blurring limits the
spatial resolution to about 1mm, which can be improved if contrast agents are
used. Thus, MRI might be an alternative to X-Ray modalities (CBCT) in
situations when ionizing radiation is to be avoided. Acknowledgements
Grant
support from the Deutsche Forschungsgemeinschaft (DFG) under grant numbers BO
3025/8-1 and UL 1187/6-1 is gratefully acknowledged.References
[1] Martins J et al.
Prevalence Studies on Root Canal Anatomy Using Cone-beam Computed Tomographic
Imaging: A Systematic Review. J Endod
45, 372-386.e4 (2019).
[2] Timme M. et al. Imaging of root canal
treatment using ultra high field 9.4T UTE-MRI - a preliminary study. Dentomaxillofac Radiol 20190183
(2019)
[3] Colakoglu G et al.
Evaluation of the internal anatomy of paramolar tubercles using cone-beam
computed tomography. Surg Radiol Anat
(2019)
[4] De Paula-Silva, F. W.
G., Wu, M.-K., Leonardo, M. R., da Silva, L. A. B. & Wesselink, P. R.
Accuracy of periapical radiography and cone-beam computed tomography scans in
diagnosing apical periodontitis using histopathological findings as a gold
standard. J Endod 35,
1009–1012 (2009).
[5] Hövener JB et al. "Dental
MRI: imaging of soft and solid components without ionizing
radiation." Journal of
Magnetic Resonance Imaging 36.4: 841-846. (2012)
[6] Gatehouse PD et al. Magnetic resonance imaging of
short T 2 components in tissue. Clinical
radiology, 58(1), 1-19. (2003)
[7] Robson et
al. Clinical ultrashort echo time imaging of bone and other connective tissues.
NMR in Biomedicine, 19(7), 765-780 (2006).
[8] Rahmer J et al. Three‐dimensional radial ultrashort echo‐time imaging with T2 adapted sampling. Magnetic Resonance in Medicine: An Official
Journal of the International Society for Magnetic Resonance in Medicine 55.5:
1075-1082. (2006)