Berk Silemek1, Lukas Winter1, Frank Seifert1, Harald Pfeiffer1, and Bernd Ittermann1
1Physikalisch-Technische Bundesanstalt, Berlin, Germany
Synopsis
A Measurement-based Temperature-Matrix approach is presented that enables a fast, patient and exam specific estimation and mitigation
of RF hazards of implants. Various locations in phantom are tested using an 8-channel (300MHz) implant safety testbed. Heating reduction Based on T-Matrix Measurements resulted >3 times heating reduction vs. circularly-polarized mode and >19 times vs. worst-case mode. 2-channel MRI (3T) feasibility experiments using high temperature resolution showed good correlation with transmitted power. In addition, T-matrix-based temperature increase predictions successfully demonstrated. As summary, an easy to implement, cheap,
sensor-based method, the T-matrix, to investigate, characterize and
mitigate RF heating of implants is introduced.
Introduction
Parallel transmission (pTx) is an effective method to
substantially reduce RF induced heating in medical implants1–3. Simulation and in-vitro-based
mitigation approaches cannot provide a patient and exam specific assessment. Approaches
that rely on integrated sensors in and around the implant allow for a
patient-specific, real-time mitigation, while simultaneously the sensors provide
an online safety monitoring4–6.
In this work, we introduce the temperature matrix (T‑matrix), a pTx monitoring and mitigation approach based on a
single thermistor (0.1$) that can easily be integrated within the implant. This
approach enables a fast, patient and exam specific estimation and mitigation of
RF hazards, in principle without the need for prior simulations or in vitro
testing. Feasibility is demonstrated by applying the T-matrix for heating prediction and mitigation using an 8-channel
pTx coil at 300MHz in an implant-safety testbed and a 2-channel body coil
inside a 3T MRI scanner. Methods
Instrumentation:
An implant safety testbed with an 8-channel 7T pTx RF coil6 and a cylindrical PVP based phantom is
used for benchtop experiments (Fig.1a-b).
A semi-rigid coaxial cable with an uninsulated tip is
used as a guidewire substitute (Fig. 1b). An NTC thermistor (NCP18XH103F03RB,
Murata) is connected between inner and outer conductor at the tip of the cable
(Fig. 1c).
T-matrix:
Tip heating impulse responses based on a complex voltage excitation vector $$$\vec{u}$$$(n) of an n-channel array are measured and used
to calculate the Hermitian T-matrix $$$\overset{\text{$\leftrightarrow$}}{T_M}$$$: $$\Delta T = \vec{u}^H\overset{\text{$\leftrightarrow$}}{T_M}\vec{u} \label{eq:Tmat} $$
To determine $$$\overset{\text{$\leftrightarrow$}}{T_M}$$$, RF pulses with phase differences 0° (i<j)
or 90° (i>j) are applied to channels i,
j and the corresponding temperature
rise, ΔTij, is recorded. In
order to demonstrate a proof of concept mitigation of RF induced heating, an
orthogonal projection (OP) mode was applied, which is the projection of the
circular polarized (CP) vector onto the subspace orthogonal to the $$$\overset{\text{$\leftrightarrow$}}{T_M}$$$ eigenvector with the largest eigenvalue. OP
allows to mitigate RF induced heating, while maintaining sufficient B1+ homogeneity
for imaging6,7. The worst case
(WC) is defined using the phases of the $$$\overset{\text{$\leftrightarrow$}}{T_M}$$$ eigenvector with the largest eigenvalue and
equal amplitudes, which corresponds to the worst possible RF induced heating using
a pTx transmission system with limited forward power per channel.
Benchtop Experiments (300MHz):
Repeatability was tested with 10 $$$\overset{\text{$\leftrightarrow$}}{T_M}$$$
measurements
at the same location (64 cycles: 2s RF heating/2s cooling). The T-matrix (64 cycles: 1s RF-heating/1s
cooling) was evaluated on three locations within the phantom. The transmitted
average power over the acquisition window was 3.14W. RF heating experiments in CP,
OP and WC mode were performed with the same total forward power of 2.06W at Pos-1,
2.47W at Pos-2 and 2.41W at Pos-3 (Fig.3).
MRI Experiments (3T):
To investigate the feasibility of the $$$\overset{\text{$\leftrightarrow$}}{T_M}$$$
concept under
pulsed MRI conditions, a 3T scanner (Siemens, Verio) with a 2-channel body coil
(quadrature splitter of the system was removed, and two ports were connected to
a pTx system) was utilized Fig.4a. A conventional GRE sequence was used (TR:15ms,
pulse length=2ms) to determine the relation between
pulsed power vs. temperature increase at the tip. Results
Benchtop Experiments (300 MHz):
The temporal resolution of the thermistor reading was ~70ms with a temperature
resolution of ~100μK. Repeatability experiments showed <3% standard
deviation in slopes and eigenvalues for 10 $$$\overset{\text{$\leftrightarrow$}}{T_M}$$$ measurements
(Fig.2).
was acquired in 128s with a total temperature
rise of <0.1°C at the guidewire tip. Based on the $$$\overset{\text{$\leftrightarrow$}}{T_M}$$$, the
calculated OP mode was able to substantially reduce RF induced heating at the
guidewire tip at all three locations (>3 times vs. CP mode and >19 times
vs. WC mode, Fig. 3).
MRI Experiments (3T): MR B1+ image artifacts due to the induced current on
the guidewire are displayed in Fig. 4b-g. The temperature increase during a 1s
GRE sequence correlates well with transmitted power (Fig. 5a-b), demonstrating
the feasibility of using the thermistor in MR experiments. $$$\overset{\text{$\leftrightarrow$}}{T_M}$$$ was
acquired at two positions (Figs. 5c,e) in a total of 32s and with an overall
temperature increase at the tip of <0.1°C. $$$\overset{\text{$\leftrightarrow$}}{T_M}$$$ contains
the full system information and can also be used to predict heating. These
predictions correlate well with actual measurements for different pTx driving
conditions changing both amplitude (Fig. 5d) and phase (Fig. 5f). Discussion
The T-matrix can
be acquired with good repeatability and speed, requiring only low flip angles.
The pTx-based method is able to detect, predict and mitigate guidewire tip
temperatures as demonstrated for a 2-channel body-coil at 3T and an 8-channel
pTx RF coil for 7T. Based on temperature measurements (not fields or currents),
it allows to directly monitor and manage the applied thermal dose9.
In future experiments more implant locations, implant orientations and implant
geometries will be tested with the introduced method. Furthermore, B1+
homogeneity will be included in the cost function to reduce tip heating.Conclusions
An easy to implement,
sensor-based method, the T-matrix, to investigate, characterize and
mitigate RF heating of implants is introduced. The technique is fast and does
not require additional simulations or prior in vitro testing. It is thus attractive
for pTx based temperature control and monitoring of implant temperatures for
patient-specific and exam-specific safety assessments. Acknowledgements
This work was funded by the EMPIR grant 17IND01
MIMAS. The EMPIR initiative is co- funded by the European Union’s Horizon 2020
research and innovation program and the EMPIR participating states.References
1. Eryaman Y,
Guerin B, Akgun C, et al. Parallel transmit pulse design for
patients with deep brain stimulation implants. Magn Reson Med.
2015;73(5):1896-1903.
2.
McElcheran CE, Golestanirad L,
Iacono MI, et al. Numerical Simulations of Realistic Lead Trajectories and an
Experimental Verification Support the Efficacy of Parallel Radiofrequency
Transmission to Reduce Heating of Deep Brain Stimulation Implants during MRI. Sci
Rep. 2019;9(1).
3.
Guerin B, Angelone LM, Dougherty D,
Wald LL. Parallel transmission to reduce absorbed power around deep brain
stimulation devices in MRI: Impact of number and arrangement of transmit
channels. Magn Reson Med. 2020;83(1):299-311.
4.
Etezadi-Amoli
M, Stang P, Kerr A, Pauly J, Scott G. Controlling radiofrequency-induced
currents in guidewires using parallel transmit. Magn Reson Med.
2015;74(6):1790-1802.
5.
Silemek B, Acikel V, Oto C, et al.
A temperature sensor implant for active implantable medical devices for in vivo
subacute heating tests under MRI. Magn Reson Med. 2018;79(5).
6. Winter L, Seifert F, Hoffmann W, et al. Parallel transmission medical implant
safety testbed: First application using time-domain E-field probes to measure
and mitigate RF induced currents. Proc . ISMRM 2019:0728
7.
Seifert F,
Weidemann G, Ittermann B. Q matrix approach to control implant heating by transmit
array coils. Proc. ISMRM 2015:3212.
8. Han
H, Moritz R, Oberacker E, Waiczies H, Niendorf T, Winter L. Open Source 3D
Multipurpose Measurement System with Submillimetre Fidelity and First Application
in Magnetic Resonance. Sci Rep. 2017;7(1):13452.
9.
Van Rhoon GC, Samaras T, Yarmolenko
PS, Dewhirst MW, Neufeld E, Kuster N. CEM43°C thermal dose thresholds: A
potential guide for magnetic resonance radiofrequency exposure levels? Eur
Radiol. 2013;23(8):2215-2227.