Rasim Boyacioglu1, Megan Poorman2,3, Kathryn Keenan2, and Mark Griswold1
1Radiology, Case Western Reserve University, Cleveland, OH, United States, 2Physical Measurement Laboratory, National Institute of Standards and Technology, Boulder, CO, United States, 3Department of Physics, University of Colorado Boulder, Boulder, CO, United States
Synopsis
Conventional temperature
monitoring is based on measurement of off-resonance via gradient-echo phase scans
for non-adipose tissue. MRF with quadratic RF phase (MRFqRF) simultaneously quantifies
off-resonance, T1, T2, and T2*. For a proof of principle thermometry experiment
with MRFqRF, an ex-vivo aqueous sample was heated with laser ablation,
temperature was tracked, and multiple continuous MRFqRF scans were obtained
with different temporal resolutions. Scanner frequency drifts were removed
automatically with Independent Component Analysis. Residual changes in
off-resonance predict the temperature change. However, MRFqRF temporal
resolution (~10s) needs to be increased further for clinical relevance.
Introduction
Magnetic Resonance Fingerprinting1
(MRF) is a quantitative imaging technique for simultaneous mapping of multiple
tissue properties and system parameters. Recently, MRF was extended to map T2*
and off-resonance (Δf) alongside T1 and T22. MRF with quadratic RF
pulses (MRFqRF) sweeps the on-resonance frequency linearly between -1/(2*TR) to
1/(2*TR) in time by modulating the RF phase with a quadratic function (Figure
1). Dynamic sweeping of off-resonance frequencies makes the signal evolutions
different for all the time points and stable against random noise (Figure 2). In
this work, stability of MRFqRF for off-resonance mapping was explored for use
in temperature monitoring of aqueous tissues. However, the conventional temporal
resolution of MRFqRF is much slower than what is required for temperature
monitoring of thermal therapies (40s vs 3s). Here, we performed proof of
concept experiments with the MRFqRF framework to optimize it for fast and
continuous mapping of Δf for temperature quantification.Methods
Experimental setup
A tofu phantom was immersed in a water
bath and placed at isocenter of a 3T MRI scanner (Skyra, Siemens). The laser
ablation device’s applicator (LaserPro 980, PhotoMedex) and a fiber optic temperature
probe (Luxtron, LumaSense Technologies) were placed together in the center of the
phantom. A second temperature probe was positioned inside the phantom away from
the site of the heating for reference measurements. For each experiment, the tofu
was heated with the laser ablator for ~20 s at 1.5 W of Continuous Wave power,
followed by subsequent cooling (total of
6:40 minutes per experiment).
Data acquisition
To approach the temporal
resolution of conventional proton resonant frequency-shift thermometry (~3s), the
number of time points per one MRFqRF scan was reduced. This was achieved by
sweeping the off-resonance frequency faster with respect to MRFqRF with 3516
time points (0.72 Hz/TR) by adjusting the quadratic RF function (2x in 1752
time points with 1.44 Hz/TR and 4x in 876 time points with 2.88 Hz/TR). The
original flip angle pattern was resampled, and the same four block structure
was preserved for consistency. The phantom setup described above was scanned
for the 6:40 minutes with three different 2D MRFqRF versions during heating and
cooling of the phantom.
Reconstruction and post-processing
2D MRFqRF data were reconstructed
after compression with randomized SVD3. The compressed data were
matched to a compressed dictionary, which was also partially undersampled in
the tissue dimension. The low resolution in the tissue dimension was recovered
with quadratic interpolation4 after the matching.
The relatively strong gradients
needed for MRFqRF acquisition can cause eddy currents and frequency drifts over
time (~4 Hz/min), which need to be removed to see underlying off-resonance
changes due to temperature. For automated estimation of frequency drift, Independent
Component Analysis (ICA)5 was run on the dynamic Δf
series of an 11x11 ROI from a region where no temperature change is expected.
The component that describes the linear frequency drift was then removed from
all voxels in the dynamic Δf series.
Finally, the reconstructed
off-resonance maps were converted to ΔT with the following equation6,
where $$$γ$$$ is the gyromagnetic ratio in MHz/T, B0
is the field strength in Tesla, and n is the dynamic scan number.
$$
\triangle T_n=\frac{\triangle f_n - \triangle f_0}{-0.01*\gamma*B_0} $$Results
Figure 3 compares the signal
evolutions of two different dictionary entries with matching T1 and T2 for MRFqRF
3516/1752/876. Inner product of signal evolutions is similar across number of
time points, but SNR is reduced as the number of time points decrease. The ΔT of
the laser ablation experiment computed from MRFqRF Δf maps using 3516/1752/876 time
points is plotted in Figure 4. All three acquisitions were able to detect
heating. Reducing the number of time points brings finer resolution sampling of
the heating curve through time, but also increased noise, which can be inferred
from the standard deviation over voxels around the not heated point.Discussion
While heating occurred relatively
quickly (~1 °C/s),
cooling was much slower, allowing even the slowest MRFqRF version (40s per
scan) to capture the general trend of temperature. Relatively stable MRFqRF 876
can sample the heating curve 4x faster, but it is also significantly slower
than clinical need. The SNR with faster MRFqRF translates into slightly noisier
ΔT
estimations, but the relatively stable inner product is still the dominant
factor. This observation might be useful for future optimization.
Frequency drift correction is extremely
important to be able to focus on small off-resonance fluctuations triggered by changes
in temperature. Simple summation of voxel time series or fitting a linear
function might be suboptimal considering the large fluctuations that are
possible (Figure 5, left). ICA is able to detect the frequency drift component
consistently (Figure 5, right).Conclusions
This work demonstrates the first
steps for temperature monitoring with an MRF framework. However, clinically-feasible
ablations would benefit from further increases in temporal resolution with
faster sampling of the heating curve. Additionally, no information can be
obtained about heating in adipose tissue with this current implementation due
to the use of off-resonance changes for temperature computation. Improvements to
acquisition speed and contrast in multiple tissue types is necessary, before it
can be used in real-time treatment monitoring and is the subject of future work7.Acknowledgements
The work was supported by Siemens
Healthcare and a cooperative grant between the University of Colorado and NIST.
Authors would like to thank Will Grissom for use of the temperature probes.References
1. Ma D, Gulani V, Seiberlich N,
et al. Magnetic resonance fingerprinting. Nature 2013;495: 187–192.
2. Wang C, Coppo S, Mehta B, et al. Magnetic resonance
fingerprinting with quadratic RF phase for measurement of T2* simultaneously
with δf
,T1, and T2. Magn Reson Med 2019 81(3):1849-1862.
3. Yang M, Ma D, Jiang Y, et al. Low rank approximation
methods for MR fingerprinting with large scale dictionaries. Magn Reson Med
2018 79(4):2392-2400.
4. McGivney
D, Boyacioglu R, Jiang Y, et al. Towards Continuous Tissue Property Resolution
in MR Fingerprinting using a Quadratic Inner Product Model. ISMRM proceedings
2019, p4526.
5. FastICA
toolbox (http://research.ics.aalto.fi/ica/fastica/)
6. Rieke V and Butts Pauly K. MR Thermometry. Journal of
Magnetic Resonance Imaging 2008; 27(2):376-390
7. A
feasibility study of Magnetic Resonance Fingerprinting for multi-contrast
temperature mapping in both aqueous and adipose tissues. Submitted to ISMRM 2020.