Marlon Pérez-Rodas1,2, Hildegard Schulz1, Rolf Pohmann1, Klaus Scheffler1,3, and Rahel Heule1
1High-Field MR Center, Max Planck Institute for Biological Cybernetics, Tübingen, Germany, 2Graduate Training Centre of Neuroscience, IMPRS for Cognitive and Systems Neuroscience, University of Tübingen, Tübingen, Germany, 3Department of Biomedical Magnetic Resonance, University of Tübingen, Tübingen, Germany
Synopsis
To fully understand the neurovascular fingerprint observed in
BOLD experiments, extravascular and intravascular contributions have to be
identified separately. Balanced steady-state free precession (bSSFP) imaging
has demonstrated the ability for distortion-free fMRI with high microvascular
sensitivity. However, the underlying intravascular contribution to BOLD bSSFP
is not yet entirely known as literature R2 relaxation rates do not
reflect the apparent diffusion-related R2 decrease in blood with
shorter bSSFP refocusing intervals (TRs). This work thus focuses on
characterizing the oxygen sensitivity of bSSFP in blood samples at high to
ultra-high fields by means of passband signal differences and intrinsic R2
estimation.
Introduction
Balanced steady-state free precession (bSSFP)
imaging is an interesting alternative for functional MRI1,2,3,4
since, in contrast to standard EPI-based techniques, it provides
distortion-free images with high SNR efficiency. Moreover, Monte Carlo
simulations have revealed that BOLD-related signal changes measured with bSSFP
are selective to microvessels5,6 similar to the spin echo and are thus
believed to be closer to the neuronal event than the gradient echo. A full
understanding of measured neurovascular fingerprints requires a separate
consideration of extravascular as well as intravascular contributions to the
BOLD effect. In case of rapidly refocused bSSFP, the intravascular contribution
is controlled by the transverse relaxation rate R2, which not only
strongly depends on the oxygenation level (Y) but also on the refocusing
interval (TR)7 due to diffusion narrowing effects around red blood
cells (RBCs) that contain paramagnetic deoxyhemoglobin. The bSSFP signal in
blood can thus not simply be estimated using literature R2 values,
which are commonly derived from spin-echo techniques. In this work, the oxygen sensitivity
of bSSFP in human blood samples at high to ultra-high fields (i.e., 3 T, 9.4 T,
and 14.1 T) depending on TR is assessed by 1) direct calculation of passband
bSSFP signal differences and 2) intrinsic R2 estimation from
phase-cycled bSSFP scans using MIRACLE8, 9.Methods
Fresh venous blood from healthy adult volunteers was
collected at the local blood bank. The oxygenation levels (Y) were adjusted by
bubbling air through the blood samples. The physiological hematocrit (Hct)
level of the blood was not altered. The Y and Hct values were assessed quantitatively
using a radiometer before and after the MR measurement session. For each
session of experiments (one session at each field strength), four blood samples
were prepared in 3 ml tubes; two with Y ≈ 70% (referred to as deoxygenated) and
two with Y ≈ 94% (referred to as oxygenated). Average Hct level of the measured
blood samples was ≈ 42%.
Imaging of the prepared blood samples was performed at 3 T,
9.4 T, and 14.1 T MR scanners. The blood probes were placed into a cylindrical
phantom along with a manganese-doped reference probe (0.2 mM MnCl2
in H2O). Temperature-controlled water was circulated through the
phantom to heat the blood samples to a temperature of 37°C. The temperature
inside the phantom container was permanently monitored during the MR
experiments with an MR-compatible fiber optic temperature sensor. The phantom
was rotated in-between subsequent scans by a dedicated robot-arm to avoid RBC
sedimentation.
3D phase-cycled bSSFP experiments were performed using 20 RF
phase increments ϕ in the range (0°, 360°) for different TRs up to 10 ms and a
flip angle of α = 10° as outlined in more detail below:
3 T: TR = [ 4, 5,
6, 8, 10] ms. Resolution: 1.2x1.2x2 mm3 (matrix size: 128x64x18).
9.4 T: TR = [ 3, 4, 5, 6, 8,
10] ms. Resolution: 1.2x1.2x2 mm3 (matrix size: 128x64x18).
14.1 T: TR = [2.5, 3, 4, 5, 6, 8, 10] ms. Resolution: 0.7x0.7x1 mm3
(matrix size: 85x85x26).
Gold standard R1 and R2 were calculated
based on single-slice inversion recovery with variable inversion times and
single-slice single-echo spin-echo with variable echo times, respectively. Mean
gold standard R1 and R2 values measured in deoxygenated
and oxygenated blood probes along with mean Y and Hct levels are summarized in
Table 1.
Based
on a Fourier transform of the measured complex bSSFP frequency profile, R2
values were estimated using MIRACLE9 with R1 fixed to the
respective measured gold standard values. Signal differences of the passband
bSSFP signal were derived after dividing the signal by the proportionality
factor S0 as: ΔS = (Sox - Sdeox)/S0,
where Sox and Sdeox are the mean passband bSSFP signal
levels of oxygenated and deoxygenated blood probes, respectively, and S0
is a scaling factor accounting for proton density and coil sensitivity derived
separately for each field strength based on a MnCl2-doped reference
probe.Results
Representative bSSFP frequency profiles acquired
by a series of scans with different RF phase increments for a range of TRs in
deoxygenated and oxygenated blood samples at 3 T, 9.4 T, and 14.1 T are shown
in Figure 1. Mean bSSFP passband signal differences ΔS
between oxygenated and deoxygenated probes are displayed in Figure 2 versus TR,
revealing the expected increase in ΔS
versus TR, in particular for higher field strength. The same trend is observed for
the quantitative R2 assessment in the deoxygenated probes with
MIRACLE, i.e., a diffusion-related increase in R2 for longer TRs,
especially prominent at 14.1 T, while the R2 in oxygenated blood is
highly TR-insensitive (cf. Fig. 3). Note that for TRs ≥ 8
ms at 14.1 T, the R2 fitting is becoming unreliable since the signal
in deoxygenated blood reaches the noise level as reflected by increased standard
deviations.Discussion and Conclusion
The increased oxygen sensitivity of bSSFP for longer
refocusing intervals was evident and assessed based on passband signal
differences as well as quantitatively using intrinsic R2 estimation
based on MIRACLE. In a further step, the obtained results can be combined with,
e.g., Monte Carlo simulations to determine the overall BOLD bSSFP effect,
including a differentiation between extravascular and intravascular
contributions.Acknowledgements
No acknowledgement found.References
-
Scheffler K, Seifritz E, Bilecen D, Venkatesan R,
Hennig J, Deimling M, Haacke EM. Detection of BOLD changes by means of a frequency‐sensitive trueFISP
technique: preliminary results. NMR Biomed 2001;14(7‐8):490-496.
- Dharmakumar R,
Hong J, Brittain JH, Plewes DB, Wright GA. Oxygen-sensitive contrast in blood
for steady-state free precession imaging. Magn Reson Med 2005;53(3):574-583.
- Miller KL.
FMRI using balanced steady-state free precession (SSFP). Neuroimage
2012;62(2):713-719.
- Scheffler K,
Ehses P. High-resolution mapping of neuronal activation with balanced SSFP at
9.4 Tesla. Magn Reson Med 2016;76(1):163-171.
- Baez-Yanez MG,
Ehses P, Mirkes C, Tsai PS, Kleinfeld D, Scheffler K. The impact of vessel
size, orientation and intravascular contribution on the neurovascular
fingerprint of BOLD bSSFP fMRI. Neuroimage 2017;163:13-23.
- Scheffler K,
Heule R, Báez-Yánez MG, Kardatzki B, Lohmann G. The BOLD sensitivity of rapid
steady‐state sequences. Magn Reson Med 2019;81(4):2526-2535.
- Gardener AG, Francis
ST, Prior M, Peters A, Gowland PA. Dependence of blood R2 relaxivity on CPMG
echo-spacing at 2.35 and 7 T. Magn Reson Med 2010;64(4):967-974.
- Pérez-Rodas M,
Schulz H, Pohmann R, Scheffler K, Heule R. R1 and R2 characterization of human
blood with phase-cycled balanced steady-state free precession (bSSFP). Proc.
Intl. Soc. Mag. Reson. Med. 27 (2019), p. 3766.
- Nguyen
D, Bieri O. Motion-insensitive rapid configuration relaxometry. Magn Reson Med
2017;78(2):518-526.