Julian A. J. Richter1,2, Tobias Wech1, Andreas M. Weng1, Manuel Stich1,3, Ning Jin4, Thorsten A. Bley1, and Herbert Köstler1
1Department of Diagnostic and Interventional Radiology, University Hospital Würzburg, Würzburg, Germany, 2Comprehensive Heart Failure Center Würzburg, Würzburg, Germany, 3Siemens Healthcare, Erlangen, Germany, 4Siemens Medical Solutions USA, Inc., Chicago, IL, United States
Synopsis
The
wave-CAIPI technique was applied to aortic 4D flow MRI. Three healthy
volunteers were examined and flow parameters as well as hemodynamic flow
patterns were derived from the measured data. The acquisitions were
retrospectively accelerated and compared to conventional Cartesian 2D-CAIPI
sampling. Using wave-CAIPI sampling, the deviations between flow parameters of
the 6-fold accelerated scans and the references (2-fold accelerated) could be reduced by up to 47%
compared to Cartesian sampling. As a consequence, the acquisition time of
aortic 4D flow acquisitions could be decreased to 3.5 minutes with higher
precision, concerning the calculated flow parameters and hemodynamic flow
patterns.
Purpose
Dynamic 3D
phase-contrast (4D flow) MRI can be used to evaluate flow parameters and to
visualize complex hemodynamic flow patterns. As velocity encodings must be
performed in all 3 spatial dimensions, a total of 4 images (3 velocity-encoded,
1 flow-compensated) must be acquired for each cardiac phase, leading to considerable
scan times. Parallel Imaging methods can be applied to accelerate the image
acquisition; however, a significant noise penalty arises for extensive
undersampling. We propose the application of the wave-CAIPI k-space trajectory1–4 for
optimized parallel Imaging of accelerated aortic 4D flow acquisitions to
achieve an acquisition time of less than 4 minutes.Methods
To assess the
performance of the wave-CAIPI technique in aortic 4D flow acquisitions, three healthy volunteers were examined using both wave-CAIPI and Cartesian 2D-CAIPRINHA (Controlled Aliasing in Parallel Imaging Results in Higher Acceleration), or 2D-CAIPI, sampling5,6 in
a 3D velocity-encoded gradient echo pulse sequence. All measurements were
performed on a 3T clinical MR scanner (MAGNETOM Prismafit, Siemens Healthcare, Erlangen, Germany). Images were
acquired in sagittal orientation with a k-space matrix of 128×76×28-30 and a
corresponding resolution of 2.5×3.5×3.5 mm. Further sequence parameters were:
flip angle 7°, TE 3.40 ms, TR 6.03 ms, temporal resolution 48.24 ms, readout bandwidth
500 Hz/pixel, FOV 320×260×98-105 mm, maximum encoded velocity VENC 200 cm/s. For the wave-CAIPI, 4
complete wave cycles with a maximum oscillation amplitude of 9 mT/m were
employed during readout. The average acquisition time was (10:40
± 01:18) minutes. The same acquisition window of 700 ms was
used for all volunteers and 14 cardiac phases were reconstructed using
prospective ECG-gating. To account for respiratory motion, the acquisition was
prospectively gated, using a navigator at the lung/liver interface with an
acceptance window of 8 mm. To mitigate residual motion artifacts, the ReCAR method7, a real-time
k-space ordering technique that samples the central k-space in end-expiration
and the peripheral k-space during inspiration, was applied. Images were
acquired with an undersampling factor of 2 in the phase-encoding
(anterior-posterior) direction. Undersampled data sets were reconstructed with
an iterative SENSE algorithm8. Coil
sensitivity maps were estimated from an additional low-resolution scan with
increased FOV in the partition-encoding (left-right) direction, using the Berkeley
Advanced Reconstruction Toolbox9. No slice
oversampling was employed to account for the imperfect slab-selective RF pulse
– instead, parallel Imaging (iterative SENSE8) was used to
remove the additional aliasing. The scanner-specific gradient system transfer
function was employed to correct for gradient distortions, in order to avoid
image artifacts10–13. The
acquired data was additionally undersampled retrospectively in partition
direction with an acceleration rate of 3, simulating a scan time reduction from
(10:40
± 01:18) minutes to (03:33 ± 0:26) minutes, and flow
parameters, as well as hemodynamic flow patterns, visualized via streamlines,
were calculated. To ensure comparability between the wave-CAIPI and Cartesian
2D-CAIPI results, the segmentation of the aorta, as well as the placement of
analysis planes was performed identically for both sampling strategies. Results
Figure 1 shows
the location of 8 analysis planes in the ascending and descending aorta of a
healthy volunteer, as well as the calculated net flow rates (NFR) within 3
exemplary planes for wave-CAIPI and Cartesian 2D-CAIPI sampling. The 2×3
accelerated Cartesian acquisition shows a larger deviation from the 2-fold
accelerated reference scan than the wave-CAIPI technique. Bland-Altman plots
comparing flow parameters of the 2×3 accelerated acquisitions with the
parameters of the 2-fold accelerated scans are presented in Figure 2 (net flow
rate) and Figure 3 (net average through-plane velocity, $$$v^\perp$$$). The
respective data points represent flow parameters calculated in 8 analysis planes
(as in Figure 1) and 14 cardiac phases. The wave-CAIPI scans of all three volunteers exhibited a lower standard deviation in both NFR and $$$v^\perp$$$. For the net
flow rate, the standard deviation between the two wave-CAIPI data sets was 32 –
42% lower than the corresponding standard deviation in the Cartesian 2D-CAIPI case. For the net average through-plane
velocity, the wave-CAIPI standard deviation was between 34 and 47% lower than
the Cartesian 2D-CAIPI standard deviation. Hemodynamic flow patterns are visualized
in Figure 4 by means of streamlines for wave-CAIPI and Cartesian 2D-CAIPI sampling
at peak systole for all three volunteers.Discussion & Conclusion
Due to the
wide voxel-spreading properties of the wave-CAIPI point spread function,
aliasing artifacts are distributed in all spatial dimensions, and variations in
coil sensitivities are used more efficiently in parallel Imaging
reconstructions1. As a
consequence, the noise enhancement induced by parallel Imaging is significantly
reduced, and accelerated phase difference images can be reconstructed with
higher precision. Therefore, the calculation of flow parameters using the wave-CAIPI
sampling technique is more stable in the case of undersampling. Differences in
streamline visualizations between the 2-fold and the 2×3-fold accelerated
acquisitions are more pronounced in the Cartesian 2D-CAIPI case. By employing
the wave-CAIPI sampling technique, the acquisition time could, in theory, be
reduced to (03:33 ± 0:26) minutes with minimal differences in flow
parameters and streamline flow visualizations.Funding
Comprehensive Heart Failure Center Würzburg, Grant BMBF 01EO1504Acknowledgements
No acknowledgement found.References
1. Bilgic B, Gagoski BA,
Cauley SF, et al. Wave-CAIPI for highly accelerated 3D imaging. Magn Reson
Med. 2015;73(6):2152-2162. doi:10.1002/mrm.25347
2. Cauley
SF, Setsompop K, Bilgic B, Bhat H, Gagoski B, Wald LL. Autocalibrated
wave-CAIPI reconstruction; Joint optimization of k-space trajectory and
parallel imaging reconstruction. Magn Reson Med. 2017;78(3):1093-1099.
doi:10.1002/mrm.26499
3. Polak
D, Setsompop K, Cauley SF, et al. Wave-CAIPI for highly accelerated MP-RAGE
imaging. Magn Reson Med. 2018;79(1):401-406. doi:10.1002/mrm.26649
4. Polak
D, Cauley S, Huang SY, et al. Highly-accelerated volumetric brain examination
using optimized wave-CAIPI encoding. J Magn Reson Imaging JMRI. February
2019. doi:10.1002/jmri.26678
5. Breuer
FA, Blaimer M, Mueller MF, et al. Controlled aliasing in volumetric parallel
imaging (2D CAIPIRINHA). Magn Reson Med. 2006;55(3):549-556.
doi:10.1002/mrm.20787
6. Breuer FA, Blaimer M, Heidemann
RM, Mueller MF, Griswold MA, Jakob PM. Controlled
aliasing in parallel imaging results in higher acceleration (CAIPIRINHA) for
multi-slice imaging. Magn Reson Med. 2005;53(3):684-691.
doi:10.1002/mrm.20401
7. Pruessmann
KP, Weiger M, Börnert P, Boesiger P. Advances in sensitivity encoding with
arbitrary k-space trajectories. Magn Reson Med. 2001;46(4):638-651. doi:10.1002/mrm.1241
8. Ma
LE, Markl M, Chow K, et al. Aortic 4D flow MRI in 2 minutes using compressed
sensing, respiratory controlled adaptive k-space reordering, and inline
reconstruction. Magn Reson Med. 2019;81(6):3675-3690. doi:10.1002/mrm.27684
9. Martin
Uecker, Frank Ong, Jonathan I Tamir, et al. Berkeley Advanced Reconstruction
Toolbox. In: Proc. Intl. Soc. Mag. Reson. Med. Vol 23. ; 2015.
10. Stich
M, Wech T, Slawig A, et al. Gradient waveform pre-emphasis based on the
gradient system transfer function. Magn Reson Med. 2018;80(4):1521-1532. doi:10.1002/mrm.27147
11. Vannesjo SJ, Haeberlin M, Kasper
L, et al. Gradient system characterization by impulse
response measurements with a dynamic field camera. Magn Reson Med. 2013;69(2):583-593.
doi:10.1002/mrm.24263
12. Vannesjo SJ, Graedel NN, Kasper
L, et al. Image reconstruction using a gradient impulse
response model for trajectory prediction. Magn Reson Med.
2016;76(1):45-58. doi:10.1002/mrm.25841
13. Campbell-Washburn
AE, Xue H, Lederman RJ, Faranesh AZ, Hansen MS. Real-time distortion correction
of spiral and echo planar images using the gradient system impulse response
function. Magn
Reson Med.
2016;75(6):2278-2285. doi:10.1002/mrm.25788