Wissam AlGhuraibawi1, Kevin Godines1, Mark Velasquez1, Sinyeob Ahn2, Wolfgang Rehwald3, and Moriel Vandsburger1
1Bioengineering, University of California Berkeley, Berkeley, CA, United States, 2Siemens Healthineers, Concord, CA, United States, 3Siemens Healthineers, Durham, NC, United States
Synopsis
CEST-MRI is an emerging molecular imaging method for non-invasive
assessment of cardiomyocyte metabolites. In cardiac CEST-MRI, spatial B1
inhomogeneity across the myocardium significantly reduces the accuracy of
measured CEST contrasts. Deviation from the prescribed B1 leads to altered
creatine CEST contrast due to both reduced labeling efficiency and heightened
magnetization transfer and direct water direct saturation across the heart. The
final impact is measurement of falsely and substantially reduced creatine CEST
contrast in the healthy heart.
Introduction
Chemical exchange saturation transfer (CEST) MRI is an emerging
molecular imaging method for non-invasive assessment of cardiomyocyte
metabolites including creatine, and has been used to assess changes in the
setting of obesity or myocardial infarction. 1–3 While correction for spatial B0 inhomogeneity
is routinely performed in cardiac CEST, the impact of spatially varying B1
fields, which can vary 40% across the left ventricle (LV), 4 is generally overlooked. Importantly, the CEST
contrast generated for any metabolite, magnitudes of magnetization transfer
(MT) and direct water saturation are a non-linear function of saturation power
used during CEST preparation. Subsequently, measured CEST contrasts reflect a
combination of underlying substrate concentrations and spatially specific
saturation power. In this study, we test the impact of spatial B1
inhomogeneity across the healthy heart upon assessment of creatine CEST and MT
contrasts.Methods
12 healthy volunteers with no history of cardiovascular disease,
diabetes, or smoking aged 20–37 were recruited. Cardiac MRI was performed on a
3T Siemens Trio scanner (Siemens Medical Systems, Erlangen, Germany). B0
and B1 field maps were acquired in one midventricular short-axis slice.
For CEST imaging, saturation offsets ranged from -10ppm to +10ppm using a 1311ms
train of gaussian saturation pulses (pulse duration = 36ms, duty cycle = 0.63,
and peak B1 = 1.2µT), followed by a segmented echo planar imaging
gradient echo readout (FOV = 300 x 253mm, spatial resolution = 1.56 x 1.56mm,
slice thickness = 8mm, TR = 4.7ms, TE = 2.59ms, and FA = 25°) timed to end-diastole.
Z-spectra were generated in six anatomical segments of the LV and fit with a multi-pool Lorentzian
algorithm 5 to quantify creatine and APT contrasts, nuclear Overhauser effect, MT
and direct saturation of water (Figure 1). The goodness of Lorentzian fit was
assessed using normalized mean square error (NMSE) of the raw data to the
generated Lorentzian fit.6 Segments with noticeable susceptibility artifacts and an NMSE value
< 0.998 were excluded from the analysis. In-vitro studies were performed in
two phantoms with sub-phantoms containing creatine concentrations of 0 to 30mM
dissolved in (a) phosphate buffer solution (PBS) or (b) 10% cross-linked bovine
serum albumin (BSA).7 For each participant, ‘true’ creatine concentrations were calculated
from the application of septal CEST contrasts to the regression from 10% BSA + creatine
phantoms. For each individual, segmental creatine concentrations were
calculated from the same regression, and the accuracy determined based on the
percent error in each segment compared to each individual’s ‘true’ value.Results
Representative images (Figure 2) and Z-spectra (Figure 1) highlight the
impact of B1 inhomogeneity upon CEST/MT contrasts. The average
magnitude of spatial B1 variation across the six LV segments was
22.03 ± 8.2%, with consistently higher peak B1 observed in the
lateral wall segments (Figure 3). Spatial B1 inhomogeneity resulted
in significantly elevated MT and direct water saturation, and significantly
reduced creatine CEST contrast in lateral wall segments compared to septal
segments (Figure 3). The
impact of B1 variation on CEST contrast generated by a fixed
creatine concentration can be estimated using a two-pool stimulation (Figure 4). However, measured creatine CEST contrasts are significantly lower in the presence
of MT (Figure 4). Significant correlations existed between the magnitude of B1
deviation and all measured CEST contrasts (Figure 5). Similarly, the accuracy
of in vivo quantification of creatine content was significantly and inversely
correlated with the magnitude of B1 deviation (Figure 5).Discussion
Generation of creatine CEST contrast requires that the correct peak B1
saturation power be spatially consistent across the heart. Spatial B1
inhomogeneity generates inconsistent creatine CEST labeling and further
complicates the detection of creatine via increased MT, direct saturation, and
spillover from APT labeling. Two-pool simulations demonstrate that a 20% shift
in B1 from prescribed amplitude will reduce CEST contrast in the
presence of consistent underlying creatine concentration. However, phantom and
in vivo data reveal that the combined impact of increased MT and reduced
labeling efficiency further abrogate creatine CEST contrast to a level
consistent with infarcted tissue despite normal underlying concentrations in
the hearts of healthy individuals.Conclusion
These results in this study implies that cardiac CEST-MRI requires both
more advanced design of saturation pulses to maximize spatial uniformity, and
analysis methods that account for B1 inhomogeneity in quantification
of underlying metabolite contrasts.Acknowledgements
No acknowledgement found.References
1.
AlGhuraibawi W, Stromp T, Holtkamp
R, et al. CEST MRI reveals a correlation between visceral fat mass and reduced
myocardial creatine in obese individuals despite preserved ventricular
structure and function. NMR Biomed. 2019;32(7):1-10.
2.
Haris M, Singh A, Cai K, et al. A
technique for in vivo mapping of myocardial creatine kinase metabolism. Nat
Med. 2014;20(2):209-214.
3.
Zhou Z, Nguyen C, Chen Y, et al.
Optimized CEST cardiovascular magnetic resonance for assessment of metabolic
activity in the heart. J Cardiovasc Magn Reson. 2017;19(1):95.
4.
Sung K, Nayak KS. Measurement and
characterization of RF nonuniformity over the heart at 3T using body coil
transmission. J Magn Reson Imaging. 2008;27(3):643-648.
5.
Deshmane A, Zaiss M, Lindig T, et
al. 3D gradient echo snapshot CEST MRI with low power saturation for human
studies at 3T. Magn Reson Med. 2019;81(4):2412-2423.
6.
Chen L, Barker PB, Weiss RG, van
Zijl PCM, Xu J. Creatine and phosphocreatine mapping of mouse skeletal muscle
by a polynomial and Lorentzian line-shape fitting CEST method. Magn Reson
Med. 2019;81(1):69-78.
7. Xu J, Yadav NN, Bar-Shir A, et al.
Variable delay multi-pulse train for fast chemical exchange saturation transfer
and relayed-nuclear overhauser enhancement MRI. Magn Reson Med.
2014;71(5):1798-1812.