Thomas Puiseux1,2, Anou Sewonu1,3, Ramiro Moreno1,3,4, Simon Mendez2, and Franck Nicoud2
1SPIN UP, Toulouse, France, 2IMAG, Univ. Montpellier, CNRS, Montpellier, France, 3I2MC, INSERM U1048, Toulouse, France, 4ALARA Expertise, Strasbourg, France
Synopsis
The present study proposes a
novel approach to efficiently simulate 4D Flow MRI acquisitions in realistic
complex flow conditions. Navier-Stokes and Bloch equations are simultaneously
solved with Eulerian-Lagrangian coupling. A semi-analytic solution for the
Bloch equation as well as a periodic particle re-injection strategy are
implemented to reduce the computational cost. The Bloch solver and the velocity
reconstruction pipeline were first validated in a steady flow configuration. The
coupled 4D Flow MRI simulation procedure was validated in a complex pulsatile
flow phantom cardiovascular-typical experiment. Besides, we compared simulated
MR velocity data with experimental 4D Flow MRI measurements.
Introduction
Time-resolved 3D phase-contrast Magnetic Resonance Imaging (or 4D Flow MRI)
is a promising tool for in-vivo non-invasive blood flow quantification. As it
relies on complex physical principles and multimodal acquisition, different
types of velocity measurement errors arise, which can impair diagnosis. Numerical
simulation of the MRI acquisition process enables to reconstruct synthetic
images free of experimental errors in order to identify their origins. Although
many simulation frameworks have been developed for static tissues imaging1,2,
flow MRI modeling remains challenging mainly because of the need to account for
the spin dynamics, which results in noticeable increase in computational load.
The objective of this work is to present a 4D flow MRI simulation workflow of
a realistic pulsatile flow typical of the cardiovascular system.Methods
We designed a well-controlled experiment delivering pulsatile
blood-mimicking fluid flow within a rigid phantom and carried out PC-MRI experiments.
An image-based CFD analysis was performed from the PC-MRI data prescribed as
inlet velocity profile3. The Bloch equations were solved at each
iteration and the spin positions updated “on the fly” from CFD velocity hence
eliminating the need to store particles trajectories. The experimental setup is
depicted in figure 1.
Simulation of flow MR imaging (illustrated in figure
2): Np particles are injected with a
uniform random distribution into each cell of a fixed unstructured numerical
mesh. At each iteration, the Navier-Stokes equations are solved on the
unstructured mesh using YALES2BIO solver4. The CFD velocity was
interpolated on each particle to update their position. Each particle is
associated with magnetic properties (T1, T2, M0), an isochromat volume, and an injection magnetization Minj = (0, 0, Mzss) where Mzss is the steady
state magnetization. At each
spoiling event, particles are suppressed and Np particles are reinjected at
the same location as initially. This key step of the methodology allows to keep
the particles distribution homogeneous and avoid spurious-signal areas. An
effective magnetic field imposed by the (idealized) MR sequence designed with JEMRIS2
was prescribed to each particle at each instant. From a multi-criterion
time-stepping approach, the magnetization vector was either advanced with a
full numerical integration (4th-order Runge-Kutta) of the Bloch equations
whenever the Radio-Frequency field (RF) is on, or by the analytical solution
(given that the gradient waveform can be analytically described.
MR parameters: A RF-spoiled time-resolved 3D
Gradient echo sequence with three velocity encoding gradients was simulated. A
2 mm3 voxel size was set with a flip angle =15°, TR=6 ms, ∆T = 58
ms, and a VENCx,y,z=60 cm/s.
Reconstruction: The resulting MR signal was Fourier-transformed
in order to reconstruct synthetic MR images and velocity field. As the
simulated MR images are stripped of experimental errors, the sources of
observed discrepancies can be better highlighted.
Validation: The simulated MR velocity field (uSMRI) was compared with the phase-averaged
CFD velocity fields (uCFD) to validate the CFD coupling. Prior to the comparison, CFD velocity field
were downsampled to the MRI spatial resolution. In addition, uSMRI was
compared to 4D Flow MR measurements (uMRI) to identify the major
discrepancy sources.Results
Figure 3 shows excellent agreement between imposed uCFD and uSMRI regardless of the phase (peak and
mean correlation coefficients: r2 = 0.978 and 0.966). While the MRI simulation
suppresses many sources of discrepancies, some still remain such as
misregistration artifacts, intravoxel phase dispersion, Gibbs artifacts or
velocity fluctuations effects.
To identify the largest velocity error contribution, the highest error
patterns were compared to the highest CFD acceleration patterns in figure 4.
The high visual correlation confirms that acceleration-induced artifacts
contribute for the main errors.
Note also that the sparse particles distribution near the inlet boundary induces high velocity error.
The velocity error produced by MRI simulation with respect to CFD was qualitatively compared
to the error raised between MR measurements (uMRI) and CFD in figure 5. Generally speaking, a
good agreement and reproducible error patterns are generally noticed with a net
error decrease in uSMRI. It is suggested that the highest error site
located at the collateral elbow (and well reproduced by the simulation) is the
result of both misregistration artifacts and intravoxel dispersion.Discussion
In this study, we presented a workflow for simulating realistic 4D flow MR
acquisitions. Several coupled CFD-MRI simulations of a well-controlled flow
phantom experiment were performed and compared to the phase-averaged CFD
velocity. Reported at peak systole, the largest velocity discrepancies were
mainly located in the regions of high flow acceleration. It was suggested that
these errors appear since the particle acceleration is not accounted for in the
phase-velocity relationship. Acceleration-sensitive acquisitions could be carried
out to account for the flow high-order motion5 or pre-processing
corrections could be applied to the velocity field6. The experimental
MRI was also compared to the simulated MRI and similar error patterns were
reported as compared to the CFD, were small rotating structures and high
convective acceleration induce image artifacts. However, some unquantified
errors still arise because of the differences between experimental and
simulated MR sequences. A crucial step to compare MRI simulation to MRI measurements is to simulate the same MR sequence as the one applied
experimentally. This work constitutes a preliminary phase to further exploit
the MRI simulation capabilities.Acknowledgements
No acknowledgement found.References
1. Bittoun et al, MRI, 1984.
2. Stoecker et al, MRM, 2010.
3. Puiseux et al., NMR Biomed., 2019.
4. Mendez
et al, IFMBE, 2015.
5. Bittoun et
al, MRM, 2000.
6. Thunberg et al, MRM, 2000.