Brendan Lee Eck1,2, Jeehun Kim2, Mingrui Yang2, and Xiaojuan Li2
1Cardiovascular and Metabolic Sciences, Cleveland Clinic, Cleveland, OH, United States, 2Biomedical Engineering, Cleveland Clinic, Cleveland, OH, United States
Synopsis
Quantitative T1ρ
imaging has been studied for evaluating changes in tissue composition, in
particular for detecting early cartilage degeneration in osteoarthritis.
Magnetic resonance fingerprinting (MRF) provides a framework for rapid, robust
acquisition of quantitative tissue property maps. Simulation experiments using
spin-lock prepared MRF with different pulse schedules were conducted to
demonstrate the feasibility of quantification of T1ρ
in addition to T1 and T2. All tested sequences were
sensitive to T1ρ and produced tissue property
maps with major structures of interest. Differing accuracy and precision between
sequences suggests opportunities for optimizing MRF for simultaneous T1,
T2, and T1ρ quantification.
Introduction
Quantitative MRI can provide complementary
information to morphologic imaging for the evaluation of degenerative
musculoskeletal disease as well as injury. T1ρ,
the spin-lattice relaxation in the rotation frame, and T2 relaxation
times can probe early changes in proteoglycan-collagen matrix during cartilage
degeneration prior to morphological or clinical changes in osteoarthritis1,2. However, T1ρ
and T2 mapping techniques are in general time consuming and have
shown variations using different MR systems3,4. Magnetic resonance
fingerprinting (MRF) has been developed in recent years and has been shown to
be an efficient quantitative imaging technique to provide T1 and T2
measures (and other quantitative measures such as perfusion and diffusion) that
are reproducible across scanners5. However, no studies have yet
quantified T1ρ using MRF. The addition of T1ρ
quantification to the MRF framework would enable the quantification of multiple
tissue properties in a single scan, thus simplifying the scan and image analysis
workflows.Methods
Simulated MRF scans at 3T were performed using a digital
knee phantom with a sagittal slice orientation. The MRF
sequences evaluated in this work were based on the MRF fast imaging with
steady-state procession (FISP) sequence6. Four sequence variants were
used (Table 1) to evaluate the effects of changing the flip angle (FA) pattern,
TR pattern, and spin-lock preparation pulse schedule. A variable density spiral
trajectory was used for all acquisitions. Pattern matching reconstruction was
used with dictionary values of T1=100:50:2000 (ms), T2=2,5,10:10:200
(ms), and T1ρ=2,5,10:10:200 (ms). T1, T2, and T1ρ (at spin-lock frequency of 500Hz) values at 3T as reported in the literature were used as ground truth for the simulated tissues (Table 2). MRF sequences were compared in terms of relaxation time quantification using normalized root-mean-square error (NRMSE), average relaxation times within each tissue type, and standard deviation of relaxation times within each tissue type.Results
All tested sequences
were sensitive to T1ρ
values and produced tissue property maps with major structures of interest
(Figure 2). Example signal evolutions for the sequences are shown in Figure 3, indicating
that shorter TR and more frequent spin-lock preparation pulses can enhance
sensitivity to T1ρ. Sequence
D had lowest NRMSE for T1ρ
estimation (Table 2), suggesting greatest overall quality. All sequences were
unable to estimate the very short T1ρ in tendon or precisely estimate the long T1ρ in bone marrow. The optimal sequence in terms
of accuracy and precision depended on the tissue type and relaxation time of
interest. Changes in TR pattern, FA pattern, and spin-lock preparation pulse
schedule was observed to impact precision and accuracy of relaxation time
estimates to different degrees.Discussion
Spin-lock prepared MRF can enable the
quantification of T1ρ in addition to T1
and T2. While both T2 and T1ρ
cause exponential signal decay, the simulations showed that both can be quantified
with the use of spin-lock preparation in MRF sequences. All tested sequences had differing advantages for relaxation time quantification depending on tissue type and relaxation property. The MRF Sequence D with
short, constant TR and spin-lock preparation every 100 time frames provided
best overall T1ρ quantification. The flexibility in the MRF framework may present an opportunity for SAR reduction as most T1ρ
mapping techniques in the literature use spin-lock preparation every <100
TRs7,8. These findings demonstrate the feasibility of T1ρ
quantification with MRF. Future refinements to the technique are anticipated
via sequence optimization and improved image reconstruction. Evaluation of T1ρ
dispersion via MRF may also be explored within this framework by incorporating
different spin-lock frequencies.Conclusion
Magnetic resonance fingerprinting with spin-lock
preparation pulses can be used to quantify T1ρ
relaxation time in addition to T1 and T2. Further
optimization of MRF sequences for simultaneous T1ρ,
T1, and T2 quantification are warranted.Acknowledgements
The
authors would like to thank Dr. Mark Griswold and Dr. Dan Ma for helpful
discussions.References
1. Atkinson,
H. F. et al. MRI T2 and T1ρ relaxation in patients at risk for knee
osteoarthritis: a systematic review and meta-analysis. BMC Musculoskelet
Disord 20, 182 (2019).
2. MacKay,
J. W. et al. Systematic review and meta-analysis of the reliability and
discriminative validity of cartilage compositional MRI in knee osteoarthritis. Osteoarthr.
Cartil. 26, 1140–1152 (2018).
3. Balamoody,
S. et al. Magnetic resonance transverse relaxation time T2 of knee
cartilage in osteoarthritis at 3-T: a cross-sectional multicentre, multivendor
reproducibility study. Skeletal Radiol. 42, 511–520 (2013).
4. Kim, J. et al. Multi-vendor
Multi-site T1ρ and T2 Quantification of Knee Cartilage. ACR Meeting
Abstracts. (2019)
5. Buonincontri,
G. et al. Multi-site repeatability and reproducibility of MR
fingerprinting of the healthy brain at 1.5 and 3.0 T. NeuroImage 195,
362–372 (2019).
6. Jiang,
Y., Ma, D., Seiberlich, N., Gulani, V. & Griswold, M. A. MR fingerprinting
using fast imaging with steady state precession (FISP) with spiral readout. Magnetic
resonance in medicine 74, 1621–1631 (2015).
7. T1ρ‐prepared
balanced gradient echo for rapid 3D T1ρ MRI - Witschey - 2008 - Journal of
Magnetic Resonance Imaging - Wiley Online Library.
8. Li,
X., Han, E. T., Busse, R. F. & Majumdar, S. In vivo T1ρ mapping in
cartilage using 3D magnetization‐prepared angle‐modulated partitioned k‐space
spoiled gradient echo snapshots (3D MAPSS). Magnetic Resonance in Medicine:
An Official Journal of the International Society for Magnetic Resonance in
Medicine 59, 298–307 (2008).
9. Hamilton,
J. I. et al. MR fingerprinting for rapid quantification of myocardial
T1, T2, and proton spin density. Magnetic resonance in medicine 77,
1446–1458 (2017).
10. Gold,
G. E. et al. Musculoskeletal MRI at 3.0 T: Relaxation Times and Image
Contrast. American Journal of Roentgenology 183, 343–351 (2004).
11. Ma,
Y.-J. et al. 3D adiabatic T1ρ prepared ultrashort echo time cones
sequence for whole knee imaging. Magnetic Resonance in Medicine 80,
1429–1439 (2018).
12. Krämer,
M. et al. T1 and T2* mapping of the human quadriceps and patellar
tendons using ultra-short echo-time (UTE) imaging and bivariate relaxation
parameter-based volumetric visualization. Magnetic Resonance Imaging 63,
29–36 (2019).
13. Carl, M., Chiang, J., Han, E., Bydder,
G. & King, K. Bloch simulations of UTE, WASPI and SWIFT for imaging short
T2 tissues. in 884 (2010).