Ruvini Navaratna1,2, Timothy J Colgan1, Ruiyang Zhao1,2, Houchun Harry Hu3, Mark Bydder4, Takeshi Yokoo5, Mustafa R Bashir6,7,8, Michael S Middleton9, Suraj D Serai10, Daria Malyarenko11, Thomas Chenevert11, Mark Smith3, Walter Henderson9, Gavin Hamilton9, Yunhong Shu12, Claude B Sirlin9, Jean A Tkach13, Andrew T Trout13, Jean H Brittain14, Diego Hernando1,2, and Scott B Reeder1,2,15,16,17
1Radiology, University of Wisconsin-Madison, Madison, WI, United States, 2Medical Physics, University of Wisconsin-Madison, Madison, WI, United States, 3Radiology, Nationwide Children's Hospital, Columbus, OH, United States, 4Radiological Sciences, University of California - Los Angeles, Los Angeles, CA, United States, 5Radiology, University of Texas Southwestern Medical Center, Dallas, TX, United States, 6Radiology, Duke University Medical Center, Durham, NC, United States, 7Division of Gastroenterology, Duke University Medical Center, Durham, NC, United States, 8Center for Advanced Magnetic Resonance Development, Duke University Medical Center, Durham, NC, United States, 9Radiology, University of California - San Diego, San Diego, CA, United States, 10Radiology, Children's Hospital of Philadelphia, Philadelphia, PA, United States, 11Radiology, University of Michigan, Ann Arbor, MI, United States, 12Radiology, Mayo Clinic, Rochester, MN, United States, 13Radiology, Cincinnati Children's Hospital Medical Center, Cincinnati, OH, United States, 14Calimetrix, LLC, Madison, WI, United States, 15Biomedical Engineering, University of Wisconsin-Madison, Madison, WI, United States, 16Medicine, University of Wisconsin-Madison, Madison, WI, United States, 17Emergency Medicine, University of Wisconsin-Madison, Madison, WI, United States
Synopsis
Chemical
shift-encoded MRI (CSE-MRI) is well-established to quantify proton density fat-fraction
(PDFF) as a quantitative biomarker of hepatic steatosis.1 However, temperature
is known to affect the accuracy and precision of PDFF quantification.2 In
this study, we aim to characterize the effects of temperature on PDFF
quantification using computer simulations, temperature-controlled phantom
experiments, and a multi-center phantom study. Further, we present a novel
method to minimize temperature-related fat quantification bias for magnitude-based
CSE-MRI methods.
Introduction
CSE-MRI is well
established as an accurate and precise method to quantify proton density fat-fraction
(PDFF) as a quantitative biomarker of hepatic steatosis.1 PDFF is reconstructed using a multi-peak
spectral model:3 $$S(TE_n)=[W+F{c_N}]e^{i{2 \pi}{f_B}TE_n}e^{-R_2^*TE_n}$$
where $$$W$$$ and $$$F$$$ are the complex signal amplitudes of water and
fat, respectively, $$$c_N=\sum_{p=1}^P {\alpha_p}e^{i{2\pi}{f_{F,p}}TE_n}$$$ is the multi-fat peak model with $$$P$$$ peaks of amplitude $$$\alpha_p$$$ and frequencies $$$f_{F,p}$$$ relative to the water resonance, $$$f_B$$$ is the bulk field shift, and $$$TE_n$$$ is the $$$n$$$th echo time. The PDFF can be
calculated as: $$PDFF=100\times\frac{F}{W+F}$$
Generally, the
spectral model of fat assumes an in vivo temperature of 37.0°C with the relative water
and fat methylene peak separated by 3.4ppm.2
However, the proton resonance frequency of water will change with
temperature.4,5 The temperature-corrected signal can be modeled: $$S(TE_n)=[We^{i{2\pi}{f_{T}}TE_n}+F{c_N}]e^{i{2\pi}{f_B}TE_n}e^{-R_2^*TE_n}$$
where $$$f_T$$$ is the frequency offset of water due to
temperature, which shifts by −0.01 ppm/°C.
In ex vivo tissue or phantom
studies, where the object is not at body temperature (eg. 21°C), PDFF estimation with a temperature-uncorrected
model may lead to meaningful bias.2
Here, a novel method to
minimize global temperature-related bias in PDFF estimation is proposed and
evaluated using simulations, temperature-controlled experiments, and as part of
a multi-center round-robin phantom study.Methods
Simulations:
Noiseless
multi-echo fat-water signals were simulated at 1.5T and 3T with $$$R_2^*=0 s^{-1}$$$ and $$$f_B = 0 Hz$$$ at temperatures from 0°C to 37.0°C for
fat-fraction values of 0-50%. Six echoes were assumed with echo times typical
for magnitude-based CSE-MRI PDFF estimation, with an initial TE of 2.3ms
(1.15ms) and $$$\Delta$$$TE of 2.3ms (1.15ms) at 1.5T (3T). PDFF estimation was performed using magnitude
nonlinear least-squares fitting, $$$|S(TE_n)|$$$ from Eq. 1, using a
nonlinear least-squares fit to estimate PDFF, without temperature.
Phantom Setup:
A fat phantom
previously described as part of a multi-center Quantitative Imaging Biomarkers
Alliance (QIBA) study,6 was used for a set of temperature-controlled
experiments. Twelve vials with PDFF ranging from 0 to 100% (Figure 1) were
fixed in a spherical phantom housing (Model 300, Calimetrix, Madison, WI).
Temperature-Controlled
Phantom Experiments:
The phantom was imaged
at 1.5T and 3T (1.5T 450w, 3.0T Signa Premier, GE Healthcare, Waukesha, WI)
using standard head coils (1.5T - 8 Channel Hi-Res Brain Coil, 3.0T - 48 channel
head coil) at three temperatures, nominally 2°C,
21°C, and 37.0°C.
Surface temperature of the phantom was measured using an infrared temperature
probe (Lasergrip 1080, Etekcity, Anaheim, CA). Data were acquired using a 2D
multi-echo acquisition to provide magnitude only data. Acquisition parameters
are listed in Table 1.
Multi-Center Phantom
Experiments:
Multi-center
validation of the temperature-corrected model was performed using data acquired
from the QIBA PDFF study,6 which used the same phantom. 68 magnitude datasets,
acquired on 24 unique MR systems, were acquired at 9 different sites across the
U.S., using 3 different vendors (GE, Siemens, Philips), at both 1.5T and 3T. Acquisition
parameters are listed in Table 1. All acquisitions were performed at room
temperature.
Reconstruction
and Analysis:
Magnitude fitting
was performed using Eq. 3, assuming temperatures from 0°C to 37.0°C
to generate PDFF maps. For each reconstruction, the slope of apparent vs true
PDFF was recorded. Slope difference,
ie: true slope – apparent slope, or 1 – apparent slope, was recorded as an
error metric. Voxel-wise fit error, or the sum of squared residuals, $$$\sum_{TE} (S_{est}(TE) - S_{meas}(TE))^2$$$ was also measured, and
the average over all ROIs was used as a second error metric.
To minimize
temperature-related PDFF estimation bias, we determine the minimum of these
global fit errors over 0-37.0°C. The temperature at which fit error is
minimized is used for the temperature-corrected signal to minimize PDFF bias,
even if the true temperature of the phantom is unknown.Results
Simulations:
Simulations showing
PDFF bias as a function of temperature are plotted in Figure 2. PDFF bias
increases as the temperature deviates further from 37.0°C and at higher PDFF.
Temperature-Controlled
Phantom Experiments:
For the temperature-controlled
experiment, the fit error for each temperature-controlled scan is shown in
Figure 3. The fit error was minimized close to the true temperature of the
phantom vials.
Multi-Center Phantom
Experiments:
An example of
apparent vs. true PDFF is shown in Figure 4a. This demonstrates the error
metric of slope difference.
A summary of the
PDFF bias, shown as the deviation in true slope (1.0) and apparent slope,
before and after correction for temperature, is shown for all 68 acquisitions
performed in the QIBA round-robin study in Figure 4b. The PDFF bias is greatly reduced
after correction for temperature using the proposed method (p = 0.001).Discussion and Conclusions
In this work we
investigated the effects of temperature on the accuracy of magnitude based
CSE-MRI to quantify PDFF in phantoms. Our simulations demonstrate that PDFF bias
is highly dependent on temperature. Further, we have proposed a strategy that
minimizes the effects of temperature by identifying the temperature that
minimizes fitting error. PDFF bias from a large multi-site round-robin phantom
study was reduced by correcting for the temperature through a proposed
technique that minimizes global fitting error, without any a priori information
on the phantom temperature.Acknowledgements
The authors wish
to acknowledge support from the NIH (R41-EB025729, R01 DK088925, K24 DK102595),
as well as GE Healthcare and Bracco Diagnostics
who provides research support to the University of Wisconsin, and from
Calimetrix for providing use of the phantom used in this study. Finally, Dr.
Reeder is a Romnes Faculty Fellow, and has received an award provided by the
University of Wisconsin-Madison Office of the Vice Chancellor for Research and
Graduate Education with funding from the Wisconsin Alumni Research Foundation.References
- Yokoo
et al. Linearity, Bias, and Precision of Hepatic Proton Density Fat Fraction
Measurements by Using MR Imaging: A Meta-Analysis. Radiology; 2017.
- Hernando
et al. On the Confounding Effect of Temperature on Chemical Shift Encoded Fat
Quantification. Magn Reson Med; 2014.
- Hernando
et al. Chemical shift-based water/fat separation: a comparison of signal models.
Magn Reson Med; 2010.
- Kuroda
et al. Temperature mapping using the water proton chemical shift: a chemical
shift selective phase mapping method. Magn Reson Med 1997; 38:845–851
- Sprinkhuizen
et al. Temperature‐induced tissue susceptibility changes lead to significant
temperature errors in PRFS‐based MR thermometry during thermal interventions.
Magn Reson Med 2010;64:1360–1372.
- Hu et
al. Multi-Site, Multi-Vendor, and Multi-Platform Reproducibility and Accuracy
of Quantitative Proton-Density Fat Fraction (PDFF) at 1.5 and 3 Tesla with a
Standardized Spherical Phantom: Preliminary Results from a Study by the RSNA
QIBA PDFF Committee. International Society for Magnetic Resonance in Medicine
Annual Meeting. Montreal, Canada. May 16, 2019. Program Number: 1023.
- Reeder et al. Quantitative Assessment of
Liver Fat with Magnetic Resonance Imaging and Spectroscopy. J Magn Reson Imaging. 2011;34(4):729–749.