Isabelle Heukensfeldt Jansen1, Luca Marinelli1, Ek Tsoon Tan2, Robert Y Shih3,4, J Kevin DeMarco3,4, J Kent Werner3,4, Vincent B Ho3,4, and Thomas Foo1
1GE Global Research Center, Niskayuna, NY, United States, 2Hospital for Special Surgery, New York, NY, United States, 3Uniformed Services University of the Health Sciences, Bethesda, MD, United States, 4Walter Reed National Military Medical Center, Bethesda, MD, United States
Synopsis
We demonstrate a method for simultaneous imaging
of diffusion and slow motion in vivo. We use both the magnitude and phase
information from image data to reconstruct coherent and incoherent motion (flow
and diffusion). We modified a PGSE diffusion imaging sequence so that b-value
and encoded velocity can be set independently. We imaged healthy volunteers with a 2-shell sequence with bmax=2000 sec/mm2 and
venc=0.24 mm/s at multiple phases during the cardiac cycle using peripheral
gating. Results show a distinct periodic motion around the ventricles with RMS
speed 0.065 mm/s, moving laterally during systole
and medially during diastole
Introduction
Recent interest in understanding the waste removal processes
in the brain has highlighted the need for in vivo imaging of
sub-millimeter/second flow1-4. Diffusion imaging utilizes
second-order water flow to characterize white matter microstructures and
tractography. Conventional diffusion image processing relies on magnitude
information; by using the remaining phase information, it becomes possible to
simultaneously extract coherent motion profiles for the tissue, which can be
related to CSF flow.
In this work, we
present an analysis for the selection of timing parameters of a diffusion
imaging pulse sequence to optimize the simultaneous collection of both
magnitude and phase data. We also discuss the subsequent reconstruction for
coherent motion that can run in parallel to existing diffusion imaging
reconstruction.Methods
Standard diffusion imaging uses a pair of trapezoidal pulses
to encode a given b-value with minimum TE. We modified a single-spin echo
diffusion imaging sequence to create the SCIMI (Simultaneous Coherent-Incoherent
Motion Imaging) sequence, which allows setting a given b-value and an encoded
velocity independently by appropriately tuning timing parameters. For a given b-value,
minimum TE is achieved by varying $$$\delta$$$ while keeping the spacing between the two lobes of the encoding gradient
fixed, $$$\Delta=\delta+t_{180}$$$, where $$$t_{180}$$$ is
the minimum time needed for the refocusing pulse and crusher gradients
surrounding it. By allowing $$$t_{180}$$$ to
stretch, the sequence gains the flexibility to simultaneously choose a venc value, at the cost of lengthening TE. To
encode for both a given b-value and venc, the timing for
the diffusion-encoding trapezoids is modified using the roots $$$\delta$$$ in
the following equation:
$$$c_2\delta^3+(\frac{c_2\zeta^2}{2}-3c_1)\delta+3c_1^2-c_2\zeta^3/10=0$$$
where $$$c_1=2bv_{enc}/\pi\gamma G$$$, $$$c_2=4bv_{enc}^2/\pi^2$$$, and $$$\zeta$$$ is
the trapezoid ramp time. The combination of the values $$$c_1\propto bv_{enc}$$$ and $$$c_2\propto bv_{enc}^2$$$ encode the relevant parameters,
while the ramp time can always be set using the maximum desired slew rate to
keep TE as low as possible. The real positive values of
are then used to determine the required $$$\Delta$$$ values according to
$$$\Delta=\frac{\delta^3-\zeta^3/10+\delta\zeta^2/2}{3\delta(\delta-c_1)}$$$
subject to $$$\Delta\ge\delta+t_{180,min}$$$ and $$$\delta\gt\zeta$$$.
Imaging with such a sequence allows for conventional
diffusion image reconstruction while a separate phase-sensitive reconstruction
pipeline is used to measure coherent motion. Figure 1 details the method to
reconstruct velocity profiles in 3D from the images. Using this, we imaged 5 healthy
volunteers on a GE Healthcare Discovery MR750 (Waukesha, WI) scanner using a
custom high-performance gradient coil set for brain imaging5 with a
gradient strength of 200 mT/m and slew rate of 500 T/m/s. We imaged with a
2-shell sequence (30 directions: 8 in inner shell, 22 in outer shell) with bouter=2000sec/mm2, binner=1400sec/mm2, and venc=0.24mm/s. For each series of
images, we varied the trigger delay (peripheral gating) to image axial slices
containing the lateral ventricles at various phases in the cardiac cycle, using
delay times of 14ms, 64ms, 114ms, 164ms, 214ms, and 514ms delays from the R-wave.Results & Discussion
Figures 2 and 3 show some of the constraints when
simultaneously setting the b-value and the venc. In
Figure 2, several sample encoding pulses are shown for a variety of different venc
values. As the venc decreases,
the two lobes of the encoding pulses become narrower and further apart,
increasing TE overall. The minimum TE sequence will always have the maximum venc
allowed for a given b-value and gradient strength, as shown in Fig. 3. At
higher b-values, the increase in TE is only nominal, while at low b-values and
venc, this increase may make imaging impossible. With a stronger gradient
coil, the maximum allowed venc increases and the effect of lowering
the venc for a given b-value on TE is reduced.
Figures 4 and 5 show the result of the velocity
profile reconstruction. In Fig. 4, the difference between gated and ungated
acquisition clearly demonstrates that with gating, there is a net coherent
motion with RMS 0.065 mm/s in the region of interest. In Fig. 5, the different trigger
delays provide a sequence of motion at different phases in the cardiac cycle.
Immediately after R-wave (systole phase), there is significant lateral motion
around the ventricles. This motion fades, and a clear reversal is seen at 164
ms, demonstrating the expected brain pulsatility motion. This motion accounts
for both tissue movement due to pulsation and CSF flow through the tissue. Both
the motion sequence and the measured velocities agree with independent
phase-contrast scans performed on the same volunteers.Conclusion
Our method for simultaneous imaging of coherent and
incoherent (diffusion) motion in vivo uses magnitude reconstruction of
diffusion data and phase-sensitive reconstruction to image coherent motion.
Adjusting timing parameters of a diffusion pulse sequence to set b-values and venc
values independently allows low venc values on the order of 0.1-0.5
mm/s. This range is well suited to studying CSF pulsatile flow through brain
parenchyma.
Analysis of this motion over the course of several hours,
for example during a sleep study, may provide insight into the waste removal
mechanisms of the brain. Slowly varying CSF mean velocity may be extracted by
averaging the signal over multiple heartbeats to study CSF flow in interstitial
spaces.Acknowledgements
Funding for this research was provided in part by CDMRP W81XWH-16-2-0054.
The views expressed in this abstract are those of the authors and do not reflect the official policy or position of the Uniformed Services University of the Health Sciences, Walter Reed National Military Medical Center, the Department of Defense, of the U.S. Government.
References
1. Jessen
NA, Munk AS, Lundgaard I, et al. The Glymphatic System: A Beginner's Guide.
Neurochem Res. 2015 Dec;40(12):2583-99.
2. Iliff
JJ, Lee H, Yu M, et al. Brain-wide pathway for waste clearance captured by
contrast-enhanced MRI. J Clin Invest. 2013 Mar;123(3):1299-309.
3. Soellinger,
M., Rutz, A., Kozerke, S., Boesiger, P. (2009). 3D cine displacement-encoded
MRI of pulsatile brain motion Magnetic Resonance in Medicine 61(1), 153 162.
https://dx.doi.org/10.1002/mrm.21802
4. Yildiz,
S., Thyagaraj, S., Jin, N., Zhong, X., Pahlavian, S., Martin, B., Loth, F.,
Oshinski, J., Sabra, K. (2017). Quantifying the influence of respiration and
cardiac pulsations on cerebrospinal fluid dynamics using real-time
phase-contrast MRI Journal of Magnetic Resonance Imaging 46(2), 431 439.
https://dx.doi.org/10.1002/jmri.25591
5. Foo
TK, Tan ET, Vermilyea ME, et al. Highly efficient head-only magnetic field
insert gradient coil for achieving simultaneous high gradient amplitude and
slew rate at 3.0T (MAGNUS) for brain microscopic imaging. Magn Reson Med. 2019 (in
press).