Jochen Keupp1, Bernhard Gleich1, and Ulrich Katscher1
1Philips Research, Hamburg, Germany
Synopsis
A combined
acquisition of distortion-free diffusion-weighted images and tissue
conductivity maps is explored using a fully balanced double echo steady state
(DESS) sequence. Banding artifacts are avoided using sufficiently high gradient
moments of the diffusion gradient, such that the banding is contained within
single voxels. The stability of the B1 transceive phase measurement by the
balanced DESS sequence allows the derivation of quantitative tissue
conductivity based second derivative using standard EPT (electrical properties
tomography) methods. Feasibility of simultaneous DWI and EPT is shown on a 3T
MRI system in phantom and volunteer experiments (head).
Introduction
Diffusivity and tissue conductivity are physiologic
parameters with manifold applications e.g. in tumor characterization,
typically assessed by Diffusion Weighted Imaging (DWI) and Electrical
Properties Tomography (EPT) in separate sequences. EPI based DWI sequences
often suffer from geometric distortions (magnetic field inhomogeneity).
Diffusion weighted dual-echo steady-state (DW-DESS) MRI using unipolar gradients provides
a distortion free alternative1,
but is inherently sensitive to motion and does not exploit the steady-state
signal, because of non-balanced gradients. In this study, a balanced DW-DESS
sequence was developed using bipolar DW gradients while avoiding dark-band
artefacts. EPT is based on the transceive phase φ purely related to B1
(not impacted by B0) as in spin-echo (SE) based sequences as well as in
balanced steady state sequences2.
This study investigates the use of φ from balanced DW-DESS as basis for EPT,
which would synergistically allow to assess two relevant physiological
parameters from a single MR acquisition.Methods
A fully-balanced DW-DESS sequence is used in
combination with bipolar DW gradients (example pulse sequence diagram in
Figure 1). For (single echo) balanced SSFP, the dark band spacing only depends
on TR and off-resonances (frequency spacing 1/TR), as the echo is fully
refocused at TE=TR/2. The dark-bands of S+ and S- in fully-balanced DESS also
depend on the moments of the diffusion-weighting gradient lobes, which appear
as an additional off-resonance effect in the gradient direction. At high
gradient moments, the orientation and spacing of the dark bands is dominated by
the gradient effect and less influenced by actual off-resonances. This study
applies sufficiently high gradient moments of the bipolar gradient lobes such
that the spatial distance of the dark band artefacts is reduced to a value
smaller than the imaging voxel size. Dark bands within a voxel partly reduce
the overall signal intensity but are not visible as artefacts. Thus, a strong
diffusion weighting can be induced at high SNR efficiency by combining all
coherence pathways for the overall FID (echo1, S+) and ECHO (echo2, S-) of the
DESS acquisition.
From the acquired DESS signals, conductivity σ was
calculated via $$$σ = \nabla^{2}φ/(2μω)$$$ (with vacuum permeability μ and
Larmor frequency ω) in combination with a bilateral denoising filter2. A
phantom was composed of polyvinylpyrolidon (P), gelatine (G), NaCl (S) and H2O
(W) with different diffusion and conductivity values in an outer and inner
compartment (inner: D=1.04×10-3 mm2/s,
σ=0.66 S/m, P/G/S/W=5/3/0.5/91.5 mass%; outer: D=0.8×10-3 mm2/s,
σ=0.42 S/m, P/G/S/W=25/3/0.3/71.7 m%).
Combined DW-DESS and EPT acquisition was
tested on a 3T MRI system (Achieva TX, Philips, NL) on the phantom and in a
volunteer head examination (male, age 50 yrs), with written consent
obtained, using the following imaging parameters: 3D balanced dual-echo SSFP, 8-channel
head coil, TR/TE1/TE2=31/1.8/26 ms (phantom: 53/1.85/50.8 ms) , FOV
224×224×120 mm3, pixel 1.8×1.8 mm2 ,
reconstruction 224×224, 24 slices (5mm in vivo, 1.8mm phantom), pixel bandwidth
1.3 kHz, bipolar or unipolar diffusion gradients (3 simultaneous
directions, duration 2×11 ms (phantom: 2×22 ms), slopes 0.4 ms, strength
18 mT/m), two signal averages (phantom: 6), total scan duration 2min 55s
(phantom: 7min). Diffusion weighted images were computed as ratio S+/S-.Results
Phantom results are shown in Fig.2. While the first
echo (S+, FID, Fig.2a) shows minimal contrast between the two compartments, the
second echo (S-, Fig.2b) has overall a lower signal and shows further lowered
signal from in the inner compartment with larger diffusivity. The resulting
image (S+/S-, Fig.2c) confirms the diffusion weighting and can be compared with
and ADC map obtained using a standard DWI sequence (EPI, 8 b-values 0…1400,
Fig.2d). Fig.2e shows the conductivity map obtained from S+ and the measured σ values,
inner/outer=(0.77±0.02)/(0.31±0.06)S/m, correspond to the phantom preparation.
Volunteer results are shown in Fig.3 and Fig.4. Using bipolar (Fig.3 a/b)
instead of unipolar diffusion weighting gradients (Fig3. c/d) is much more
robust to motion effects, in particular for the second echo (S-). In Fig. 4, a
DW-DESS image of the volunteer is shown together with the corresponding phase
map and the resulting conductivity map, yielding conductivity values as
expected from literature4.Discussion & Conclusion
A balanced DW-DESS acquisition could be successfully
implemented using large bipolar DW gradients that avoid banding artifacts and
show low motion sensitivity. Although the SNR is lowered by dark band contents
within the voxels, image quality was clearly improved as compared to unipolar
gradients. A drawback of bipolar DW-DESS is given by its limitation of the
achievable b-values. In this initial demonstration, the diffusion weighted
images also include a considerable T2 weighting, because of the long second
echo time (26 or 51 ms). Multiple b-values with the same echo time could be
used to reduce the T2 weighting (b=0 cannot be used because of banding
artifacts).
The transceive phase of DW-DESS can be used for EPT, yielding conductivity maps
with comparable quality as previously obtained in the brain5.
Because of the lower overall SNR in S- images (DW and long TE), EPT
reconstructions were preferably calculated from the first echo S+.
In conclusion, DW-DESS is able to produce distortion-free diffusion weighted
images and conductivity maps simultaneously. It is thus expected to be a
valuable sequence particularly for tumor characterization.Acknowledgements
Cordial
thanks to Christian Stehning and Philipp Karkowski for technical help.References
1. Granlund KL et al., High-resolution, three-dimensional
diffusion-weighted breast imaging using DESS. Magn Reson Imaging. 2014; 32:330
2. Stehning C et al., Real-time conductivity mapping using balanced SSFP
and phase-based reconstruction. ISMRM 2011; 19: 128
3. Katscher U, van den Berg CAT. Electric properties tomography:
Biochemical, physical and technical background, evaluation and clinical
applications. NMR Biomed. 2017; 30:3729
4. Gabriel S et
al., The dielectric properties of biological tissues II. Measurements in the
frequency range 10Hz to 20GHz. Phys Med Biol. 1996;41: 2251
5. Tha KK et al.,
Noninvasive electrical conductivity measurement by MRI: a test of its validity
and the electrical conductivity characteristics of glioma. Eur Radiol. 2018;
28:348