Qingjia Bao1, Ricardo Martinho 1, and Lucio Frydman1
1Weizmann Institute of Science, Rehovot, Israel
Synopsis
A
challenge in functional, diffusivity and spectroscopic MRI studies of mouse
body regions is B0 inhomogeneity, especially at ultrahigh fields. To
map and correct these DB0s, phase difference images are usually acquired
using gradient echo sequences. These, however, are hard to obtain with good
quality in abdominal studies due to motion artifacts. In this study we report a
fully automated 3D map-based shimming method based on an ultrafast sequence that
can overcome these artifacts, delivering optimal B0 homogeneity over the targeted
ROIs. This technique is exemplified with mice studies at 15.2T, where its
usefulness and reproducibility is demonstrated.
INTRODUCTION
Due to recent advances in genetic engineering, transgenic and knockout
mice have become critical models for studying diseases. Alongside this, the
availability of high and ultrahigh field magnets and cryocoils makes the
sensitivity of these preclinical studies increasingly attractive. However, the
small sizes involved in mouse organs and the strong susceptibility gradients
arising at high fields, making it sometimes difficult to reliably obtain a
sufficiently homogenous static magnetic field B0 over the regions of
interest (ROIs)1,2. The optimization of B0’s homogeneity is particularly
important for diffusion, functional and spectroscopic studies. Traditional
iterative methods based on manually maximizing the area under an FID usually
fall short of these needs, as they are time-consuming and often produce sub-optimal
results. A variety of methods have therefore been proposed to provide an automated
adjustment of these shims, based on a concurrent mapping of the ΔB0
inhomogeneities. One widespread such method is FASTMAP3,4, which uses six “pencils” to sample DB0
in a localized sphere. FASTMAP’s accuracy, however, like that of other methods,
typically underperform when applied to motion-prone regions such as the abdomen.
This study introduces a fast, fully
automated 3D map-based shimming method based on spatiotemporal encoding (SPEN)5. SPEN is an ultrafast 2D MRI sequence that can overcome motion
artifacts as well as certain image distortions arising from local dephasings,
thanks to its possibility of performing acquisitions in a “fully-refocused”
mode6 where T2* effects echo throughout the entire FID.
This makes SPEN suitable for targeting challenging abdominal regions. The idea
of the method here introduced is to calculate phase differences between
fully-refocused and non-fully refocused SPEN 3D images, and use the ensuing
multi-slice maps to calculate and correct for ΔB0 distortions. We
have applied this technique to studies of the abdominal region in mouse and
have evaluated its usefulness and reproducibility at 15.2T. Method
Figure
1 shows the SPEN-based pulse sequence used for B0 field mapping:
Fig. 1a presents a fully refocused experiment whereby static T2* effects
are refocused throughout each time point of the acquisition process –and not
just at the echo center as in k-space encoding; Fig 1b introduces a ΔB0-derived the phase difference in the image by altering the echo delay from its
fully-refocused, Ta/2 value. The phase difference between both SPEN experiments
is thenψ=2*π*γΔB(r)*τ; on the basis of repeating this experiment over a series of
phase-encoded values and processing it as described in7, volumetric maps of the DB0 can be obtained. A data
acquisition protocol based on these ideas was written for a Bruker preclinical
scanner running Paravision 6; the processing written for evaluating this data
on the same scanner is schematized in Figure 2. The first part entails a 3D image
reconstruction and 3D phase unwrapping procedure; the latter was based on phase
spatial smoothness properties expected from a physical ΔB0(r),
and was conducted in Matlab. The second part of the procedure involved calculating
the shimming currents required to compensate the resulting ΔB0(r),
and we wrote it based on a regularization algorithm provided by Bruker Paravision
6. RESULTS & DISCUSSION
Figure 3 shows experimental results comparing our new SPEN-based
field mapping method with that supplied by a conventional gradient echo approach
for an in vivo mouse brain. On the upper row are the anatomical images
in the different orientations, whereas in the lower row are the respective field
maps. After the shimming based on SPEN pulse sequence, we can notice the width
of the water peak obtained by PRESS(18*18*8 mm3 voxel) is narrowed, from 58.6 Hz to 31.4 Hz.
The new method was also tested for in in vivo abdominal studies in a naïve
mouse, where GRE gave poor
results as a consequence of motion. Fig. 4a shows EPI images obtained
before the implementation of the SPEN shimming; notice the substantial organs’
distortions. Figures 4b and 4c show the EPI
and SPEN images that resulted after implementing the automated SPEN-based shimming
protocol. The procedure yielded a narrowing of the abdominal water peak (PRESS, 30*30*4 mm3
voxel) from 235.5Hz to 110.8Hz. The ensuing improvement in the EPI image
is evident; shown for the sake of comparison is also a single-shot 2D SPEN
image collected on the same ROI at similar shimming conditions.CONCLUSION
An automatic shimming method based on
ultrafast 3D SPEN was proposed and demonstrated. The advantage of the shimming
method stems from its speed, its robustness to T2* effects, and its ability to
overcome motion artifacts arising when using other 3D field mapping methods. The technique
was found particularly valuable in mouse abdominal studies, where its
usefulness and reproducibility at 15.2 T were demonstrated. Similar advantages are expected to be valid
at higher preclinical fields, or in human studies at ≥7T. Acknowledgements
We
are grateful to Dr. Markus Wick (Bruker BioSpin), and Mr. Koby Zibzener
(Weizmann) for technical assistance. Financial support came from the NIH human
placenta project (R01HD086323), the Thompson Family
Foundation, the Kimmel Institute for Magnetic Resonance and the generosity of
the Perlman Family Foundation.References
1. Miyasaka, N., Takahashi, K. &
Hetherington, H. P. Fully automated shim mapping method for spectroscopic
imaging of the mouse brain at 9.4 T. Magn. Reson. Med. 55,
198–202 (2006).
2. Hetherington,
H. P., Chu, W. J., Gonen, O. & Pan, J. W. Robust fully automated shimming
of the human brain for high-field 1H spectroscopic imaging. Magn. Reson.
Med. 56, 26–33 (2006).
3. Gruetter,
R. Automatic, localized in Vivo adjustment of all first‐and second‐order shim coils. Magnetic
Resonance in Medicine 29, 804–811 (1993).
4. Gruetter,
R. & Tkáč, I. Field mapping without reference scan using asymmetric
echo-planar techniques. Magn. Reson. Med. 43, 319–323 (2000).
5. Tal,
A. & Frydman, L. Single-scan multidimensional magnetic resonance. Prog.
Nucl. Magn. Reson. Spectrosc. 57, 241–292 (2010).
6. Schmidt,
R. & Frydman, L. New spatiotemporal approaches for fully refocused,
multislice ultrafast 2D MRI. Magn. Reson. Med. 71, 711–722
(2014).
7. Yon,
M., Bao, Q., Henriques, R., Shemesh, N. & Frydman, L. SPatiotemporal
ENcoding ( SPEN ) at ultra-high fields : Applications to high resolution ( <
100 µm isotropic ) in vivo mouse brain DTI. Proc. Intl. Soc. Mag. Res. Med.
1021 17–19 (2019).