Yong Chen1, Xiaopeng Zong2, Dan Ma1, Weili Lin2, and Mark Griswold1
1Radiology, Case Western Reserve University, Cleveland, OH, United States, 2Radiology, University of North Carolina at Chapel Hill, Chapel Hill, NC, United States
Synopsis
In
this study, we developed a 3D MRF method in combination with fat navigator to
improve its motion sensitivity for neuroimaging. A rapid fat navigator sampling
was achieved at 3T by using the stack-of-spirals acquisition and non-Cartesian
spiral GRAPPA. The improvement in motion robustness was achieved without
increasing the scan time for quantitative tissue mapping. Our preliminary
results demonstrate that 1) the added fat navigator sampling does not influence
the accuracy of T1 and T2 quantification, and 2) the
motion robustness for quantitative tissue mapping using MRF was largely
improved with the proposed method.
Introduction
Subject motion is ubiquitous in clinical imaging and presents one
of the major challenges for MR imaging of children as well as many other difficult
patient populations. MR Fingerprinting (MRF) is a quantitative imaging
technique which can provide simultaneous quantification of multiple tissue
properties (1). Compared to conventional MR imaging, MRF often utilizes a
non-Cartesian spiral trajectory for in-plane encoding, which is known to yield
better performance in the presence of motion. The template matching algorithm
used to extract quantitative tissue properties also provides a unique
opportunity to reduce motion artifacts (1). However, with the prolonged 3D MRF acquisitions
(typically around 5~10 min) (2-4), the scan is more susceptible to subject motions
and further improvement in motion robustness is needed. In this study, we
aimed to improve the motion sensitivity of 3D MRF technique by using fat
navigator accelerated with spiral GRAPPA.Methods
All measurements were performed on a Siemens 3T Skyra scanner using a 20-channel
head coil. A 3D MRF protocol using variable flip angles and golden-angle spiral
trajectories was adopted from the literature (5). The imaging parameters
included: FOV, 30×30 cm; matrix size, 256×256; slice thickness, 3 mm; number of
slices, 72; flip angles, 5°~12°; MRF time frame, 576. Similar to the original
MRF method, each MRF time frame was highly undersampled in-plane by acquiring
only one spiral arm (1). The partition direction is linearly encoded and
acquired sequentially (Fig. 1a). A 2-sec waiting time was applied at the end of
acquisition of each partition for longitudinal signal recovery. The total
acquisition time for one 3D acquisition was about 11 min.
To improve the motion
robustness of 3D MRF, a fat navigator (spatial resolution, 2×2×3 mm3)
was inserted during the 2-sec waiting time at the end of each partition
acquisition (Fig. 1a) (6). A 1-2-1 binomial RF pulse was used for fat
excitation. Compared to previous implementation of a fat navigator at 7T, the
chemical shift between fat and water is reduced from ~1000 Hz to 440 Hz at 3T,
which increases the duration of the fat excitation pulses (~3.3 msec). To
accelerate the 3D fat navigator to minimize the sensitivity of navigator itself
to motion, a stack-of-spiral trajectory was used along with the spiral GRAPPA
technique (Fig. 1b) (7,8). To identify the optimum setting for combined
in-plane and through-plane accelerations, a 3D MRF scan was performed on a
normal volunteer and the fat navigator was fully sampled during the scan.
Retrospective undersampling was performed with various combination of in-plane
and through-plane reduction factors and the results were compared to the
reference results obtained with no acceleration.
We
first validated the quantitative accuracy of the proposed method on phantom
experiments and the results were compared to those acquired from the standard
3D MRF without the fat navigator. The proposed method was further tested on
three normal volunteers (male; mean age, 48±9 years). The subjects were
instructed to move intentionally during the MRF scans. Based on the reconstructed
fat navigator images, motion signal was extracted using the SPM12 toolbox. Both
translational and rotational motions were corrected in k-space followed by non-Cartesian
imaging reconstruction. T1 and T2 maps were computed from
3D MRF datasets using pattern matching. Results
Fig. 2 shows the
phantom results obtained with and without fat navigator. A good agreement was
observed in both T1 and T2 values, which suggests that
the application of fat navigator does not influence the tissue quantification
based on water signal. Fig.3 shows the results of different acceleration
schemes to acquire the fat navigator. A combined in-plane and through-plane
acceleration of R6×3 was selected in the study based on the accuracy of motion
tracking and the acquisition speed (~0.5 sec per navigator module). Representative
T1 and T2 maps obtained with and without motion
correction from a normal volunteer are presented in Fig. 4 along with the estimated
motion curves. Fig. 5 shows an animation of T1 and T2
maps before and after motion correction from another volunteer. All these
results demonstrate that the proposed method can be applied to effectively
improve motion robustness for the 3D MRF method in neuroimaging.Discussion and Conclusion
In this study, a fat
navigator was integrated at each partition with 3D MRF to effectively reduce its
motion sensitivity in neuroimaging. In combination with non-Cartesian spiral
GRAPPA, a rapid fat navigator sampling (~0.5 sec) was achieved at 3T, reducing
its sensitivity to potential motion. The improvement in motion robustness was
achieved without increasing the scan time for quantitative tissue mapping. Since
the current study was focused on evaluation of motion sensitivity of the
proposed method, recent advances to accelerate 3D MRF itself were not applied
(2-4). These methods can be combined with the proposed method to reduce the
total acquisition time and motion artifacts in the future. Future studies will also
be performed to evaluate its performance with various types of head motions, its
accuracy for high-resolution quantitative imaging, and its application with
challenging populations such as children.Acknowledgements
No acknowledgement found.References
1. Ma D, et al. Nature, 2013; 187–192.
2. Ma D, et
al. MRM, 2018; 2190-2197.
3. Liao C, et al. NeuroImage, 2017; 13-22.
4. Chen Y,
et al. ISMRM 2019, p17.
5. Chen Y, et al. Radiology, 2019; 33-40.
6. Gallichan
D, et al. MRM, 2016; 1030-1039.
7. Seiberlich N, et al. MRM, 2011;1682-1688.
8.
Chen Y, et al. Invest Radiol, 2015; 367-375.