Evgeniy Alekseevich Koreshin1, Mikhail Zubkov1, Alexander Yurievich Efimtcev2, Alexandr Mikhailovich Gulko3, and Irina Valerievna Melchakova1
1Faculty of Physics and Engineering, ITMO University, Saint-Petersburg, Russian Federation, 2Department of Radiology, Federal Almazov North‐West Medical Research Center, Saint-Petersburg, Russian Federation, 31st urology department, City Center of Endourology and New Technologies, Saint-Petersburg, Russian Federation
Synopsis
We
present a new design of a radiofrequency resonator for urological applications.
The resonator functions by inductively coupling to the body coil of a 1.5 T
MR-scanner. This configuration of the Tx-Rx path allows increasing the
transmission efficiency and signal
to noise ratio (SNR) while reducing the specific absorption rate (SAR). Phantom
and in-vivo imaging shows that the
resonator provides around 100-fold SAR reduction and 10-fold
transmission efficiency increase. Phantom imaging shows doubled SNR compared to the commercial flexible 4-element Rx coil. In-vivo imaging shows only a 50%
increase in SNR, improved patient positioning and reduced the artifacts
rate.
Introduction
MR imaging of the male reproductive
system pathologies is a rare occurrence in clinical practice, often due to the weak
fit of the standard radiofrequency (RF) coils to imaging of the penis. In order
to diagnose penis-associated disorders (e.g., fractures1, Peyronie
disease2 etc.)
with MRI a surface coils is commonly used with the patient in supine position while
penis is being fixed to the lower patient body3, which
leads to the compression of the tissues, a decrease in blood flow and an
uncomfortable position for the patient, provoking increased motion artifacts. In the present study, these
problems have been sought to be overcome by developing a new RF resonator based
on the birdcage coil operation principles4. The
study presents some of the results obtained during simulation and in-vivo imaging to show that the
proposed resonator in combination with a scanner body coil (BC) improves
patient positioning, signal-to-noise ratio (SNR) and RF-safety.Methods
The resonator (Figure 1 A) design was based on a low-pass birdcage coil
and includes 8 legs (wires) and 8 capacitors, two circle end rings of the birdcage coil are formed
in the current design by a metal screen and full metallization on the opposite sides of the PCB (Rogers RO4003, thickness 0.508mm, permittivity of 3.38 and tan
δ 0.0027). Geometrical parameters of the resonator were dictated by the
operation frequency of 63.8 MHz and the target organ size. This led
to the length of the wire H = 120mm, wire gap L = 33mm and structural capacity
of the PCB capacitor of 27.5 pF. The tuning in experiments was provided via adjusting the length of the resonator
legs implemented as telescopic brass tubes. The resonator operated via
inductive coupling of the first eigenmode of the resonator (similar to magnetic
dipole TE01δ) at the 63.8 MHz frequency to the BC
of a 1.5 T MR-scanner. This configuration of the Tx-Rx path allowed increasing
the transmission efficiency, SNR and reducing the SAR. The coil was initially
simulated in CST Microwave Studio; coil tuning, field homogeneity and SAR (whole body and local maximum 1g SAR) levels were
assessed. Electromagnetic (EM) simulations of the Tx-Rx path
included realistic BC, voxel human model with an additional penis model and the
resonator under study (Figure 1 B). Next, a human subject was scanned with the proposed coil. The study was
approved by the local ethics committee and the volunteer provided written
informed consent. The subject was scanned in the prone position elevated above the scanner table on a specially manufactured
platform, which incorporated the designed resonator. The subject’s penis was
freely lowered into the resonator cavity (Figure 1 C). MR-images
were obtained on a clinical MR-scanner (1.5 T Siemens Magnetom Espree) with standard
gradient (FLASH, FA = 70°, TR = 710 ms, TE = 7.2 ms, matrix 384x269, slice thickness 3 mm) and
spin echo (TSE, FAexc = 90°, FAref = 150°, TR = 5300 ms,
TE = 104 ms, ETL = 23, matrix 320x275, slice thickness 3 mm) pulse sequences. Three
RF configurations were considered: BC for transmission and reception, BC with
resonator for transmission and reception, BC for transmission and a flexible 4-element coil for reception. When the resonator was used, the power supplied to the
BC was adjusted to achieve the same B1+ field in the
center of the penis. The SNR was calculated by dividing average value of the signal in corpus
spongiosum along the penis length by the standard deviation of noise in signal-free
regions5.Results
The magnetic field
simulations and experimental flip angle measurements can be seen in Figure 2, showing
high similarity between the simulated and experimental data (with
coefficient of determination of 0.94).
As can be seen, the
size of the homogeneous
region (i.e., where field
value deviation is does
not exceed 0.1 of
the average
field value) inside the resonator is
75 mm long. Assessment of the transmission efficiency shows a 9.8 times
increase (simulation confirmed by experiment) compared to the BC-only case. Maximum
(Figure 3) and whole body SAR decrease 95 and 46 times respectively. SNR
measurements in
phantom (Figure 4) with three different
Tx-Rx path configurations show
a 14-fold sensitivity gain for the resonator compared with the BC-only case,
and around 2-fold increase for the resonator compared with the 4-channel flexible coil.
In-vivo measurements (Figure 5) show
only 27% SNR increase when comparing the resonator against the flexible
4-channel coil in T1-weighted images
and 50% rise in T2-weighted images. Finally,
MR-images of the penis obtained with
the resonator contain fewer motion artifacts due to a more comfortable
patient positioning and smaller sensitivity to flow and motion outside the
imaging region.Conclusion
It was demonstrated here
that using the proposed RF
resonator design in urology and andrology
MRI applications can improve
image quality when compared to the body coil and to a more clinically relevant
case of surface coils. The improved image quality can
be used to provide better
resolution or shorter scanning times.
At the same time SAR reduction shows the scanning procedure to
become safer.Acknowledgements
This work was supported by the Russian Science Foundation (Grant No. 18-79-10167)References
1. Pretorius ES, Siegelman ES, Ramchandani P, Banner MP. MR
Imaging of the Penis. RadioGraphics 2013; 21(1): 283–298.
2. Helweg G, Judmaier W,
Buchberger W, Wicke K, Oberhauser H, Knapp R et al. Peyronie’s disease:
MR findings in 28 patients. Am J Roentgenol 1992; 158(6):
1261–1264.
3. Guler I, ödev K,
Kalkan H, Simsek C, Keskin S, Kilinç M. The value of magnetic resonance imaging
in the diagnosis of penile fracture. International Braz J Urol 2015; 41: 325-328
4. Hayes CE, Edelstein
WA, Schenck JF, Mueller OM, Eash M. An efficient, highly homogeneous
radiofrequency coil for whole-body NMR imaging at 1.5 T. J Magn Reson 1985; 63(3): 622–628.
5. Dietrich O, Raya JG,
Reeder SB, Reiser MF, Schoenberg SO. Measurement of signal-to-noise ratios in
MR images: Influence of multichannel coils, parallel imaging, and
reconstruction filters. J Magn Reson Imaging 2007; 26(2): 375-385