Yonghyun Ha1, Kartiga Selvaganesan1, Charles Rogers III1, Baosong Wu1, Sajad Hosseinnezhadian1, Gigi Galiana1, and R. Todd Constable1
1Department of Radiology and Biomedical Imaging, Yale School of Medicine, New Haven, CT, United States
Synopsis
In this
work, we designed a frequency division duplex RF system for frequency encoding
using Bloch-Siegert shift at very low field, with a modification of dual-band
pass filters. Although the off-resonance frequency (870 kHz) is very close to
the Larmor frequency (1 MHz), the applied off-resonance signal can be filtered
out by a modified dual-band pass filter on the receive path. This system allows
us to apply a 870 kHz transmit pulse while receiving 1 MHz signal from the RF
coil.
Purpose
By
applying off-resonant radiofrequency (RF) pulses, a spatially dependent
frequency shift of the Larmor frequency is introduced1. Using this
so-called Bloch-Siegert (BS) shift, it is possible to perform spatial frequency
encoding without gradient coils2. However, simultaneous transmit
(Tx) at the off resonance, and receive (Rx) at the resonance frequency is
required for BS frequency encoding. One possible approach is using a frequency
division duplex (FDD) RF system with RF filters. For our low field magnet3, it is challenging to design
the RF filters since the two frequencies are close. In this work, we introduce a FDD
RF system, which can be utilized to transmit a BS pulse during MR signal
reception. The design goal of this system is to achieve a small insertion loss
at 1 MHz and high insertion loss (> 58dB) at 870 kHz, between Tx and Rx
ports. Methods
Fig. 1 shows the block diagram of the FDD
RF system. A 1 MHz signal can be applied from the power amplifier (A) to the
coil (C) during spin excitation (red line on Fig. 1), while this signal on
receive pass (C to B) is blocked by receive switch. During signal reception, 870
kHz off-resonance signal for BS shift can be applied to the coil (A to C) while
1 MHz of the signal from the coil (C to B) is received by the preamplifier.
Filter 1 and filter 2 are blocking the signal at 870 kHz and 1 MHz,
respectively. A Dual-tuned coil can
be used with this system. The maximum output power of the RF power
amplifier (AR 0.5-15-1E3-3C, PCP, NY, USA) is 1 kW (60 dB) and the output P1dB
and gain of the preamplifier (P4.2VD NMR, Ar2, CT, USA) are +22 dB and 20 dB,
respectively. Thus, if the maximum power of the off-resonance signal is applied
to the coil and the insertion loss of the filter 2 at 870 kHz is less than 58
dB, the signal can be saturated after amplification by preamplifier (60 dB - 58
dB + 20 dB = 22 dB). However, in practice, it is not simple to build the filter
with greater than 58 dB insertion loss at 870 kHz while keeping small insertion
loss at 1 MHz due to the frequencies being very close. In this work, dual-band
pass filter4 was modified for this purpose (Fig. 2). For filter 1, f1 and f2 were tuned at 870 kHz and 1 MHz, respectively. A 3rd
order dual-band pass filter, with two series filters and a shunt filter, was
used for filter 2. The f2
and f3 of the series
filters were tuned to 870 kHz and 1 MHz, respectively and the same configuration of
filter1 was used as shunt filter of the filter 2. Figure 3 shows the schematic
diagram and the photo of the whole RF system. It is designed that the diodes D2, D3, and D4
are tuned on/off simultaneously. During
the spin excitation, 200 mA of the forward current can be applied to the D2 and split into D3 and D4. However, RF signal applied from port A can be
transmitted to port B through the DC bias line. For this reason, a 3rd order
band stop filter was added between D2
and D3. The S-parameters
between the ports A, B, and C were measured for Tx and Rx conditions using vector
network analyzer.Results
For the
spin excitation, the port A and the port C are connected by turning on the D2, while the impedance
between the port B and the port C is high by the receive switch with turning on
D3 and D4. As shown in Fig. 4, the
insertion loss between port A and C was 1.11 dB at 1 MHz, and 69.24 dB between port A and B. Thus, an RF pulse with a frequency of 1MHz can be applied
from port A to port C, without it reaching port B. The insertion loss
between port A and C was 1.88 dB at 870 kHz and 66.55 dB between port A
and B. When D1 is turned
on, off-resonance signal can be applied from port A to port C but is
blocked by filter 2. At the same moment, receive signal from port C with 1
MHz frequency can be transmitted to port B but is blocked by filter 1. The
insertion loss between port B and C was 67.61 dB and 1.70 dB at 870 kHz and 1 MHz, respectively and it between port A and C was 39.53 dB during the signal reception.Discussion / Conclusions
We have built and demonstrated the performance of an FDD
system for frequency encoding using Bloch-Siegert shift. The insertion loss
from Tx port to Rx port at 870 kHz was about 69 dB which is sufficient for
amplification of the Rx signal without saturating the preamplifier. If
necessary, the isolation can even be improved further by increasing filter 2 to
a higher order filter. Insertion losses for other signal paths (1 MHz Tx,
870 kHz Tx, and 1 MHz Rx) were less than 2 dB which is sufficient for RF signal
transport.Acknowledgements
No acknowledgement found.References
1. F. Bloch and A. Siegert, “Magnetic resonance for
nonrotating fields,” Phys. Rev., vol.
57, no. 6, pp 522-527, Mar. 1940.
2. Z. Cao, E. Y. Chekmeney, and W. A. Grissom, “Frequency
encoding by Bloch-Siegert shift,” in Proc.
22nd Annu. Meet. Int. Soc. Magn. Res. Med., Milano, Italy, May
10-16, 2014.
3. R. T. Constable, C. Rogers III, B. Wu, K. Selvaganesan, and G. Galiana, "Design of a novel class of open MRI devices with nonuniform B0, field cycling, and RF spatial encoding," in Proc. 27th Annu. Meet. Int. Soc. Magn. Res. Med., Montreal, Canada, May 10-13, 2019.
4. H. Joshi, and J. Chappell, “Dual-band
lumped-element bandpass filter,” IEEE
Trans. Microw. Theory Tech., vol. 54, no. 12, pp. 4169-4177, Dec. 2006.