Philipp Eirich1,2, Tobias Wech1, Julius F. Heidenreich1, Manuel Stich1,3, Nils Petri4, Peter Nordbeck4, Thorsten A. Bley1, and Herbert Köstler1
1Department of Diagnostic and Interventional Radiology, University Hospital Würzburg, Würzburg, Germany, 2Comprehensive Heart Failure Center Würzburg, Würzburg, Germany, 3Siemens Healthcare, Erlangen, Germany, 4Department of Internal Medicine I, University Hospital Würzburg, Würzburg, Germany
Synopsis
An automatic pre-emphasis based on the Gradient
System Transfer Function (GSTF) was applied to real-time cardiac MR imaging at 3T to
compensate for deviations of spiral k-space trajectories. This
yielded cine series with coverage of the whole heart in free-breathing, less
than 40 s total scan time and a temporal resolution of 50 ms. The
developed framework was compared to a gated Cartesian acquisition in multiple
breath-holds, in one healthy volunteer and one patient suffering from cardiac
arrhythmia.
Introduction
Non-Cartesian data acquisition suffers from
imperfections of the dynamic gradient system, which lead to k-space
misregistrations and consequently to image artifacts. Measuring the systems impulse
response in terms of the Gradient System Transfer Function (GSTF) and using it
for trajectory correction can mitigate image artifacts not only in post-processing
but also as a pre-emphasis1,2.
In this study, the workflow developed in [2] was
applied to spiral real-time cardiac functional MR imaging in free-breathing and
the results were compared to an ECG-gated Cartesian cardiovascular product sequence
with multiple breath-holds, as typically applied in clinical routine.Methods
The first order GSTF self-terms of the x-, y-
and z-axis of a 3T system (MAGNETOM Prismafit, Siemens Healthcare,
Erlangen, Germany) were determined by measuring the phase evolutions in two
parallel slices of a spherical phantom. 12 triangular input gradients with
varying duration were switched as test waveforms (TR = 1 s, read-out BW = 100 Hz/px,
dwell-time = 8.7 µs, slice thickness = 3 mm, slice distance from isocenter = ± 16.5
mm, flip-angle = 90°, 100 measurements)1,3,4. The pre-emphasis of
the applied spiral read-out gradients was realized by a projection on the
physical x-, y- and z-axis and a subsequent multiplication with the respective
inverse GSTF in the frequency domain.
Slew-rate
optimized spiral gradients with variable density were generated using an open
source Matlab toolbox (Hargreaves B., https://mrsrl.stanford.edu/~brian/vdspiral/) and implemented in a 2D FLASH
sequence. Ten
consecutively acquired and equally distributed spiral interleaves form a three-fold
undersampled k-space, which represents one time frame during image
reconstruction (TR = 4.96 ms, temporal resolution < 50 ms). The
inter-frame k-space angle increment was defined as φ = (2·GA)/10, with the golden angle GA = (2·π)/(1+√5) ≈ 111.25°.
Further measurement parameters were: TE = 0.84 ms,
read-out BW = 320 Hz/px, dwell-time = 2.2 µs, flip-angle = 15°, FOV = 480 mm × 480 mm, in-plane spatial resolution = 1.34 mm × 1.34 mm, slice thickness = 8 mm.
The framework was tested in one healthy volunteer and
one patient suffering from heart failure with reduced ejection fraction (HFrEF)
and intermittent arrhythmia. An ECG-gated Cartesian cardiovascular
sequence with multiple breath-holds served as reference. All investigations
were performed in short-axis orientation covering the whole heart by means of
11 slices. Spiral acquisitions were performed in free-breathing and 3.5 s scan-time
per slice (i.e. all slices < 40 s measurement time). The Cartesian reference
required breath-holds of 10 s per slice, leading to a total measurement
time of 3-5 min including additional pauses for breathing and recovery.
For the Cartesian reference technique the measurement
parameters were adjusted individually to match the spatial resolution of the spiral
acquisition and to avoid aliasing: volunteer / patient: TE = 2.36 ms / 2.29 ms,
TR = 4.70 ms / 4.60 ms, read-out BW = 707 Hz/px, dwell-time = 3.4 µs,
flip-angle = 15°, FOV = 310 mm × 253 mm / 360 mm × 294 mm, temporal resolution
= 42.2 ms / 27.4 ms, in-plane spatial resolution = 1.49 mm × 1.49 mm / 1.73 mm ×
1.73 mm, slice thickness = 8 mm, T-GRAPPA = 2.
Spiral acquisitions were reconstructed offline using
GRAPPA operator gridding (GROG)5 and a low rank plus sparse compressed
sensing model6.Results
Video 1 shows the results of both methods in a
midventricular slice of the healthy volunteer in short-axis orientation: the
breath-held gated Cartesian
reference technique on
the left and two adjacent
R-R intervals of the free-breathing spiral real-time sequence on the right. In
general, comparable image quality can be observed. No artifacts corresponding
to trajectory errors were traceable. Solely, off-resonance in regions with fat
tissue led to blurring artifacts and a slight jitter associated with residual
aliasing remained. Respiratory motion was resolved and can be monitored in the
image series. The ejection fraction (EF) was 63.8 % for the Cartesian reference
and 60.4 % for the spiral real-time technique.
Video 2 assembles all short-axis slices of the healthy
volunteer by means of one synchronized R-R interval using the spiral real-time
technique in free-breathing. Video 3 displays the cine series of a
midventricular slice of the patient suffering from HFrEF in accordance to video
1. Apparent arrhythmia led to corrupted triggering and thus impaired image
quality in the gated Cartesian reference technique, while the arrhythmic
cardiac kinetics could be resolved by the spiral real-time sequence. Quantification
led to an EF of 21.5 % for the Cartesian reference and 23.0 % for the spiral
real-time technique.Discussion & Conclusion
The GSTF-based pre-emphasis serves as a robust
setup to mitigate trajectory errors when using non-Cartesian k-space trajectories.
A framework was established to resolve cardiac motion in real-time with 50 ms
temporal resolution. Whole heart coverage can be achieved in less than 40 s in
free-breathing, which increases time efficiency as well as patient comfort. Furthermore,
the proposed real-time imaging allows for the omission of ECG-gating and
thereby enables proper cardiac cine imaging in patients with cardiac
arrhythmia. Quantitative analysis yielded similar ejection fractions for both
methods.Funding
Comprehensive Heart Failure Center Würzburg,
Grant BMBF 01EO1504Acknowledgements
No acknowledgement found.References
1. Stich M. et al., Gradient waveform
pre-emphasis based on the gradient system transfer function, Magnetic Resonance
in Medicine, 80(4):1521-1532 (2018)
2. Eirich
P. et al., Spiral imaging using a fully automatic pre-emphasis based on the
Gradient System Transfer Function (GSTF), Abstract #3850, ISMRM 2019
3. Liu H. et al., Accurate Measurement of
Magnetic Resonance Imaging Gradient Characteristics, Materials, 7:1-15 (2014)
4. Campbell-Washburn A. E. et al., Real-Time
Distortion Correction of Spiral and Echo Planar Images Using the Gradient
System Impulse Response Function, Magnetic Resonance in Medicine,
75(6):2278-2285 (2016)
5. Seiberlich N. et al., Non-Cartesian data reconstruction
using GRAPPA operator gridding (GROG), Magnetic Resonance in Medicine,
58:1257-1265 (2007)
6. Otazo M. et al., Low-rank plus sparse
matrix decomposition for accelerated dynamic MRI with separation of background
and dynamic components, Magnetic Resonance in Medicine, 73(3):1125-1136 (2015)