Kristina Sonnabend1, Gerrit Brinker2, David Maintz1, Alexander Bunck1, and Kilian Weiss1,3
1University of Cologne, Faculty of Medicine and University Hospital Cologne, Institute for Diagnostic and Interventional Radiology, Cologne, Germany, 2University of Cologne, Faculty of Medicine and University Hospital Cologne, Department of General Neurosurgery, Center for Neurosurgery, Cologne, Germany, 3Philips GmbH, Hamburg, Germany
Synopsis
Intraspinal
compliance is related to neurological diseases and can be measured by pulse
wave velocity (PWV). A multiband CINE phase-contrast MRI sequence was developed
to measure the intraspinal PWV between two simultaneously acquired slices along
spine. The method was evaluated in-vitro, in healthy-subjects and in a normal
pressure hydrocephalus patient. In-vitro results show good reproducibility and
dependency on transmural pressure in agreement with theory. A higher PWV compared
to healthy subjects is observed in the patient. A decline in PWV after shunt
surgery is detected, making it a promising tool for investigation and treatment
follow-up of neurological diseases.
Introduction
The
compliance describes the ability to accommodate an increase of volume by a
specific increase of pressure in a vessel. The intracranial and intraspinal
compliance are related to neurological diseases, e.g. normal pressure
hydrocephalus (NPH) and Chiari malformation1,2. As the
compliance cannot be measured directly, indirect methods were developed.
However, these methods are either based on invasive measurements3,4 or multiple
non-invasive measurements in combination with complex models using a number of
assumptions2,5 An alternative and comparative simple way to
indirectly measure compliance and transmural pressure are pulse wave velocity
(PWV) measurements in accordance with the Bramwell-Hill equation6. However, high pulse wave velocities and
physiological variations (e.g. heart rate) make PWV measurements of the
cerebrospinal fluid (CSF) along the spine very challenging. To this end, we
developed a new method that enables PWV measurements of the CSF along the
spinal cord within one measurement of two simultaneously acquired transversal
slices using high temporal resolution multiband CINE phase-contrast MRI (PCMRI)
velocity measurements7-9. The new
method was evaluated in-vitro, in healthy volunteers and in a normal pressure
hydrocephalus patient.Methods
All measurements were carried out on a clinical 3T MRI system (Ingenia, Philips Healthcare, Best, The Netherlands). For PWV measurements two transversal imaging slices were acquired simultaneously using multiband CINE PCMRI with imaging parameters shown in Table 1. To allow for highest temporal resolution as possible only velocity-encoded images were acquired and velocity-compensated acquisition were waived for the PWV measurements. The background phase was removed by subtraction of the first dynamic from all subsequent dynamics. For in-vivo measurements, heart rate was recorded using a wireless pulse oximeter. PWVs were calculated as shown in Figure 1 using a foot-to-foot algorithm10.
In-vitro data were acquired using a tube system and a flow pump
(CardioFlow 5000, Shelley Medical, London, Ontario, Canada) with 6 different
transmural pressure levels, increasing in steps of 5 cmH20, relative to the initial pressure. At each pressure level, the PWV
sequence was repeated five times.
Measurements of PWV in 20 healthy subjects were performed in two
subsequent scanning sessions for each subject. In each scanning session, the
PWV sequence was repeated five times, resulting in ten measurements per
subject.
One 76-year-old male patient suffering from an NPH was included in the
study. The patient data were acquired one day before and five days after CSF
shunt surgery. The multiband CINE PCMRI sequence was repeated three times in
each scan session. Results
PWVs of the in-vitro study are shown in Figure 2. The minimum and maximum measured PWVs were 2.45 m/s and 3.78 m/s,
respectively. The standard deviations for five repeated measurements at the
same pressure level reach from 2.36% to 3.24%. Pearson correlation between
relative pressure levels and PWV² averaged over repeated measurements was calculated as R = 0.93 (p =
0.008).
The mean PWV over all healthy subjects was 5.92 ± 3.81 m/s with. Figure 3 shows the average PWV for each session and subject. The mean absolute
difference between average PWVs of session 1 and 2 was 1.53 ± 1.31 m/s, which was found to
be not significant using students t-test (p= 0.609).
The mean PWV of the NPH patient measurements before and after surgery
were 33.80 ± 6.75 m/s and 15.69 ± 3.37 m/s, respectively (Figure 3).Discussion
The in-vitro study showed good reproducibility with standard deviations
under 5% for repeated measurements. The increasing PWV with increasing pressure
seem to follow a square-root-function, in line with the Bramwell-Hill equation6. This is confirmed by the excellent linear correlation between pressure
and PWV².
The in-vivo data showed no significant differences between the averages
of two sessions with 5 repeated PWV measurements. The mean PWV over all
subjects is within the range of estimated PWV in previous studies11-13. However,
large standard deviations between repeated measurements were found in in-vivo
data. As the in-vitro results show smaller standard deviations, these
variations may mainly be attributed to physiological variations of the PWV,
e.g. due to differences in respiration and heart rate between the repeated
measurements, instead of low measurement accuracy. Future studies to
investigate the interconnection between physiologic parameters such as heart
and respiratory rate and intraspinal compliance and pressure in healthy
subjects are warranted.
CSF PWV in the NPH
patient proved to be markedly higher compared to those in healthy subjects providing evidence
for a significantly reduced compliance in NPH. A decrease in PWV was observed
after CSF shunt surgery.Conclusion
A method to measure the PWV of the CSF within one PCMRI measurement was
successfully developed and evaluated in-vitro, in healthy-subjects and in an
NPH patient. In-vitro results are in good agreement with theory of the
Bramwell-Hill equation. In-vivo measurements including the preliminary results
in an NPH patient indicated that the presented technique is a very promising
tool for investigation, detection, and treatment follow-up of the
interconnection between intraspinal compliance and physiological factors or
neurological diseases. Further studies will show, if the spinal CSF PWV, which
can be assessed rapidly and non-invasively by the presented technique, may be a
relevant parameter in the diagnosis of neurological diseases and replace
invasive pressure measurements in the future. Acknowledgements
No acknowledgement found.References
1. Bothwell
SW, Janigro D, Patabendige A. Cerebrospinal fluid dynamics and intracranial
pressure elevation in neurological diseases. Fluids Barriers CNS 2019;16(1):9.
2. Alperin N, Sivaramakrishnan A, Lichtor
T. Magnetic resonance imaging-based measurements of cerebrospinal fluid and
blood flow as indicators of intracranial compliance in patients with Chiari
malformation. Journal of Neurosurgery 2005;103(1):46-52.
3. Marmarou A, Shulman K, LaMorgese J.
Compartmental analysis of compliance and outflow resistance of the
cerebrospinal fluid system. J Neurosurg 1975;43(5):523-534.
4. Cardoso ER, Rowan JO, Galbraith S.
Analysis of the cerebrospinal fluid pulse wave in intracranial pressure. J
Neurosurg 1983;59(5):817-821.
5. Alperin NJ, Lee SH, Loth F, Raksin PB,
Lichtor T. MR-Intracranial pressure (ICP): a method to measure intracranial
elastance and pressure noninvasively by means of MR imaging: baboon and human
study. Radiology 2000;217(3):877-885.
6. Bramwell JC, Hill AV. The Velocity of
the Pulse Wave in Man. Proceedings of the Royal Society B: Biological Sciences
1922;93(652):298-306.
7. Barth M, Breuer F, Koopmans PJ, Norris
DG, Poser BA. Simultaneous multislice (SMS) imaging techniques. Magn Reson Med
2016;75(1):63-81.
8. Breuer FA, Blaimer M, Heidemann RM, et
al. Controlled aliasing in parallel imaging results in higher acceleration
(CAIPIRINHA) for multi-slice imaging. Magn Reson Med 2005;53(3):684-691.
9. Breuer FA, Blaimer M, Mueller MF, et
al. Controlled aliasing in volumetric parallel imaging (2D CAIPIRINHA). Magn
Reson Med 2006;55(3):549-556.
10. Latham RD, Westerhof N, Sipkema P, et al.
Regional wave travel and reflections along the human aorta: a study with six
simultaneous micromanometric pressures. Circulation 1985;72(6):1257-1269.
11. Jackson JR, Williams B. Errors in
velocity measurement by the Pitot principle in fluids with slowly propagated
pressure waves. J Biomed Eng 1979;1(1):50-54.
12. Kalata W, Martin BA, Oshinski JN, et al.
MR measurement of cerebrospinal fluid velocity wave speed in the spinal canal.
IEEE Trans Biomed Eng 2009;56(6):1765-1768.
13. Martin BA, Kalata W, Loth F, Royston TJ,
Oshinski JN. Syringomyelia hydrodynamics: an in vitro study based on in vivo measurements.
J Biomech Eng 2005;127(7):1110-1120.