Jinmin Xu1,2, Jason Stockmann2, Berkin Bilgic2, Thomas Witzel2,3, Jaejin Cho2, Congyu Liao2, Zijng Zhang1,2, Huafeng Liu1, and Kawin Setsompop2,3,4
1State Key Laboratory of Modern Optical Instrumentation, College of Optical Science and Engineering, Zhejiang University, Hangzhou, China, 2Department of Radiology, A.A. Martinos Center for Biomedical Imaging, Massachusetts General Hospital, Charlestown, MA, United States, 3Harvard Medical School, Boston, MA, United States, 4Massachusetts Institute of Technology, Harvard-MIT Health Sciences and Technology, Cambridge, MA, United States
Synopsis
Wave-CAIPI is a parallel imaging technique that can provide high accelerations with negligible g-factor and artifact penalties. However, gradient amplitude & slew rate limits impose a limitation on the amount of wave-encoding and hence the acceleration capability of this technique. In this study, we propose a multi-frequency wave-encoding method (mf-wave) that uses both the gradients and a combined RF and B0 shim array (32-channel) to perform wave-encoding simultaneously and synergistically at different wave frequencies during the acquisition. We demonstrate that mf-wave can enable ~20-fold acceleration with an acceptable g-factor noise penalty at 3T in vivo.
Introduction
The application of sinusoidal encoding during data acquisition was first proposed in Bunched Phase Encoding
1 with G
y sinusoidal encoding, which was then extended to G
y and G
z sinusoidal encodings with a quarter-cycle offset to
create corkscrew sampling in wave-CAIPI
2. Subsequently, sinusoidal encoding was performed using second-order nonlinear shim gradients in FRONSAC
3 and more recently using local B
0 field coil arrays in Spread-spectrum MRI
4. All these approaches have been shown to improve acceleration, however, the synergistic combination of them has not been investigated. In this work, we propose a multi-frequency wave-encoding method (mf-wave) that exploits both linear gradients and a combined RF and B
0 shim array
5 of single-turn loops on a close-fitting helmet to perform wave-encoding simultaneously and synergistically at different wave frequencies during the acquisition. Due to their low inductance, local shim coils can be driven at much higher frequencies than the scanner gradients. Here, the additional high-frequency wave-encoding from our 32-channel shim array enhances the voxel-spreading effects and broadens the k-space coverage achieved via the standard low-frequency wave-encoding from the gradients. mf-wave was demonstrated to reduce g-factors in high-acceleration in both 2D and SMS acquisitions.
Method
Theory: Fig1.A shows the hardware for generating the time-varying field needed for mf-wave-encoding. Fig1.B shows the sequence diagram for mf-wave where wave encodings are performed on both the gradients and shim array at different frequencies. Fig1.C shows the resulting k-space trajectories in both 2D and 3D/SMS acquisitions. Here slow-frequency wave-encoding is applied on the gradient hardware to provide large k-space wave modulation while high-frequency wave-encoding is performed with the shim array to further broaden the k-space coverage of this modulation. The slow frequency was chosen for the gradients as it works well with the slew and
Peripheral Nerve Stimulation (PNS) limits, while the higher frequency is applied on the shim array as shim fields are not restricted by these constraints. As an initial optimization along this direction, we simulated mf-wave encoding at various frequency combinations and found that the high-frequency shim-wave-encoding combined with lower-frequency gradient-wave-encoding can provide the best performance in this case (Fig. 2).
Image reconstruction: We generalized the forward-model of the voxel-spreading effects along the readout direction in wave-encoding from Bilgic et al
2 to mf-wave-encoding as:
$$W = F^{-1}(F\odot{E})m$$
Where $$$W$$$ is the vector along readout of the voxel-spreading image at the location $$$(y,z)$$$, $$$F$$$ is the 1D discrete Fourier transform matrix and $$$F^{-1}$$$ is the 1D inverse Fourier transform matrix, $$$E$$$ is a spatial-varying encoding matrix and $$$m$$$ is the corresponding row of the image data to be reconstructed at location $$$(y,z)$$$. Analogous to conventional wave-CAIPI reconstruction, this formulation makes the reconstruction process separable in the phase-encoding direction for mf-wave-encoding, thus allowing each set of collapsed $$$(y,z)$$$ rows to be solved independently, using e.g. lsqr in MATLAB. G-factor can also be calculated analogously.
In-vivo experiments
MATLAB’s quadprog function was used to find optimal shim currents for generating a close approximation to the desired wave encoding field, subject to practical current constraints (The chosen maximum of I
max= 8A per channel, while beyond the capabilities of the existing 32ch coil (4A max), could be realized in future designs using thicker wiring). The fields generated from this calculation are used in our g-factor simulation for optimizing the mf-wave-encoding.
We are currently modifying our shim amplifier controller to
create time-varying output currents necessary for dynamic wave-encoding. However, to verify the efficacy of our technique, we retrospectively construct mf-wave-encoding data from in-vivo shim-encoded scans in a “static” manner as shown in Figure 3.
The imaging parameters were: GRE acquisition, FOV = 220x220x150 mm; resolution = 1.7x1.7x2 mm, TR/TE = 50/30ms, BW = 100Hz, FA = 18°,10ms readout. For gradient-wave-encoding, 6-cycles were employed with G
max = 7mT/m and S
max = 150 T/m/s. For shim-wave-encoding, an approximately linear shim field was generated to approximate cos(wt)x and cos(wt)z modulations (48 cycles, 4.8kHz). A 6x readout oversampling was used to increase FOV
x and capture the voxel spreading effects.
Results
Figure 4 shows the reconstructions and corresponding 1/g factor maps of the acquired in-vivo 2D data at Rinplane =10-fold. At this extreme acceleration, the standard acquisition produces an unusable image, while wave-encoding produces images with substantial noise. The use of mf-wave-encoding produces images with significantly less noise, at a maximum g-factor that is ~2x lower than with wave-encoding.
Figure 5 shows the reconstructions and corresponding 1/g factor maps of the acquired in-vivo SMS data at Rinplane x MB =8 x 3. Again, the images from the standard acquisition are unusable. Here wave-CAIPI provides large improvements, but still results in poor reconstruction, while mf-wave further improves upon this, with some artifacts remaining likely from minor deviations in the encoded field vs. theoretical field used in the reconstruction. Discussion and Conclusion
mf-wave-encoding extends the wave-encoding technique to simultaneous wave-encoding on both gradients and multi-coil shim hardware. Preliminary exploration in this direction so far has shown promising results that point to the potential for significant increases in acceleration capability using this approach, with further optimization and dynamic experiments planned for future work. A further extension to include additional hardware components such as non-linear shim encoding gradients is also feasible and could potentially provide even further acceleration capability.Acknowledgements
This work is supported in part by the National Natural Science Foundation of China (No: U1809204, 61525106, 61427807, 61701436), by the National Key Technology Research and Development Program of China (No: 2017YFE0104000, 2016YFC1300302), and by Shenzhen Innovation Funding (No: JCYJ20170818164343304, JCYJ20170816172431715), by NIH R01EB019437, R01EB020613, R01 MH116173, U01EB025162 and NIH NIBIB R00EB021349. Jinmin Xu was supported by the China Scholarship Council for 2 years at Massachusetts General Hospital.
References
- Moriguchi, Hisamoto,
and Jeffrey L. Duerk. "Bunched phase encoding (BPE): a new fast data
acquisition method in MRI." Magnetic Resonance in Medicine: An Official
Journal of the International Society for Magnetic Resonance in Medicine 55.3
(2006): 633-648.
- Bilgic, Berkin, et al.
"Wave-CAIPI for highly accelerated 3D imaging." Magnetic resonance in
medicine 73.6 (2015): 2152-2162
- Wang, Haifeng, et al.
"Fast rotary nonlinear spatial acquisition (FRONSAC) imaging."
Magnetic resonance in medicine 75.3 (2016): 1154-1165.
- Scheffler, Klaus, et
al. "Spread‐spectrum magnetic resonance imaging." Magnetic resonance
in medicine 82.3 (2019): 877-885.
- Stockmann, Jason P., et
al. "A 32‐channel combined RF and B0 shim array for 3T brain
imaging." Magnetic resonance in medicine 75.1 (2016): 441-451.