Bastien Guerin1,2, Eugene Milshteyn1,2, Lawrence L Wald1,2, and Jason Stockmann1,2
1Martinos Center for Biomedical Imaging, Massachusetts General Hospital, Charlestown, MA, United States, 2Harvard Medical School, Boston, MA, United States
Synopsis
We compare the RF/B0 shim array composite pulses of Rudrapatna [1] to kT-point and spoke
pulses for flip-angle (FA) uniformization using a birdcage coil at 7T. Using 3 kT-points,
it is possible to obtained highly uniform flip-angle distributions in the brain
within 3ms, as long as the kT-points locations are optimized. The RF/B0 shim array
pulses perform less well than this optimized 3-kT-points strategy for
non-selective flip-angle mitigation, but better than optimized 3-spoke
pulses for flip-angle mitigation in a slice. The DOFs provided by shim
array coils could prove invaluable for integrated B0 shimming and flip-angle
uniformization at 7T.
Introduction
Rudrapatna et
al [1] proposed a composite pulse approach that leverages the additional
degrees-of-freedom provided by multi-coil B0 shim arrays (SA) in order to create
uniform flip-angle excitations at 7T. The approach interleaves RF sub-pulses (rect
for non-selective, sinc for slice-selective) and periods during which the shim-array
is driven alone thus dephasing spins by different amounts in different regions
of the field-of-excitation. They show that the ideal target phase pattern
between the RF sub-pulses of a two-subpulse experiment (RF-SA-RF) can be derived analytically. Thus, they drive
the SA so as to best approach this distribution. We show here that the RF-SA-RF
experiment cannot yield inversion pulses, and that an RF-SA-RF-SA-RF (three sub-pulses)
excitation must be used instead. In this case, an analytical expression for the target
phase does not exist, the problem is instead cast as a joint optimization of
the RF sub-pulse weights and shim currents. We implement this approach in a
realistic head phantom and compare the performance of the RF/B0 shim array composite
pulses to kT-points [4] and spoke [5] pulses for non-selective and
slice-selective FA uniformization at 7T, respectively.Methods
Shim array: We
use a 7T B0 shim array with 31 loops supporting both RF and DC current (“AC/DC”
coil) patterned on a close-fitting helmet. The B0-field maps
(Hz/Amp) of all channels were measured in lengthy scans in a silicone-oil phantom
using the vendor-provided field-mapping sequence (2-mm iso., dual-echo, ΔTE=1.02ms). These maps are independent of the load and need to
be acquired only once. RF transmission is by a birdcage coil.
Pulse optimization: We cast the RF/B0 shim array pulse design
as an optimization problem. Specifically, the forward model is a Bloch
simulation in the spinor domain [3]. We jointly optimize the RF sub-pulse
weights and shim currents subject to maximum RF amplitude (400V), maximum shim
current (3Amp per channel) and maximum total current (30Amp) hard constraints using
Matlab’s fmincon. We constrain the phase created by the SA between RF sub-pulses
to below 100Hz in order to avoid large rotations of the magnetization that may
lead to local minima. We initialize the pulse design using manual (first RF set
to the target flip-angle, all shim currents to zero) and random initial guesses.
Numerical evaluation: We used B1+ and B0 field maps acquired
in a human subject at 7T in all the numerical evaluations. We designed 90
degrees non-selective (RF-SA-RF), 180 degrees non-selective (RF-SA-RF-SA-RF) as
well as 20 degrees slice-selective (RF-SA-RF) pulses using the RF/shim array
method. We also designed standard kT-points [4] and spoke pulses [5] for
comparison. kT-point and spokes were optimized using a greedy approach whereby sub-pulses
were added one at a time, using the previous estimate of the pulse as the initial
guess for the placement of the next sub-pulse.
Phantom measurements: We performed RF-SA-RF experiments in a
head phantom at 7T. The birdcage’ B1+ map was measured using the product
saturated-TFL sequence [6] (TR=10 seconds, TE=2ms, resolution=3mmx3mmx5mm). The
B0 map was measured using the product GRE field-mapping sequence (same resolution).
Maps were interpolated in a common reference frame along with the shim array element
field maps. We inserted the SA composite pulses in a FLASH sequence and assessed
the improvement in contrast weighting uniformity on PD-weighted images (flip-angle
20 or 90 degrees, TR=100ms, TE=10ms). The SA was triggered by the sequence. Results & discussion
Fig. 1 shows that the RF-SA-RF strategy
seems to outperform the optimized 2-spoke excitation for this particular
phantom and slice position (10 degrees slice-selective excitations), which can
be seen both on the flip-angle maps and the PDw-FLASH images. Fig. 2 shows a
comparison of the target phase method [1]
and our joint RF/shim current optimization for a non-selective excitation (90
degrees). The joint optimization performs better, especially in regions with significant
B0 inhomogeneities. The waveform solutions created by both approaches are quite different,
indicating different mechanisms.
Fig. 3 shows an RF-SA-RF-SA-RF pulse for non-selective
inversion of the magnetization. Good homogenization is obtained, which is not
possible using only two RF sub-pulses (RF-SA-RF). Figs. 4 & 5 show that the
performance of the RF/shim array strategy is slightly inferior to that of the
best kT-point pulse, although this requires carefully choosing the kT-point locations
(greedy optimization). In contrast, the slice-selective RF/shim array strategy performed
better than the duration-matched bipolar optimized 3-spoke pulse.Conclusion
The RF/B0 shim array flip-angle mitigation strategy proposed by
Rudrapatna et al [1] seems
to outperform traditional spoke pulses for slice-selection, but not kT-points
when those are properly positioned. This is due to the fact that the degrees-of-freedoms
of the shim array can be used to create complicated phase patterns in 2D, but
less so in 3D (this is similar to the superior performance of slice-by-slice B0
shimming compared to global shimming). For magnetization inversion and, we hypothesize,
refocusing, a 3-sub-pulse strategy is required (RF-SA-RF-SA-RF). Flip-angle
mitigation using shim arrays is potentially simpler than parallel transmission
as the safety of this technique is the same as that of the birdcage coil and
the SA field maps need only be acquired once (same for all subjects).Acknowledgements
JPS: NIH NIBIB R00EB021349; BG: NIH NIBIB R00-EB019482References
[1] Umesh
Rudrapatna, S., et al. (2016).
"Dynamic multi‐coil
tailored excitation for transmit B 1 correction at 7 Tesla." Magnetic
Resonance in Medicine 76(1): 83-93.
[2] Stockmann, J. P., et al. (2016). "A
32‐channel
combined RF and B0 shim array for 3T brain imaging." Magnetic Resonance
in Medicine 75(1): 441-451.
[3]
Pauly, J., et al. (1991). "Parameter relations for the Shinnar-Le Roux
selective excitation pulse design algorithm [NMR imaging]." Medical
Imaging, IEEE Transactions on 10(1): 53-65.
[4] Cloos, M., et
al. (2011). "kT‐points:
Short three‐dimensional
tailored RF pulses for flip‐angle
homogenization over an extended volume." Magnetic Resonance in Medicine
67(1): 72-80.
[5] Saekho, S.,
et al. (2006). "Fast kz three-dimensional tailored radiofrequency pulse
for reduced B1 inhomogeneity." Magnetic Resonance in Medicine 55(4):
719-724.
[6] Fautz, H., et
al. (2008). B1 mapping of coil arrays for parallel transmission. ISMRM.