Difei Wang1, Tony Stöcker1,2, and Rüdiger Stirnberg1
1German Center for Neurodegenerative Diseases (DZNE), Bonn, Germany, 2Department of Physics and Astronomy, University of Bonn, Bonn, Germany
Synopsis
By comparison to a gold standard multiparametric
mapping (MPM) protocol at 3T, this study shows that multi-echo 3D-EPI with
highly segmented CAIPIRINHA sampling can yield whole-head T1, PD*,
MTsat and R2* maps of high quality at 1mm
isotropic resolution in less than 3 minutes scan time. Even less than 1 minute
of single-echo 3D-EPI is sufficient to yield accurate quantitative T1,
PD* and MTsat maps. If necessary, SNR can be improved by
including repeated EPI measurements. Optional motion- and distortion-correction
across measurements may further improve results. Motion-robust MPM thus renders
assessing quantitative parameter maps in clinical or population studies
feasible.
Introduction
Multiparametric mapping (MPM) provides
quantitative maps of T1, effective proton density (PD*),
magnetization transfer (MTsat) and T2* based
on multi-echo (ME) T1w, PDw, and MTw images typically acquired using a 3D fast
low angle shot (FLASH) sequence1. Highly segmented 3D ME Echo Planar
Imaging (EPI) can be applied like ME-FLASH with an EPI factor to
reduce acquisition time (TA). It can provide longer TEs at higher
signal-to-noise ratio efficiency (SNR/√TA) and decreased motion
sensitivity2. Our study investigates whether such a 3D-EPI approach with
CAIPIRINHA3 sampling is suitable for MPM.Methods
All images were acquired on a Siemens
(Healthineers) Skyra 3T system using a 32 channel head receive coil. T1w, PDw and
MTw images were acquired using whole-head ME-FLASH protocols as recommended by
Callaghan et al. (vendor sequence) with 6 TEs up to 14.76ms at 1mm isotropic
interpolated resolution4,5 (experiment A). Corresponding ME-EPI
scans with 4 TEs up to 28.62ms were acquired with repeated measurements (n=4-6)
to approximate the respective FLASH TAs (experiment B). A prolonged excitation
pulse (~2.4ms) was used for fat suppression6. Finally, single-TE EPI
scans were acquired that allowed for TR and flip angles (FA) matched to the
FLASH experiment (experiment C). Each EPI scan included 10s of preparation time
(external phase correction2, steady-state “dummies”, 36x36 FLASH
autocalibration scans (ACS)7). All MPM imaging parameters are
summarized in Tab. 1. Additionally, a fast 3D-EPI-based Look-Locker (LL) scan8
with otherwise identical image parameters was acquired in 13 minutes for T1
estimation (TE=4.2ms, TR=16.0ms, EPI factor=5, turbo factor=16, 18 TIs between
143 and 4495 ms, FA=7°, 2 averages).
Multiparametric maps were calculated within the
hMRI toolbox9. A 3DREAM B1+ map, acquired at
4mm isotropic resolution (TA=6.9s)10 was utilized for flip angle
correction in MPM and for Look-Locker T1. A simultaneously acquired
B0 map was used to account for reduced off-resonant flip angles
according to Freed et al.11.Results
Fig. 1 shows representative sagittal,
coronal and axial slices of all T1w scans at TE=4.92ms. Fig. 2 shows
corresponding multiparametric maps. TA-matched ME-EPI results in superior
quality compared to FLASH, while overall the estimated T1 appears
smaller. Notably, more ventricle (CSF) voxels exceed 100 p.u. in the EPI PD*
maps. The T2* map is clearly improved with ME-EPI
(reduced ringing and noise) except for frontal brain regions, where magnetic
susceptibility gradients are strong. A single ME-EPI measurement per contrast
(n=1) leads to noisier but accurate MT, PD* and T1 maps.
T2* is still largely superior to FLASH. Fig. 3 shows
whole-brain histograms of MTsat with clear modes for CSF/WM/GM. The
FLASH histogram is shifted towards higher values and the peaks of ME-EPI (n=1)
are wider than of TA-matched ME-EPI. However, all scans result in the expected
0:1:2.2 ratio of MTsat modes12. Fig. 4 shows bar plots of
PD* (a), T2* (b) and T1 (c) in
different regions of interest (ROI) from different scans (including LL-T1).
Fig. 5 (a,b) confirm overall agreement across scans except for increased EPI PD*
estimates in the ventricles. In CSF, LL underestimates T1 (lacking
sufficiently long TIs). In all other ROIs, there is a greater T1
agreement between EPI and LL than between FLASH and LL.Discussion
Our study has shown that 1mm isotropic
multi-echo T1w, PDw and MTw measurements, altogether acquired in only 3 minutes
using 3D-EPI, are sufficient to estimate high quality T1, PD*,
MTsat and R2* maps. A high readout bandwidth
combined with high segmentation assures minimal geometric distortions. Scan
time could even be further reduced without sacrificing SNR, e.g. by sharing one set of ACS. If necessary, SNR can be improved by including repeated EPI measurements,
potentially varying across contrasts. This provides superior parameter maps and
the option of motion- and distortion-correction across measurements, as
recently shown for QSM and T2*2.
Overall, we observed slightly higher T1
results with FLASH MPM compared to all EPI MPM and the Look-Locker reference.
Increased TR and FA for ME-EPI does not appear to conflict with the “short-TR”
and “small-FA” approximations of MPM13. A violation should result in
increased T1 estimates. Although not observed here, the effect might
be present but masked by another systematic difference between FLASH and EPI.
We observed smaller MTsat values with EPI MPM. This may be due to
empirically set MT flip angles12. Also, we find that PD*
estimates tend to exceed 100 p.u. in CSF when using EPI. Future investigations
will explore how MPM parameters depend on excitation bandwidth (long water
excitation pulse), larger TRs and FAs (MPM approximations), increased susceptibility artifacts, and physiological noise in EPI MPM.Conclusion
By comparison to a gold standard MPM protocol at
3T we have validated that 3D-EPI with segmented CAIPIRINHA sampling yields
quantitative multiparametric maps of high quality in short scan time. Accurate
whole-head maps at 1mm isotropic resolution can be obtained from T1w, PDw and
MTw multi-echo 3D-EPI scans acquired in less than 3 minutes. If T2*
can be sacrificed, even less than 1 minute of single-echo 3D-EPI is sufficient
to yield accurate quantitative T1, PD* and MTsat maps.
Motion-robust MPM renders assessing quantitative parameter maps in clinical or
population studies feasible, where they offer great potential as biomarkers,
e.g. for neurodegenerative diseases.Acknowledgements
No acknowledgement found.References
1 Weiskopf N, Suckling J, Williams G,
Correia MM, Inkster B, Tait R et al.
Quantitative multi-parameter mapping of R1, PD*, MT, and R2* at 3T: a
multi-center validation. Frontiers in
Neuroscience 2013;7(June):95.
2 Stirnberg R, Deistung A,
Reichenbach J, Stöcker T. Accelerated quantitative susceptibility and R2*
mapping with flexible k-t-segmented 3D-EPI. In: Proceedings of the
International Society of Magnetic Resonance in Medicine, 2018.
3 Breuer FA, Blaimer M, Mueller MF,
Seiberlich N, Heidemann RM, Griswold MA et
al. Controlled aliasing in volumetric parallel imaging (2D CAIPIRINHA). Magnetic Resonance in Medicine
2006;55(3):549–56.
4 Callaghan MF, Lutti A, Ashburner J,
Balteau E, Corbin N, Draganski B et al.
Example dataset for the hMRI toolbox. Data
in Brief 2019;25:104132.
5 Callaghan MF, Lutti A, Ashburner J,
Balteau E, Corbin N, Draganski B et al.
MPM_protocol_vendor_seq_SIEMENS_settings.pdf. https://hmri-group.github.io/hMRI-toolbox/#hmri-example-data-and-mri-protocol. Accessed September 9, 2019.
6 Stirnberg R, Brenner D, Stöcker T,
Shah NJ. Rapid fat suppression for three-dimensional echo planar imaging with
minimized specific absorption rate. Magnetic
Resonance in Medicine 2016;76(5):1517–23.
7 Ivanov D, Barth M, Uludağ K, Poser
BA. Robust ACS acquisition for 3D echo planar imaging. In: Proc Intl Soc Mag
Reson Med 23, 2015.
8 Stirnberg R, Dong Y, Bause J, Ehses
P, Stöcker T. T1 Mapping at 7T Using a Novel Inversion-Recovery Look-Locker
3D-EPI Sequence. In: Proceedings of the International Society of Magnetic
Resonance in Medicine, 2019.
9 Tabelow K, Balteau E, Ashburner J,
Callaghan MF, Draganski B, Helms G et al.
hMRI – A toolbox for quantitative MRI in neuroscience and clinical research. NeuroImage 2019;194(January):191–210.
10
Ehses P, Brenner D, Stirnberg R, Pracht ED,
Stöcker T. Whole-brain b1-mapping using three-dimensional dream. Magnetic Resonance in Medicine
2019;82(3):924–34.
11 Freed DE, Hürlimann MD, Scheven
UM. The equivalence between off-resonance and on-resonance pulse sequences and
its application to steady-state free precession with diffusion in inhomogeneous
fields. Journal of Magnetic Resonance
2003;162(2):328–35.
12 Helms G, Dathe H, Kallenberg K,
Dechent P. High-resolution maps of magnetization transfer with inherent
correction for RF inhomogeneity and T1 relaxation obtained from 3D FLASH MRI. Magnetic Resonance in Medicine
2008;60(6):1396–407.
13 Helms G, Dathe H, Dechent P.
Quantitative FLASH MRI at 3T using a rational approximation of the Ernst
equation. Magnetic Resonance in Medicine
2008;59(3):667–72.