Eddy Solomon1, Houchun H. Hu2, Kai Tobias Block1, Daniel K. Sodickson1, and Hersh Chandarana1
1Radiology, New York University School of Medicine, New York, NY, United States, 2Radiology, Nationwide Children's Hospital, Columbus, OH, United States
Synopsis
Inversion-recovery
3D T1 gradient echo sequences are commonly used in brain examinations for their
excellent gray-/white-matter contrast. However, prominent motion artifacts can
arise during lengthy Cartesian k-space sampling (typically 5-7 minutes) if the
patient is not able to hold still, as is often the case for pediatric or
elderly patients. Here, we present an alternative based on radial
stack-of-stars imaging and show that comparable image contrast can be achieved,
with lower sensitivity to head motion. Moreover, we demonstrate how the radial
acquisition scheme can be utilized for additional retrospective motion
correction to further improve robustness without increasing acquisition time.
INTRODUCTION
Acquisition of T1-weighted images is an essential
component of routine brain MRI. Three-dimensional Magnetization-Prepared
Rapidly Acquired Gradient Echo (MP-RAGE) is a commonly used sequence that
provides excellent gray-/white-matter contrast at high resolution (1). However,
like any sequence using Cartesian phase-encoding, MP-RAGE is prone to
ghosting-like motion artifacts that can impair the diagnostic value of the
images if the patient cannot hold still during the examination, which poses a
particular problem in pediatric patients or elderly patients with tremor. As a
result, scans must be repeated or, in the case of pediatric imaging, must be
performed during sedation. Recently, it has been demonstrated that non-Cartesian
k-space acquisition schemes, such as radial sampling, provide higher inherent
robustness to motion (2), and it has been shown that using such sequences
provides value for imaging of the head and brain (3,4). The aim of this work
was to develop a variant of the MR-RAGE sequence with radial sampling of
k-space, named MP-RAVE, and to evaluate its advantages in healthy human
volunteers during head motion.METHODS
All data were
acquired on a 3T Prisma system (Siemens Healthcare, Erlangen, Germany) using a 20-channel
head coil array. The study protocol included a conventional T1-weighted MP-RAGE
sequence and the radial counterpart MP-RAVE, which has been derived from a
previously described radial stack-of-stars 3D GRE sequence (RAVE) (5) by
integrating an inversion recovery pulse that is played out once for every stack
of radial views with same angle. Common imaging parameters were TR/TI/TE=2200/900/3.2ms,
1.0mm spatial resolution, no fat suppression, and 224 slices. For MP-RAVE, 400
radial views were acquired and a centric acquisition was used along the slice
direction. Images were reconstructed by standard gridding (NUFFT). The scan time
of MP-RAGE was 4min with 2x GRAPPA acceleration whereas the initial version of MP-RAVE
did not use parallel imaging, resulting in double the acquisition time. Both
methods were tested in six healthy volunteers.
Motion-correction algorithm: While radial data acquisition is inherently
more robust to motion and does not show the MR-typical ghosting artifacts, it
is not completely immune to motion. However, due to the geometry of the radial
trajectory, which covers k-space center with each radial view, it is possible
to reconstruct time-resolved subframes from the acquisition using an iterative
reconstruction technique like GRASP (6), which utilizes compressed sensing with
a temporal total-variation constraint to calculate images from the highly
undersampling subframe data. This series of dynamic images can then be used for
estimating intra-acquisition head motion via image-based registration,
specifically rotation and translation. Once the motion dynamics has been estimated,
the k-space trajectory can then be corrected for rigid motion and a final
full-resolution motion-corrected image can be reconstructed. To demonstrate
this possibility, a volunteer scan was performed under head motion using the
RAVE sequence. Each subframe “navigator” image consisted of four radial views, sensing
motion every ~0.9 seconds. The motion-correction algorithm scheme is illustrated
in Figure 4. RESULTS AND DISCUSSION
T1-weighted anatomical brain images were
acquired by conventional MP-RAGE and the proposed
MP-RAVE sequence, showing comparable T1-weighted contrast and image quality
(Fig. 1). As a direction to further reduce the relatively longer scan time of
MP-RAVE, reconstruction of MP-RAVE data with different undersampling factors is
shown in Fig. 2, demonstrating that acceptable image quality is achieved even
at 50% undersampling. The higher motion robustness achieved with MP-RAVE compared to MP-RAGE in
presence of continuous head motion is shown in Figure 3. None of the sequences are
completely immune to motion. However, while MP-RAVE images are somewhat blurred,
MP-RAGE images show prominent ghosting artifacts along the phase-encode
direction that can potentially obscure pathologies. The latter arise because
the phase-encoding scheme is disturbed by motion-induced phase modulation,
which is avoided with radial sampling. In addition, MP-RAVE’s higher robustness
may also be attributed to oversampling of
the k-space center, which has a motion-averaging effect. To overcome residual
motion blurring with MP-RAVE, we have tested a motion-correction algorithm (Fig.
4) in a volunteer (Fig. 5). It can be seen that the correction algorithm is
able to reduce the amount of blurring and can recover anatomical information in
various regions of the brain, allowing for better delineation of specific structures
(see yellow arrows).CONCLUSION
This work describes a novel IR-prepared T1-weighted
3D GRE sequence with radial stack-of-stars sampling (MP-RAVE) as a potential
alternative to the widely used MP-RAGE sequence. It has been shown that MP-RAVE
can generate comparable image quality and contrast to conventional MP-RAGE but
with lower sensitivity to head motion. To reduce the scan time, undersampling
techniques can be utilized, such as iterative constrained reconstruction. Moreover,
to further improve the robustness to motion, retrospective motion correction
can be applied without adding to scan time or requiring additional navigator
information.Acknowledgements
We acknowledge support from NIH grant P41
EB0171813 and R01 5R01EB018308.References
- Brant-Zawadzki M, Gillan GD, Nitz WR. MP RAGE: a three-dimensional, T1-weighted, gradient-echo sequence--initial
experience in the brain. Radiology. 1992 Mar;182(3):769-75.
- Chandarana
H, Feng L, Block TK, Rosenkrantz AB, Lim RP, Babb JS, Sodickson DK, Otazo R. Free-breathing
contrast-enhanced multiphase MRI of the liver using a combination of compressed
sensing, parallel imaging, and golden-angle radial sampling. Invest Radiol.
2013 Jan;48(1):10-6.
- Hu HH,
Benkert T, Jones JY, McAllister AS, Rusin JA, Krishnamurthy R, Block KT. 3D
T1-weighted contrast-enhanced brain MRI in children using a fat-suppressed
golden angle radial acquisition: an alternative to Cartesian inversion-recovery
imaging. Clin Imaging. 2019 May - Jun;55:112-118
- Kecskemeti S, Samsonov A, Velikina J, Field AS,
Turski P, Rowley H, Lainhart JE, Alexander AL. Robust Motion Correction Strategy
for Structural MRI in Unsedated Children Demonstrated with Three-dimensional
Radial MPnRAGE. Radiology. 2018 Nov;289(2):509-516.
- Block KT, Chandarana H, Milla S, Bruno M, Mulholland T, Fatterpekar G, Hagiwara M, Grimm R,
Geppert C, Kiefer B, Sodickson DK.
Towards Routine Clinical Use of Radial Stack-of-Stars 3D
Gradient-Echo Sequences for Reducing Motion Sensitivity. Journal of the Korean
Society of Magnetic Resonance in Medicine 18 (2), 87-106.
- Feng L, Axel L, Chandarana H, Block KT, Sodickson DK, Otazo R. XD-GRASP:
Golden-angle radial MRI with reconstruction of extra motion-state dimensions
using compressed sensing. Magn Reson Med. 2016 Feb;75(2):775-88.