Wei Liu1 and Kun Zhou1
1Siemens Shenzhen Magnetic Resonance Ltd., Shenzhen, China
Synopsis
As commonly
used for intracranial vasculature, 3D TOF usually requires long acquisition
time. In this study, we implemented a 3D-iEPI sequence with partial
flow compensation, combined with partial Fourier acquisition to further reduce the
flow artifacts. In specific, each interleave is sequentially acquired twice with
alternating readout polarities to reduce the systematic inconsistencies between
odd and even echoes. We explored the feasibility of such a sequence for fast
intracranial TOF-MRA and demonstrated that the proposed sequence can reduce the
acquisition time by approximately a factor of 2 with comparable vasculature
depiction to 3D-GRE, which is promising for future applications.
INTRODUCTION
As widely used in non-contrast MR
angiography (MRA) in clinical setting, the Time-of-flight (TOF) is
conventionally implemented with the 2D or 3D-GRE sequence, in which the 3D-GRE
is most commonly used for intracranial vasculature due to well depiction of
arterial trees1. The background signal from stationary tissue is
suppressed by rapid slab-selective RF excitation pulses, while the venous
signal is suppressed using a saturation band on the venous side of the imaging
volume. By reducing the number of RF pulses for a saturation band, segmented
TOF has been proposed to increase the acquisition efficiency and alleviate the
SAR problem with equivalent contrast and suppression effects in venous regions
preserved as in the conventional methods2. Alternatively, the
multi-echo acquisition such as EPI can achieve similar or even higher scan
efficiency, as compared to the segmented TOF. However, flow artifacts are more
complicated by EPI than GRE3, therefore, improvements in EPI based
TOF-MRA sequence can be essential for reduction of flow artifacts. To our
knowledge, there are no reports on TOF-MRA of intracranial vasculature with
3D-iEPI. Therefore, in this study, we implemented a first order
gradient nulling (GMN) based partial flow compensation in 3D-iEPI4,
combined with partial Fourier acquisition in phase direction, for reduction of
the TE and flow artifacts5. We demonstrated the application of the
proposed sequence on intracranial TOF-MRA, providing
a substantial reduction in scan time compared to standard 3D-GRE, whilst maintaining
well vasculature depiction.METHODS
In the
proposed sequence (Fig.1), two shots of EPI acquisitions with partial coverage
of k-space and alternating readout polarities were acquired sequentially to reduce
phase oscillates in odd and even echoes and minimize the motion influence. Furthermore,
compared with our previous flow compensation implementation4, the
prephaser and partition encoding gradients were moved next to the readout for further reduction of flow artifacts.
All
measurements were performed on a commercial 1.5T scanner (MAGNETOM Aera,
Siemens Healthcare GmbH, Erlangen, Germany) equipped with a 20-channel head/neck
coil. Experimental data was obtained from a healthy volunteer using a standard
3D-GRE based TOF, a prototype flow compensated 3D-iEPI based TOF without and with alternating readout.
The imaging parameters were shown in Table 1. Standard multiple
overlapping thin slab acquisition (MOTSA)6 was used in both
sequences. After data collection, all MRA images were displayed using a maximum
intensity projection (MIP) algorithm.RESULTS
The
preliminary in-vivo results show that, with flow compensation and partial
Fourier acquisition, the 3D-iEPI sequence provides significant acquisition time
reduction, as well as comparable vasculature depiction, compared to
conventional TOF-MRA (Fig.2). Both 3D-GRE (Fig.2a) and 3D-iEPI (Fig.2b and
Fig.2c) show comparable intracranial vasculature representation. In addition,
no obvious flow artifacts were observed in 3D-iEPI with flow compensation and
partial Fourier acquisition. That partial Fourier acquisition can not only alleviate
the phase encode flow artifacts but also reduce the TE to be comparable with
that in 3D-GRE, which can increase the vessel intensity7. Although
the SNR loss in the magnitude images of 3D-iEPI sequence is inevitable, it has less
effect on vessel signals and thus, the corresponding MIP images are similar to
those from the 3D-GRE. It should be noticed that, both ghost artifacts and
background suppression can be improved if alternating readout scheme applied
(Fig. 2c), the latter one may be due to the reduced motion effect from the inherent
short-term average in such alternating readout scheme. Moreover, some arteries were better resolved in TOF images by 3D-iEPI sequence, e.g. external carotid artery
branch.DISCUSSION
In order to
reduce the flow artifacts in 3D-iEPI, we implemented a partial flow
compensation scheme to fully flow compensated the center echo in each shot, and
utilized partial Fourier in phase encoding direction to further reduce the flow
artifacts and minimize the TE. No obvious distortion and blurring artifacts
were observed in the in-vivo experiments, which can be due to the short echo
train used in the 3D-iEPI. By taking advantage of the highly efficient
multi-echo acquisition, the proposed 3D-iEPI based TOF-MRA can provide a
significant reduction in both SAR and san time, while maintaining similar
vasculature representation compared to conventional 3D-GRE. However, the
pulsatile or disordered flow can lead to ghosting or signal loss in the vessel due
to B0 inhomogeneityl7, which can be more severe in EPI sequence.
Less EPI factor in combination of partial Fourier acquisition can alleviate
such artifacts but probably produce slightly more blurring and ringing.
Evaluation of the diagnostic performance of 3D-iEPI TOF-MRA in patients would
be needed in future study.CONCLUSION
We demonstrated that a flow
compensated 3D-iEPI with partial Fourier acquisition allows approximate 2-fold
reduction in acquisition time with comparable artery angiogram to 3D-GRE, which
is promising for TOF-MRA applications.Acknowledgements
No acknowledgement found.References
1. Hartung MP, et al. Magnetic resonance angiography: current status and future directions. J
Cardiovasc Magn Reson 2011; 13:19.
2. Zhang ZH, et al. Segmented TOF at 7T MRI: Techniques and clinical applications. Magn Reson
Imaging 2015; 33:1042-1050.
3. Butts K, et
al. Analysis of flow effects in echo-planar imaging. J. Magn. Reason Imaging
1992; 2:285-293.
4. Liu W, et
al. 3D Flow Compensated Interleaved EPI for a Fast High-Resolution
Susceptibility-Weighted Imaging at 1.5T. Proc. Intl. Soc. Mag. Reason. Med 2019; 27:3326.
5. Luk Pat GT, et al. Reducing flow artifacts in echo-planar imaging. Magn Reson Med 1997; 37: 436-447.
6. Parker DL, et al. MR angiography by multiple thin slab 3D acquisition. Magn Reson Med
1991; 17:434-451.
7. Drangova M, et al. Artifacts and signal loss due to flow in the presence of B0 inhomogeneity.
Magn Reson Med 1996; 35:126-130.