0530

Synthetic T1rho dispersion imaging for improved myocardial tissue characterization using dispersion reconstruction
Maximilian Gram1,2, Daniel Gensler1,3, Patrick Winter1,2, Michael Seethaler2,3, Peter Jakob2, and Peter Nordbeck1,3
1Department of Internal Medicine I, University Hospital Würzburg, Würzburg, Germany, 2Experimental Physics 5, University of Würzburg, Würzburg, Germany, 3Comprehensive Heart Failure Center (CHFC), University Hospital Würzburg, Würzburg, Germany

Synopsis

T dispersion imaging is a very time-consuming process because full T-mapping at different spin-lock amplitudes is required. Due to this issue, investigation of T dispersion is hardly feasible in the limited measurement time of a small animal experiment. In this work, we present a novel approach for the rapid measurement of cardiac T dispersion called dispersion reconstruction. With our new concept a T dispersion image is generated by only acquiring a fraction of the required mapping data. Phantom and in vivo experiments confirm the applicability of our new method as part of a conventional protocol for small animal studies.

Introduction

In the last decade, cardiac MRI became the imaging modality of choice for myocardial function measurements as well as locally resolved tissue characterization. Here, quantitative methods like T1 and T2 mapping are useful tools for a precise tissue characterization. Recently, numerous studies explored the alternative contrast mechanism of the spin-lattice relaxation time T [1,2,3]. Due to the high sensitivity to low frequency processes at the molecular and cellular level, native T-mapping might be superior for accurate tissue characterization. Moreover, T gives the possibility for dispersion measurements by manipulation of the effective spin-lock (SL) pulse amplitude, which enables additional contrast mechanisms. However, due to cardiac and respiratory motion, the generation of single myocardial T-maps is very time consuming. In consequence, the investigation of myocardial T dispersion is not feasible in the limited measurement time of a common in vivo experiment.
In this work, we present a novel concept for myocardial T dispersion quantification. Our measurements are based on an accelerated cardiac T-mapping sequence and a new approach called dispersion reconstruction.

Theory

For T dispersion imaging, one typically needs to acquire a series of T-maps with various SL amplitudes fSL. With the new T dispersion reconstruction approach, a T dispersion image can be reconstructed by only acquiring a single T-map and a series of dispersion-weighted images (different fSL, constant SL time tSL). The concept is illustrated in Fig. 1. First, a reference T-map is acquired by varying tSL (constant amplitude fSL,const) and fitting a mono-exponential function:
$$I_{map}=I\left(t_{SL},f_{SL,const}\right)=I_0\cdot\text{exp}\left[-\frac{t_{SL}}{T_{1\rho}(f_{SL,const})}\right]\;\;\;\;\;\text{Eq. 1}$$
Subsequently, several dispersion-weighted images are generated by the variation of fSL at a constant SL time tSL,const.
$$I_{disp}=I\left(t_{SL,const},f_{SL}\right)=I_0\cdot\text{exp}\left[-\frac{t_{SL,const}}{T_{1\rho}(f_{SL})}\right]$$
It is crucial that one of these images is acquired using the amplitude of the reference T-map.
$$I_{ref}=I\left(t_{ref},f_{ref}\right)=I_0\cdot\text{exp}\left[-\frac{t_{ref}}{T_{1\rho}(f_{ref})}\right],\,\,\,t_{ref}=t_{SL,const},\,\,\,f_{ref}=f_{SL,const}$$
Under this condition, a new synthetic T-map can be reconstructed from each dispersion-weighted image using the following equation:
$$T_{1\rho}(f_{disp})=\left(\frac{1}{T_{1\rho}(f_{ref})}-\frac{1}{t_{ref}}\cdot\text{log}\left[\frac{I_{disp}}{I_{ref}}\right]\right)^{-1}\;\;\;\;\;\text{Eq. 2}$$
The information thus obtained can then be used to calculate a dispersion map by pixel-wise fitting a suitable dispersion model (e.g. linear function).

Methods

All measurements were performed on a 7.0T small animal imaging system (Bruker BioSpec 70/30, BioSpin MRI GmbH, Ettlingen, Germany). Our novel dispersion reconstruction approach was validated in phantom as well as in in vivo experiments. T preparation was performed by totally-balanced-spin-locking (TB-SL) [4]. Acquisition of k-space was optimized for very fast myocardial imaging using a radial spoiled gradient echo readout (TE=1.9ms, TR=4.7ms). For every T-prepared image, data acquisition was segmented into 13 preparation experiments (identical tSL and fSL), each separated by a waiting time trec for magnetization recovery. Depending on the imaging mode, 8 different SL-prepared images with different tSL (T-mapping) respectively fSL (dispersion-weighting) were acquired.
In the phantom measurements, 4 cylindrical sample tubes with various BSA concentrations were investigated and the dispersion reconstruction approach was compared with two common full-mapping techniques, once using the accelerated radial readout and once a conventional cartesian readout (in which 8 complete T-maps needs to be acquired). The waiting time trec was adjusted to 1000ms resulting in total measurement times of 3.5min (novel approach), 13.9min (full-mapping radial method) and 25.6min (full-mapping cartesian method). For the in vivo measurements, one SL preparation was performed in one heart beat (during end diastole) per respiratory cycle. Here, trec resulted in approximately 1100ms. A T-map and a synthetic T dispersion image was generated for testing the practicability and for investigation of the achievable image quality in vivo.

Results

Fig. 2 shows the results of the phantom measurements. It is apparent that the common full-mapping cartesian and radial T dispersion methods as well as the novel dispersion reconstruction approach produce identical results with high reproducibility and accuracy. Mean deviation was 0.82%, 0.91%, 1.23% and 1.63% for the four different BSA phantoms. In Fig. 3 and 4 the results of the in vivo measurements are shown. In Fig. 3 the dispersion-weighted images and the reconstructed synthetic T-maps are depicted. Fig. 4 shows a synthetic T dispersion image and the determined dispersion behaviour of blood and myocardium. In myocardial tissue, T values increased from a minimum of 29.5ms to a maximum of 44.3ms with a mean slope of 5.66±0.60ms/kHz. The mean slope in the blood pool was determined to be 34.2±3.1ms/kHz.

Discussion

In this work, a new synthetic dispersion imaging method is presented that is multiple times faster than common full-mapping techniques by comparable accuracy and usability. This acceleration is based on two steps. First, an acceleration of data sampling using optimized radial imaging, and second the implementation of the dispersion reconstruction procedure. Compared to the common cartesian full-mapping method, an acceleration by a factor of 7.4 was achieved. Furthermore, the experiments confirm the applicability of our new method in vivo, which enables the inclusion of T dispersion imaging in a conventional MRI study protocol. Due to the promising results using our new method, further studies on mice are in preparation for investigating the dispersion character of healthy and various diseased tissues. Moreover, our novel method is also feasible in clinical human MRI examinations, which are also planned in the near future.

Acknowledgements

This work was supported by the Federal Ministry for Education and Research of the Federal Republic of Germany (BMBF 01EO1504, MO6)

References

[1] Witschey et al. Magn Reson Med. 2010 Nov;64(5):1453-60

[2] van Oorschot et al. J Magn Reson Imaging. 2015 May;41(5):1181-9

[3] Kamesh Iyer et al. J Cardiovasc Magn Reson. 2019 Jan 10;21(1):5

[4] Gram et al. ISMRM Annual Meeting 2019. Montreal. #1215

Figures

Fig. 1) Concept of the dispersion reconstruction approach. By variation of the SL time (tSL=4…128ms, 8 steps), a T-map is generated at a constant SL amplitude fref=1500Hz. The experiment was repeated using variable SL amplitudes (fSL=750…2500Hz, 8 steps) at a constant SL time tref=72ms. Using the reference map and the dispersion-weighted images, synthetic T-maps for different SL amplitudes were reconstructed. FOV=40x40mm, matrix=96x96

Fig. 2) Comparison of T quantification of different methods in BSA (concentration 10%, 15%, 20% and 25%). All determined T values using the cartesian, the accelerated radial and the even faster dispersion reconstruction method are almost identical.

Fig. 3) Application of dispersion reconstruction in in vivo experiments on mice (short axis view, isotropic resolution 260μm). The reference map was measured with fref=1500Hz (tSL=4…64ms, 8 steps). The dispersion-weighted images were generated at tref=28.9ms (fSL=500…2250Hz, 8 steps). The figure shows the reconstructed synthetic maps using Eq. 2 for different SL amplitudes.

Fig. 4) Analysis of T dispersion in the myocardium based on the synthetic T-maps (cf. Fig. 3) and a linear dispersion model. The figure shows the calculated dispersion map (slope of T values [ms/kHz]) and the behaviour of the dispersion in the myocardial tissue compared to the blood pool. The required data could be acquired in less than 5 minutes.

Proc. Intl. Soc. Mag. Reson. Med. 28 (2020)
0530