Dominique Franson1, Yuchi Liu1,2, Rajiv Ramasawmy3, Adrienne Campbell-Washburn3, and Nicole Seiberlich1,2
1Case Western Reserve University, Cleveland, OH, United States, 2University of Michigan, Ann Arbor, MI, United States, 3National Heart, Lung, and Blood Institute, National Institutes of Health, Bethesda, MD, United States
Synopsis
Fat/water separation at low field strengths can be difficult due to the
small difference between resonance frequencies.
Rosette trajectories have previously been shown to be effective for
spectral separation and fat suppression, and the approach is not dependent on a
large frequency difference. Here, a rosette trajectory is used to significantly
suppress fat signal in water images, and to produce separate fat images at
0.55T. B0 maps are calculated from two
of the rosette echoes, and are used to improve the fat/water separation.
Initial examples are shown in an oil/water phantom, and in the heart and
abdomen.
Introduction
Fat suppression at low field strengths can be challenging due to the
small separation in resonance frequencies between fat and water protons. At 0.55T, the fat proton resonance frequency
is downshifted by approximately 80Hz. This
small separation requires precise, narrow-band frequency selection for
species-selective excitation techniques such as fat saturation pulses, and can
be unreliable.
Rosette trajectories have previously been shown to be effective for spectral
separation and fat suppression1,2.
During the acquisition, the phase of the off-resonance (fat) species evolves
relative to the on-resonance species (water).
By carefully timing the multiple echoes in each readout, the center of
k-space can be collected such that the fat signal destructively interferes with
itself, significantly reducing the signal from fat in the resulting
on-resonance images. Post-acquisition
demodulation of the data to the fat frequency makes it possible to produce a
fat image as well as a water image. B0 maps
can be calculated by pairing two of the acquisition echoes and further used to
correct the fat/water images.
A rosette-based fat suppression approach is attractive for low-field imaging
because it does not require a large frequency difference between water and fat species. It also takes advantage of the ability to use
long readout times, benefiting from the long T2* and the reduced field inhomogeneity
at lower field3. In this work, a rosette trajectory is used
for fat/water separation at 0.55T in an oil/water phantom and for in vivo abdominal
and cardiac imaging.
Methods
A
five-lobe rosette trajectory was designed to suppress signals at an
off-resonance frequency of 80Hz. The
trajectory is comprised of 179 interleaves with a linear rotation angle
ordering, assuming the following parameters: maximum gradient strength = 10mT/m, maximum slew rate = 45 T/m/s, dwell time = 2.5μs, FOV = 300x300 mm2, in-plane
resolution = 1.56x1.56 mm2. The trajectory was used in a balanced steady-state free precession (bSSFP) sequence with the following scan parameters:
flip angle = 70°,
TR/TE = 12.26/1.04 ms, slice thickness = 8 mm. All scanning was performed on an MRI system
modified to operate at 0.55T (prototype MAGNETOM Aera, Siemens, Erlangen,
Germany).
Data were acquired in an oil/water phantom, and in the heart (short
axis view) and abdomen (one axial slice) in an IRB-approved study. Images were collected during free-breathing
and without ECG-gating. Data were
gridded4 to form the water image, and demodulated to -80Hz and gridded to form the
fat image. A B0 correction was performed
by calculating the phase difference between two of the rosette lobe echoes, and
was further used to define the appropriate demodulation frequency for improved
separation of the two species.
A Cartesian gradient echo Dixon-type acquisition was also acquired for
comparison, with the following parameters: FOV = 360x270 mm2,
matrix size = 256x144 (interpolated to 1.41x1.41 mm2 in-plane
resolution), bandwidth = 592 Hz/pixel, flip angle = 20°, 15 segments, 11 heartbeats
while breath-holding.Results
The optimized rosette trajectory for fat/water
separation at 0.55T resulted in a simulated suppression of signal at 80Hz to
6.2% of its original intensity (Figure 1).
Imaging in an oil/water phantom showed significant reduction in the fat
signal using the proposed trajectory (Figure 2) and good separation of the two
species. Imaging in the abdomen (Figure
3) showed good fat/water separation in the central portion of the abdomen,
although there was some residual fat signal around the periphery. Similarly, the rosette performed well in
suppressing fat signal immediately around the heart (Figure 4). B0 correction using the field map from the
rosette trajectory further improved fat suppression in the water images in both
the abdomen and the heart (Figures 3 and 4, see a,b vs. c,d). However, there
was still some residual signal in the breast, along the back, and along the
chest wall, as can be clearly seen by comparing the water and fat images from
the rosette trajectory to those collected using a Dixon-type acquisition
(Figures 3 and 4, c,d vs. e,f). Discussion
A rosette trajectory was used to effectively acquire fat-suppressed
images at 0.55T in an oil/water phantom and in vivo. By demodulating the data
by the fat off-resonance frequency, separate fat images could also be produced
without acquiring additional data. A B0
correction can be performed to help compensate for static field inhomogeneities.
While this initial rosette trajectory did reduce fat signal significantly, residual peripheral fat signal remained in the water
images. This may be due in part to imperfect or local shimming, and
phase wrapping in the calculated B0 maps. Note that while the trajectory was
calculated for fat suppression at 0.55T, it is not yet optimized for B0 mapping.
Additionally, one assumption of the model used here is that each pixel contains
only fat or water, resulting in masking-like edges in the corrected images. Improved B0 mapping and a model that allows
for mixed voxels may provide improved corrections. Conclusion
A rosette trajectory can be used for effective fat suppression for imaging
at 0.55T. Separate water and fat images
can be produced without acquiring additional data. Initial images are shown in an oil/water
phantom, and in vivo for abdominal and cardiac imaging. Acknowledgements
We would like to acknowledge the assistance of Siemens Healthcare in
the modification of the MRI system for operation at 0.55T under an existing
cooperative research agreement (CRADA) between NHBLI and Siemens Healthcare.References
1. Noll DC. Multishot
rosette trajectories for spectrally selective mr imaging. IEEE Trans. Med.
Imaging 1997;16:372–377 doi: 10.1109/42.611345.
2. Liu Y, Hamilton J,
Griswold M, Seiberlich N. Fat/Water Separation and T1 and T2 Quantification
Using MRF with a Rosette Trajectory in the Heart and Liver. In: Proc. Intl.
Soc. Mag. Reson. Med. 27. Montreal; 2019. p. 4376.
3. Campbell-Washburn AE,
Ramasawmy R, Restivo MC, et al. Opportunities in Interventional and Diagnostic
Imaging by Using High-Performance Low-Field-Strength MRI. Radiology
2019;293:384–393.
4. Fessler JA. Michigan
image reconstruction toolbox (MIRT). https://web.eecs.umich.edu/~fessler/code/.