Huiwen Luo1, Wissam AlGhuraibawi2, Kevin Godines2, Daniel Gochberg3, Moriel Vandsburger2, and William A Grissom1
1Biomedical Engineering, Vanderbilt University, Nashville, TN, United States, 2Department of Bioengineering, University of California Berkeley, Berkeley, CA, United States, 3Radiology and Radiological Sciences, Vanderbilt University, Nashville, TN, United States
Synopsis
A tailored spectral-spatial saturation pulse was
developed to produce a flat flip angle profile across the heart and achieve
more uniform CEST saturation despite B1 inhomogeneity at 3 Tesla. The tailored
saturation pulse train was simulated for a two-pool system to evaluate the z-spectrum
at each spatial location in the heart, based on an in vivo 3 Tesla B1 map. Whereas
CEST saturation generated with a conventional Gaussian pulse yielded CEST
contrast of 2.60±1.59% across the ventricle, the tailored pulse produced more uniform
saturation across the heart which resulted in both greater and more uniform CEST
contrast of 4.64±0.34%.
Introduction
Chemical exchange saturation transfer (CEST) imaging [1]
enables in vivo detection of intracellular metabolites. While compensation for
B0 inhomogeneity is routine, CEST contrast is highly sensitive to B1
variations ([2]; see also abstract #5033) for which compensatory approaches are
only recently emerging. Parallel transmit methods have been developed but
require specialized hardware, while post-processing corrections [3,4] require
steady saturation conditions and knowledge of chemical pool parameters, and
their efficacy depends on the specific saturation and imaging parameters. Here
we propose a new approach, which is to use a tailored spectral-spatial pulse to
produce a flat flip angle profile in the presence of B1
inhomogeneity for more uniform saturation in the heart at 3T.Methods
RF Pulse Design:
Figure 1
illustrates the tailored spectral-spatial pulse construction. The pulse
comprises a train of identical spatially-selective subpulses played during
trapezoidal gradients. The subpulse is designed by solving:
$$\hat{\textbf{b}} = argmin_{\textbf{b}}{\|1-\textbf{Ab}\|_2+\lambda\|\textbf{b}\|_2}$$
where $$$\textbf{b}$$$ contains samples of the subpulse and $$$\textbf{A}$$$ is a system matrix of size $$$N_s\times N_t$$$, $$$ a_{ij}= S(\overrightarrow{x_i})e^{i\overrightarrow{x_i}\cdot\overrightarrow{k}(t_j)}$$$, Ns is the number of spatial locations, Nt
is the number of time points, $$$S(\overrightarrow{x_i})$$$ is the measured
B1 amplitude at spatial location $$$\overrightarrow{x_i}$$$, $$$\overrightarrow{k}(t_j)$$$ is the excitation k-space trajectory traced out by
the trapezoid gradient waveform,
and $$$\lambda\|\textbf{b}\|_2$$$is a Tikhonov regularization term. The
gradient direction is also optimized, and the subpulse duration is determined
by spectral sampling requirements. Once the subpulse is designed, it is
duplicated and weighted by a Gaussian or other spectral envelope.
Given a measured 3T cardiac B1 map, an RF subpulse was
computed with the following parameters: maximum duration = 250μs, maximum
magnitude = 4G/cm, maximum slew rate = 14000G/cm/s, dwell time(dt) = 4µs, gx = 0.22G/cm and gy= 0.39G/cm.
The subpulse was duplicated 145 times for a total pulse duration of 36ms. For
comparison, a conventional 36 ms Gaussian pulse was also simulated. The pulses were
scaled to a flip angle of 208°. The B1 map was measured in vivo on a Siemens 3T scanner using
a Turbo FLASH sequence proceeded by a 50° saturation pulse.
Other B1 mapping parameters were field of view (FOV) = 24.3×30cm2, matrix = 156×192, slice thickness = 8mm, TE = 2.12ms, TR = 3.07s.
Numerical
Simulations: Z-spectra were
calculated by solving the Bloch-McConnell equations [5] using the ode45
function in MATLAB, for a two-pool model with a bulk water pool (Pool a) and a myocardial
creatine pool (Pool b). The simulation parameters were: B0 = 3T, T1a
= 2s , T2a =120ms, Δωa=0 ppm, T1b
=1.4s , T2b =50ms , Kba = 200Hz, Δωb=1.8ppm
and pool size ratio (b/a) fb = 0.0018. Both pulses were repeated 25
times for saturation in the simulations, and the interpulse time delay was
21ms.
Results
Figure 2 plots Z-spectra and MTRasym curves for two
voxels at opposite edges of the heart, where Voxel 1 (orange arrow in Fig. 2a) has
lower B1 than prescribed, and Voxel 2 (blue arrow) experiences B1 consistent
with the prescribed flip angle. The effect of B1 inhomogeneity with
conventional Gaussian saturation manifests as reduced creatine CEST contrast
and reduced full width half max of water saturation at Voxel 1 (Fig 2b)
compared to Voxel 2 (Fig 2c). However,
use of a tailored pulse train at Voxel 1 (Fig 2b) restores equal CEST
saturation to that observed at Voxel 2 (Fig 2c), resulting in identical
creatine CEST contrast.
The first row of Figure 3 shows that the tailored
pulse train produces more uniform saturation at 0ppm across the heart: MTR = -0.0015±0.0036 (tailored) versus MTR = -0.0193±0.0247 (Gaussian). The second row shows normalized
flip angle maps that demonstrate substantially enhanced spatial uniformity for
the tailored pulse: the flip angle coefficients of variation were 0.05 (tailored)
and 0.16 (Gaussian). The third row shows CEST contrast maps (%) of cardiac wall
for each pulse. The CEST contrasts in the cardiac wall were 2.60±1.59% and 4.64±0.34% for the Gaussian and tailored pulses,
respectively. Figure 4 shows normalized flip angle and CEST contrast histograms
that illustrate the tighter distributions achieved with tailored saturation. Discussion and Conclusion
In this work, we have designed and demonstrated a
tailored spectral-spatial RF pulse with gradients which achieves a uniform
saturation in the heart despite B1 inhomogeneity, compared with a conventional
Gaussian pulse. In this preliminary study, the tailored and Gaussian pulses
were brought into a two-pool Bloch-McConnell model and the results showed the
advantage of the tailored pulse. Future work will include an in vivo experiment
and further multi-pool simulations fully validate the tailored pulses.
In conclusion, a tailored spectral-spatial RF pulse
solved from a subject-specific B1 map may lead to more uniform
saturation in CEST imaging. Acknowledgements
This work was supported by NIH grants R21 EB 024311,
R01 EB 016695, R01 HL128592, and UH2 EB028908.References
[1] Ward, K. M., A. H. Aletras, and Robert S. Balaban.
"A new class of contrast agents for MRI based on proton chemical exchange
dependent saturation transfer (CEST)." Journal of magnetic resonance 143.1
(2000): 79-87.
[2] Windschuh, Johannes, et al. "Correction of
B1‐inhomogeneities for relaxation‐compensated CEST imaging at 7 T." NMR in
biomedicine 28.5 (2015): 529-537.
[3] Sun, Phillip Zhe, Christian T. Farrar, and A.
Gregory Sorensen. "Correction for artifacts induced by B0 and B1 field
inhomogeneities in pH‐sensitive
chemical exchange saturation transfer (CEST) imaging." Magnetic Resonance
in Medicine: An Official Journal of the International Society for Magnetic
Resonance in Medicine 58.6 (2007): 1207-1215.
[4] Singh, Anup, et al. "On B1 inhomogeneity
correction of in vivo human brain glutamate chemical exchange saturation
transfer contrast at 7T." Magnetic resonance in medicine 69.3 (2013):
818-824.
[5] Roeloffs, Volkert, et al. "Towards
quantification of pulsed spinlock and CEST at clinical MR scanners: an
analytical interleaved saturation–relaxation (ISAR) approach." NMR in
Biomedicine 28.1 (2015): 40-53.