Johanna Lott1,2, Armin M. Nagel1,3,4, Sebastian C. Niesporek1, Thoralf Niendorf5,6, Peter Bachert1,2, Mark E. Ladd1,2,7, and Tanja Platt1
1Medical Physics in Radiology, German Cancer Research Center (DKFZ), Heidelberg, Germany, 2Faculty of Physics and Astronomy, University of Heidelberg, Heidelberg, Germany, 3Institute of Radiology, University Hospital Erlangen, Erlangen, Germany, 4Institute of Medical Physics, Friedrich-Alexander-Universität Erlangen-Nürnberg (FAU), Erlangen, Germany, 5Berlin Ultrahigh Field Facility (B.U.F.F.), Max Delbrueck Center for Molecular Medicine in the Helmholtz Association, Berlin, Germany, 6MRI. TOOLS GmbH, Berlin, Germany, 7Faculty of Medicine, University of Heidelberg, Heidelberg, Germany
Synopsis
Sodium (23Na) ion
distribution plays a fundamental role in biological processes, in particular in
myocardial function. 23Na MRI provides noninvasive information about
the total tissue sodium concentration. However, short relaxation times, low
signal-to-noise ratio, breathing and heart motion render quantitative cardiac 23Na
MRI challenging and result in long acquisition times. We present a method to
compensate for respiratory motion in 23Na MRI by adding a linear
phase in k-space with the goal to determine myocardial tissue sodium
concentration. This enables a reduced measurement time for quantitative
cardiac 23Na MRI compared to retrospective sorting into one
respiratory state.
Introduction
The tissue sodium (23Na)
concentration (TSC) is associated with the viability of cells1, in
particular in myocardial function2. In acute and chronic heart
disease, alterations in myocardial sodium concentration have been observed with
23Na MRI3-5. In
Conn’s Syndrome6 or after myocardial infarction7 the TSC
is increased8 that indicates pathological changes. Thus, 23Na
MRI may allow insights into the underlying mechanisms of sodium overload9.
However, low spatial resolution, short relaxation times, and partial volume effects
(PVE) render quantitative cardiac 23Na MRI challenging10.
The previously applied retrospective sorting methods for cardiac11
and respiratory12 motion in 23Na cardiac MRI require long
acquisition times and discard approximately 70% of the acquired projections10.
Here, we present a method that shifts projections in k-space to compensate for
respiratory motion aiming for clinically feasible acquisition times for
myocardial sodium concentration assessment. Methods
In vivo data of three healthy
volunteers were acquired on a 7T MR system (Siemens, Erlangen, Germany) with an
oval-shaped birdcage coil12 using a density-adapted 3D radial
sampling scheme13 and golden angle projection distribution14.
The following parameters were applied:TR=21ms, TE=0.97ms, α=61°, (∆x)3=(5mm)3,
tAcquisition(ACQ)=35min. To determine binary masks, required for the
correction of PVE, 1H measurements were conducted at 3T
(navigator-gated (exhaled), ECG-triggered (diastole) Flash sequence15
TR=3ms, TE=1.3ms, α=20°, (∆x)3=(2mm)3 interpolated to (∆x)3=(1mm)3 , tACQ=6-10min).
To increase the number of projections
used, all respiratory states within one cardiac phase (diastole) were considered:
All projections are shifted to a reference state (end-exhaled) by adding a
linear phase in k-space. The respiratory signal was normalized16 and
equally divided into five states. The displacement Z in z-direction within a
heart mask was determined using a linear fit of ∆Φ (=phase difference between the
shifted and the reference state) in k-space center:
$$Z=\frac{\Delta \Phi }{2\cdot\Pi \cdot \frac{1}{2\cdot \Delta x}}$$
The data were processed in the
following order and reconstructed using a NUFFT17 with Hanning
filter: 1) correction for B1+ before any processing (Double-Angle-Method18,
α=45°/90°, TR=150ms, TE=1.55ms, tACQ=12min35s, (∆x)3=(10mm)3);
2) sorting into cardiac state; 3) respiratory motion compensation; 4) B0
and 5) partial volume correction (PVC)10,19; 6) determination of myocardial
TSC with binary masks and blood as reference. This evaluation was performed for
datasets with a decreasing number of projections (full, half, quarter of tACQ).
SNR was determined by NEMA Standards20.
Results
The respiratory signal was
normalized and divided into five respiratory states, independent of the number
of projections in each state (Fig.1).
Without motion compensation, the
heart moves approximately (0.40±0.09)cm in z-direction (Fig.2A,3A) between the
end-exhaled and end-inhaled state. The amplitude of this motion was ascertained
for each respiratory state using a linear fit of the phase difference between each
state and the reference state (Fig.2B). With motion compensation, respiratory
motion is markedly reduced (Fig.2C,3B) and comparable to data with retrospectively
sorted into the reference state (Fig.2C).
The number of projections used, SNR
and myocardial TSC for reconstructions with and without motion compensation of
the full, half and quarter dataset are shown in Table1. Applying retrospective
gating uses about (49±0.29)% less projections for the reconstruction compared
to the motion compensation approach. Hence, the SNR is reduced by a factor of (1.46±0.07)
(approximately
$$$\sqrt{\frac{100\%}{49\%}}\sim\sqrt{2}$$$, Fig.4).
Despite decreased SNR and number of projections used, the determined myocardial
TSC of (32±3)mM remains similar.Discussion and Conclusion
With the presented motion
compensation method, the respiratory movement of the heart in z-direction was
detected and remarkably compensated. All projections obtained for the different
respiratory states in one cardiac phase can be used for reconstruction and
acquisition time can be reduced.
Due to the determination of the
displacement Z by applying a linear fit in k-space, no reconstruction of subsets
for an image registration in spatial domain is necessary21.
Consequently, no minimal number of projections for subsets is required. The respiratory signal was equally divided
into five states independent of the number of projections within one state.
At decreased SNR the determined
myocardial TSC remains constant. A further reduction of the number of
projections and thus of the measurement time could be feasible while still reaching
similar values for myocardial TSC. Hence, a detailed analysis of the determined
TSC and the arising variation due to an increasing noise level are
required.
However, the approach has some
limitations. To divide the respiratory signal into more than two states10,
the respiratory signal was normalized. This assumes an equable respiration over
the whole measurement time and might cause an overestimation or underestimation at some
time points. In addition, the movement in x-/y-direction was not considered.
In conclusion, with the presented
motion compensation, all projections of the different respiratory states that
belong to the desired cardiac phase can be used for reconstruction.
Consequently, measurement time can be reduced by a factor of two compared to retrospective
gating into two respiratory states. The presented influence of noise using even
smaller datasets demonstrates a consistently determined myocardial TSC. With
the analyzed size of the dataset, measurement time could be reduced to
<10min, which is suitable for a clinical assessment of myocardial sodium
concentration. Our approach is not limited to the heart but can be also applied
to MRI of e.g. the liver or abdomen.Acknowledgements
This work was supported by iMed–the
Helmholtz Initiative on Personalized Medicine.References
1.
Madelin G et al. Biomedical Applications of Sodium MRI in Vivo. J Magn
Reson Imaging. 2013; 38(3): 511-529.
2.
Sandstede JJ et al. Time Course of 23Na Signal Intensity
after Myocardial Infarction in
Humans. Magn Reson Med 2004;
52: 545-51.
3.
Barclay
JA et al. Electrolyte content of rat heart atria and ventricles. Circulation research 1960; 8: 1264-1267.
4.
Constantinides
CD et al. Noninvasive quantification of total sodium concentrations in acute
reperfused myocardial infarction using 23Na MRI. Magn Reson Med 2001; 46 (6): 1144-1151.
5.
Rochitte
CE et al. Microvascular integrity and the time course of myocardial sodium
accumulation after acute infarction. Circulation
research 2000; 87(8): 648-655.
6.
Christa
M et al. Increased myocardial sodium signal intensity in Conn's syndrome
detected by 23Na magnetic resonance imaging. Eur Heart J Cardiovasc Imaging. 2018; 30.3: 263-270.
7.
Ouwerkerk
R et al. Tissue sodium concentration in myocardial infarction in humans: a
quantitative 23Na MR imaging study. Radiology. 2008;
248: 88–96.
8.
Kim RJ
et al. Relationship of elevated 23Na magnetic resonance image intensity to
infarct size after acute reperfused myocardial infarction. Circulation. 1999; 100: 185-192
9.
Aksentijevic
D et al. Is there a causal link between intracellular Na elevation and
metabolic remodelling in cardiac hypertrophy? Biochem Soc Trans. 2018; 46(4): 817-827.
10.
Lott J et al. Corrections of myocardial tissue
sodium concentration measurements in human cardiac 23Na MRI at 7 Tesla.
Magn Reson Med. 2019; 82.1: 159-173.
11.
Resetar et al. Retrospectively-gated CINE 23Na imaging of the heart
at 7.0 Tesla using densityadapted 3D projection reconstruction. Magn Reson Imaging. 2015; 33(9):
1091-1097.
12.
Platt T
et al. In vivo self‐gated 23Na MRI at 7 T using an oval‐shaped body resonator. Magn Reson Med. 2018; 80: 1005–1019.
13.
Nagel
AM et al. Sodium MRI using a density‐adapted 3D radial acquisition technique. Magn Reson Med. 2009; 62: 1565–1573.
14.
Chan RW
et al. Temporal stability of adaptive 3D radial MRI using multidimensional
golden means. Magn Reson Med. 2009; 61: 354–363.
15.
Bi X et
al. Whole‐heart coronary magnetic resonance angiography at 3 Tesla in 5 minutes
with slow infusion of Gd‐BOPTA, a high‐relaxivity clinical contrast agent. Magn Reson Med. 2007; 58: 1–7.
16.
Behl NG et al.
Dynamic 23Na-Imaging of the human Lung with Fully Flexible Intrinsic
Respiratory Gating. Proc Intl Soc Mag
Reson Med. 2018; 26:0627.
17.
Fessler JA et al. Nonuniform fast Fourier
transforms using min-max interpolation. IEEE
Trans Signal Process. 2003; 51(2): 560-574.
18.
Insko EK et al.
Mapping of the Radiofrequency Field. J
Magn Reson Ser A. 1993; 103(1): 82-85.
19.
Niesporek SC et al. Partial volume correction for in vivo 23Na-MRI
data of the human brain. NeuroImage.
2015; 112: 353-363.
20.
Association NEM. NEMA Standards Publication MS
1-2001. 2001.
21.
Weller
DS et al., Motion-compensated reconstruction of magnetic resonance images from
undersampled data. Magn Reson Med. 2019, 55: 36-45