Jakob Slipsager1,2,3,4, Stefan Glimberg4, Liselotte Højgaard2, Rasmus Paulsen1, Andre van der Kouwe3,5, Oline Olesen1,4, and Robert Frost3,5
1DTU Compute, Technical University of Denmark, Kgs. Lyngby, Denmark, 2Department of Clinical Physiology, Nuclear Medicine & PET, Rigshospitalet, University of Copenhagen, Copenhagen, Denmark, 3Athinoula A. Martinos Center for Biomedical Imaging, Massachusetts General Hospital, Charlestown, MA, United States, 4TracInnovations, Ballerup, Denmark, 5Department of Radiology, Harvard Medical School, Boston, MA, United States
Synopsis
This work compares prospective and retrospective motion correction based
on their capabilities to remove motion artifacts from 3D-encoded MPRAGE scans. Motion
artifacts in clinical and research brain MRI are a major concern and the outcome
of this problem includes repeated scans and the need for patient sedation or
anesthesia, causing increased study time and cost. The prospective and retrospective
correction approaches substantially improve the image quality of in-vivo scans for
similar motion patterns. Prospective motion correction resulted in higher image
quality than retrospective correction for larger discrete movements, and for
periodic motion.
Introduction
Head motion is an ongoing problem in MRI brain imaging, causing reduced diagnostic
image quality or biased research results1,2. Prospective Motion Correction (MoCo)3-5 and retrospective MoCo of k-space6 have shown promising results for improving image quality. Due
to the different correction strategies of these two methods, there are both
technical and workflow-related advantages and disadvantages7. Prospective MoCo in principle updates the image encoding in real-time
to resolve k-space under-sampling that Retrospective MoCo suffers from, however,
prospective MoCo requires pulse sequence modification and the uncorrected
images are not available for comparison. Retrospective MoCo preserves the
original images and if motion is measured with an external tracking system, it
does not require customized sequences. In this work, we compared prospective and retrospective MoCo with
matched-motion experiments. In addition, we also used simulations to assess
retrospective correction.Methods
Rigid head motion was estimated using the Tracoline TCL3.01 markerless
tracking system (TracInnovations, Ballerup, Denmark)8,9. Motion tracking and prospective correction were implemented on a 3T Prisma
scanner (Siemens Healthcare, Erlangen, Germany) with a 64-channel head coil.
During the scan, 30 surface scans of the subject’s face were recorded per
second. The tracking system estimates motion by finding the rigid
transformation that aligns the current surface back to an initial
reference surface. The motion estimates were geometrically aligned to the
scanner by matching a reference surface to the surface of an MRI
calibration scan.
The k-space and motion data were saved from two scan sessions of five
MPRAGE scans of healthy volunteers, who were scanned in accordance with
Institutional Review Board guidelines. In both scan sessions, the two
volunteers were instructed to move in a repeatable pattern. In session one, the
pattern was discrete motion (Fig 2A) and in session two, it was periodic motion
(Fig 3A). Within a session, matched comparisons of medium and larger amplitude
movements were performed. In the last scan, the volunteers were instructed to
remain motionless. The MPRAGE scans have the following protocol FOV=256x256mm2,
matrix=256x256, 176x1mm sagittal slices, in-plane GRAPPA R=2, TR=2500ms,
TE=3.3ms, TI = 1070ms, bandwidth=240Hz/px, echo spacing=8ms, and turbo
factor=176.
Prospective MoCo was applied by modifying an MPRAGE sequence to adjust
the imaging Field of View (FoV) according to motion estimation received from the
tracker. The MoCo sequence adjusts its FoV before the start of each echo-train,
and every six readouts thereafter9. The prospective MoCo sequence was reconstructed with the same
reconstruction algorithm as used for the retrospective MoCo, to avoid dissimilarities
due to scanner versus offline reconstruction methods.
Retrospective MoCo was performed using the freely available retroMoCoBox
software package10. Each k-space line was aligned in time to the nearest available motion
estimate before translations were corrected by adding additional phase ramps to
each line. The non-uniform fast Fourier transformation11 was used to correct for rotations12.
Synthetic motion corrupted k-space data were simulated to further
investigate the performance of the retrospective MoCo when the exact motion is
known. Using code adapted from10, multi-channel k-space data from the motionless MPRAGE scan were used
to generate k-space data during the presence of similar motion as that performed
by the volunteers in the two sessions. The amplitude of the motion was varied. The
generated k-space data were then reconstructed with and without retrospective
MoCo. This simulation is sensitive to k-space under-sampling,
but receiver coil sensitivity effects were not simulated.
Image quality was quantified by computing the Root Mean Square Error (RMSE)
using the motionless MPRAGE scans as ground truth. A rigid registration (to
motionless MPRAGE) and image normalization (zero mean and standard deviation
one) were performed on every image prior to computing RMSE.Results
The motion simulations in Fig 1 show that retrospective MoCo removes the
artifacts in the discrete motion case, however, in the case of periodic motion,
retrospective MoCo was not able to remove all artifacts. The motion estimates
in Fig 2 and 3 show that the motion was similar for each in vivo comparison, although
there was a difference in the large discrete motion comparison. Fig 2B and 3B show that both prospective and retrospective MoCo provide improved image
quality. However, in the case of periodic and large discrete motion, the retrospectively
corrected images have some residual motion artifacts. Fig 4 shows that
prospective and Retrospective MoCo reduced RMSE for both motion patterns when
the backgrounds of the images were removed. Discussion
Prospective and retrospective MoCo both improved image quality during
the discrete and periodic motions tested here. In these preliminary
comparisons, the prospective MoCo images had less artifacts than the
retrospective MoCo images. The subject was trained to repeat motion as
similarly as possible, but for example, in the large discrete motion
comparison, the motion was larger in the scan used for retrospective
correction. Further, in vivo tests are required to definitively assess the
performance of retrospective and prospective correction. The simulations
suggest that periodic motion is more challenging to correct retrospectively.
This was confirmed in the experiments, where the motion was well matched between
the prospective and retrospective corrections.Conclusion
The tested prospective and retrospective MoCo approaches were able to
reduce motion artifacts of 3D structural MPRAGE images. Prospective MoCo resulted
in better image quality in repeated in vivo motion experiments.Acknowledgements
We are grateful for the following NIH funding:
R01HD093578, R01HD085813, R01HD099846, R42CA183150.References
1. Andre JB, Bresnahan BW, Mossa-Basha M, et al.: Toward
quantifying the prevalence, severity, and cost associated with patient motion
during clinical MR examinations. J Am Coll Radiol 2015; 12:689–695.
2. Tisdall MD,
Reuter M, Qureshi A, Buckner RL, Fischl B, van der Kouwe AJW: Prospective
motion correction with volumetric navigators (vNavs) reduces the bias and
variance in brain morphometry induced by subject motion. Neuroimage
2016; 127:11–22.
3. Thesen S, Heid
O, Mueller E, Schad LR: Prospective acquisition correction for head motion with
image-based tracking for real-time fMRI. Magn Reson Med 2000; 44:457–465.
4. Zaitsev M, Dold
C, Sakas G, Hennig J, Speck O: Magnetic resonance imaging of freely moving
objects: prospective real-time motion correction using an external optical
motion tracking system. Neuroimage 2006; 31:1038–1050.
5. van der Kouwe
AJW, Benner T, Dale AM: Real-time rigid body motion correction and shimming
using cloverleaf navigators. Magn Reson Med 2006; 56:1019–1032.
6. Gallichan D,
Marques JP, Gruetter R: Retrospective correction of involuntary microscopic
head movement using highly accelerated fat image navigators (3D FatNavs) at 7T.
Magn Reson Med 2016; 75:1030–1039.
7. Maclaren J,
Herbst M, Speck O, Zaitsev M: Prospective Motion Correction in Brain Imaging: A
Review. Magn Reson Med 2013; 69:621–636.
8. Slipsager JM,
Ellegaard AH, Glimberg SL, et al.: Markerless motion tracking and correction
for PET, MRI, and simultaneous PET/MRI. PLoS One 2019; 14:1–17.
9. Frost R, Wighton
P, Karahanoğlu FI, et al.: Markerless high-frequency prospective motion
correction for neuroanatomical MRI. Magn Reson Med 2019; 82:126–144.
10. RetroMoCoBox
Toolbox [https://github.com/dgallichan/retroMoCoBox]
11. Implementation
of Non Uniform Fast Fourier Transformation
[https://github.com/andyschwarzl/gpuNUFFT]
12. Slipsager JM,
Glimberg SL, Gallichan D, Højgaard L, Paulsen RR, Olesen OV: Clinical
Implementation of Fast Markerless Motion Correction in K-Space of Structural 3D
MR-Images of the Brain. Montreal: ISMRM; 2019.