0451

Bias Field Correction in Hyperpolarized 129Xe Gas Ventilation MRI
Junlan Lu1, Ziyi Wang2, John C. Nouls3, Kush Gulati2, Elianna Bier2, David Mummy3, and Bastiaan Driehuys3
1Medical Physics Graduate Program, Duke University, Durham, NC, United States, 2Biomedical Engineering, Duke University, Durham, NC, United States, 3Radiology, Duke University, Durham, NC, United States

Synopsis

The presence of RF inhomogeneity in hyperpolarized 129Xe ventilation MRI affects quantitative analysis and interpretation. These inhomogeneities are commonly corrected using the N4ITK algorithm, which retrospectively calculates a smoothly varying bias field with non-uniform intensity normalization. However, this algorithm has not been rigorously validated for functional imaging. Here we show, using flip angle maps derived directly from 3D-radial acquisitions of ventilation, that N4ITK may over-correct bias field and remove inherent physiological gradients. We illustrate this comparison by a combination of simulation, phantom, and in vivo ventilation imaging.

Introduction

Hyperpolarized (HP) 129Xe MRI is increasingly used for 3D visualization of regional pulmonary ventilation1. Since these images are often acquired using close-fitting transmit/receive coils, their associated B1 inhomogeneity (“bias field”) leads to regional variations in both flip angles and detection sensitivity. These factors must be corrected to achieve robust and repeatable quantitative image analysis. The current de facto standard for such bias field correction is currently N4ITK2, which retrospectively calculates a smoothly varying bias field through iteratively sharpening the high frequency content of the intensity distribution3. However, this technique has never been validated for lung MRI, where inherent physiological gradients may be mistakenly removed. To address these questions, we compare the bias fields and corrected ventilation images using N4ITK to the recently proposed approach of mapping flip angles directly from 3D radial imaging4. Using a combined approach of simulations, phantom, and in vivo imaging, we demonstrate that N4ITK bias field correction likely needs further attention as we seek to interpret more than defects from 129Xe ventilation MRI.

Methods

Digital Phantom Simulations Simulations of N4ITK versus flip angle mapping were performed using a digital Shepp-Logan Phantom(matrix=256x256), which was Fourier-transformed to generate k-space data and subjected to RF-induced magnetization decay, ignoring T1 effects. K-space was modulated with a simulated, smoothly varying bias field, with flip angles regionally varying between 2 and 7o. The data was radially sampled with 1000 radial spokes (128 points/spoke), using a 2-D golden-angle pattern. Images were reconstructed using a conjugate-gradient nonuniform FFT technique5. These were then corrected using both N4ITK and the flip angle mapping technique of Niedbalski4. N4ITK was applied using default B-spline fitting parameters and a shrink factor of 2 over a mask defining the phantom edge. The flip angle map was generated using the ratio of images reconstructed with the first and second half of k-space. This was converted to a bias field estimate of both transmit and receive sensitivity using $$B1\propto\alpha* \sin(\alpha)*(1-cos^{n_{views}}(\alpha))/(1-\cos(\alpha))$$ and smoothed with an approximating B-spline6 (spline order=3; number of levels=4x4x4; number of control points=4x4x4).

Phantom Imaging This analysis was extended to a system with a known, uniform 129Xe distribution affected only by magnetization decay. Isotopically enriched 129Xe gas was polarized to 25% in volumes of 700 mL and delivered to an evacuated 43-liter acrylic cylinder (h=61cm, D=30cm). The cylinder was wrapped with water tubes to provide coil loading and enable proton localizers. This assembly was wrapped with a quadrature flexible 129Xe transmit-receive chest coil. 129Xe imaging was performed with a 3.0T Siemens Magnetom Trio Scanner (VB19) using a randomized 3D Halton spiral radial sequence (views=1000; samples/view=128; TR/TE=7.5/0.46ms; flip angle=2o; FOV=50cm). 129Xe images were reconstructed using a kernel sharpness7 of 0.15 onto 128x128x128 matrices.  Images were bias field corrected using both methods and constrained with a mask defining the phantom volume. Corrected images were rescaled to the 99th percentile such that intensities ranged from 0-1 for comparison.

In Vivo Imaging Five subjects (3 healthy, 1 young asthmatic, 1 interstitial pneumonia) underwent radial ventilation imaging (views=3600; TR/TE=4.5/0.45ms; flip angle=1.5o; FOV=40cm) in the supine position. One healthy subject was imaged both supine and prone. Images were bias field corrected using both methods as described above, with T1 assumed to be uniform within the lung.

Results

Figure 1 depicts the digital phantom, applied bias field, and uncorrected and corrected images. Note the N4ITK approach misinterprets the bright feature in the phantom as bias field while the flip-angle approach more appropriately corrects for RF inhomogeneity. However, there are residual regions of intensity variations attributable to regions of relatively low SNR in areas of low flip angle.

Figure 2 illustrates imaging, flip angle maps, and the two types of correction for the cylindrical 129Xe spin-density phantom. Flip angle maps can also be extracted in phantoms since T1 is long with respect to imaging time. Standard deviation calculations of spin density inside the masked region demonstrate that both bias field correction methods decrease image heterogeneity compared to the uncorrected value ($$$\sigma_{uncor}=0.32,\sigma_{cor-RF}=0.28,\sigma_{cor-N4ITK}=0.31$$$). However, the two methods generate very different spatial patterns.

Figure 3 shows ventilation imaging, N4ITK-derived bias field, and bias-corrected imaging in a healthy subject scanned both prone and supine. This study illustrates how physiological ventilation gradients in the anterior-posterior (AP) direction may be incorrectly filtered out with N4ITK. Figure 4 shows radially acquired ventilation maps and distribution histograms in 4 subjects that consistently demonstrate that N4ITK applies a stronger intensity correction than is indicated by the RF-depletion approach. This is confirmed by the quantitative metrics in Figure 5.

Discussion

Digital phantom simulations and in-vivo imaging demonstrate that physiological ventilation gradients can be suppressed with standard N4ITK bias field correction as compared to corrections derived from flip angle mapping. These flip angle maps are uniquely derived from radial acquisitions4, but not from the more commonly employed GRE imaging. While these overcorrections are unlikely to materially affect calculation of ventilation defect percentages, they could have a more significant effect on analyzing other aspects of the ventilation distribution such as high and low-intensity voxels. Thus, for imaging that seeks to study and preserve physiological gradients, bias field correction based on radially acquired flip angle maps should be favored.

Acknowledgements

R01HL105643, R01HL12677, NSF GRFP DGE-1644868

References

1. Petousi, N., Talbot, N. P., Pavord, I. & Robbins, P. A. Measuring lung function in airways diseases: Current and emerging techniques. Thorax 797–805 (2019). doi:10.1136/thoraxjnl-2018-212441

2. Song, S., Zheng, Y. & He, Y. A review of Methods for Bias Correction in Medical Images. Biomed. Eng. Rev. 3, 1–10 (2017).

3. Tustison, N. J., Cook, P. A. & Gee, J. C. N4ITK: Improved N3 Bias Correction. 29, 1310–1320 (2011).

4. Niedbalski, P. J. et al. Mapping and correcting hyperpolarized magnetization decay with radial keyhole imaging. Magn. Reson. Med. 82, 367–376 (2019).

5. Lin, J. M. Python non-uniform fast fourier transform (PyNUFFT): An accelerated non-cartesian MRI package on a heterogeneous platform (CPU/GPU). J. Imaging 4, 1–22 (2018).

6. Tustison, N. J. & Gee, J. C. N-D Ck B-Spline Scattered Data Approximation. 1–13 (2006).

7. Robertson, S. H. et al. Optimizing 3D noncartesian gridding reconstruction for hyperpolarized 129 Xe MRI-focus on preclinical applications. Concepts Magn. Reson. Part A 44, 190–202 (2015).

Figures

Simulated radial sampling and bias field correction. The applied flip angle map is smoothly varying from 2 to 7o. The uncorrected image demonstrates RF inhomogeneity, that is also captured in the RF-depletion flip angle map. N4ITK incorrectly identifies a bright spot of RF inhomogeneity. By contrast, the RF-depletion correction method largely improves unwanted inhomogeneity.

a) Cylindrical 43-liter phantom with 1H localizer tubes. b) Phantom assembly fitted with RF flex coil on table (note the phantom extends significantly beyond the coil). c,e,g) show the uncorrected, RF-depletion corrected, and N4ITK-corrected images. d,f,h) show the calculated flip angle, its associated bias field map, and the N4ITK bias field map. Both methods improve the heterogeneity of the uncorrected spin density image, noted by reduced standard deviation compared to uncorrected. However, the bias fields have very different spatial patterns.

Illustration of N4ITK removal of physiological gradients in a healthy subject imaged supine and prone. a) Uncorrected ventilation in the supine position exhibits a prominent gradient in the AP direction. N4ITK bias field captures and removes this gradient from the corrected image. b) Uncorrected ventilation in the prone position shows a gradient in the opposite direction. This is again captured by the N4ITK bias field and removed. The reversal of the gradient between supine and prone positions suggests it is physiological rather than due to RF-inhomogeneity and should be preserved.

Comparison of RF-depletion-corrected versus N4ITK-corrected ventilation images and histograms for 4 subjects. These examples illustrate that the N4ITK algorithm tends to largely remove the AP ventilation gradient, whereas the RF-depletion approach changes the images only modestly. This even more readily perceived in the ventilation histograms which demonstrate that N4ITK skews the average ventilation towards significantly higher values.

Mean extracted flip angles (ignoring T1 effects) derived from subject images were approximately equal to the applied flip angle. Bias fields derived from the RF depletion method had a lower interquartile range than those from N4ITK. Corrected ventilation median from RF depletion method has a more modest increase from uncorrected ventilation median than that of N4ITK. Both indicate this algorithm applies a more aggressive correction than indicated by flip angle mapping.

Proc. Intl. Soc. Mag. Reson. Med. 28 (2020)
0451