Zhitao Li1, John Pauly2, and Shreyas Vasanawala1
1Department of Radiology, Stanford University, Stanford, CA, United States, 2Electrical Engineering, Stanford University, Stanford, CA, United States
Synopsis
A radial dual-echo IR-SPGR
technique combined with a principal component based iterative
reconstruction algorithm are demonstrated for fat and water separated
rapid high-resolution abdominal multi-parameter mapping. This method
can yield high-quality fat-signal-fraction map, B0 field map,
composite T1 map, water component T1 map as well as a R2* map from a
scan as fast as 3seconds/slice. The in-plane resolution of the resulting parameter maps is 1.25mm
and the through-plane resolution is 5.00mm. With a selective inversion
pulse, multiple slices can be acquired within a breath-hold.
Introduction
Quantitative measurements of MR
physical parameters are becoming more desirable in clinical practice.
It has been shown that T1 maps are valuable for various abdominal
diseases such as cirrhosis [1], tumors [2] and kidney dysfunction
[3]. However, accumulation of fat and iron in the abdomen can bias
the measured T1 values. To avoid the bias introduced by fat, it is
desirable, particularly for hepatic indications, to measure the T1
values of only the water component. Further, to account for the bias
introduced by iron, it is also desirable to quantify R2*
simultaneously to estimate iron concentration.
Breath-holding puts a severe
limitation on the spatiotemporal resolution for conventional
parameter mapping techniques [4-7].
Here we present a radial
dual-echo inversion-recovery (IR) Spoiled Gradient-Echo (SPGR)
technique for high-resolution multi-parameter mapping. This method is
demonstrated for breath-hold imaging studies that produced B0,
fat-signal-fraction, composite T1, water T1, and R2* maps. Data can be
acquired around 3 seconds/slice.Technique and methods
In the proposed technique, data
acquisition is initiated by a 180° adiabatic slice-selective
inversion pulse, followed by a train of fixed flip-angle dual-echo
radial views that are acquired in a single-shot. Tiny golden angle
[8] view-ordering was used to avoid eddy current artifacts (Figure
1(a)). For fat-signal-fraction estimation, composite images of the
two echoes were used. For subsequent parameter estimations, data is
partitioned into TI-groups (N=16 views/group, ~4% Nyquist); the TI
for each TI-group is approximated by the average TI of the views in
the group (Figure 1(b)).
This dual-echo radial IR-SPGR
sequence was implemented on GE 3T MR750 scanner using an EPIC wrapper
from Karolinska Foundation [9]. For a breath-hold acquisition, the
following parameters were used: TE1=1.88ms, TE2=3.35ms, TR=5.8ms,
TI=35ms, FA=10°, base resolution=256, FOV=320mm, radial views=512,
bandwidth=976.6Hz/pixel.
A four-step reconstruction
(Figure 1(c)) was used to recover multi-parameter maps. First,
composite echo-images were reconstructed using a naïve non-uniform
FFT [10]. B0 and fat-signal-fraction maps were then estimated from
these two echo-images [11].
Secondly, a principal-component
(PC) based algorithm was used for the reconstruction of TI-images
[12]. Two TI-image series were reconstructed from the dual-echo data,
but only the first echo was used to estimate the composite T1 maps
due to SNR considerations. The algorithm was able to recover 3 PC
images to reconstruct a series of 32 TI-images, which were then
fitted to the T1 model in Eq.(1) [13].
Thirdly, the fat-signal-fraction
map from step-1 and the composite T1 map from step-2 were used as the
initial condition for the non-linear fitting of the fat/water
separated T1 model described in Eq.(2) to recover water T1 map.
$$$T_{1f}^{*}$$$in the equation is treated as known and is estimated using the
composite T1 values with corresponding fat-signal-fraction values >
95%.
Lastly, the R2* map was estimated
using the last 10 dual-echo TI-images in order to avoid the low
signal intensity around the zero-crossing due to the
inversion-recovery. Before fitting for R2*, the T1 recovery effect
between the dual-echo TI-images was corrected using the estimated
composite T1 maps. The corrected data was then fitted to Eq.(3). A
clinically available R2* mapping sequence IDEAL was used as a
reference for R2* accuracy analysis.
A NIST phantom [14] was used to
verify the accuracy of the T1 mapping technique using an IR-Spin-Echo
(IR-SE) method as a reference. With IRB approval, a subject with
hepatic steatosis
was imaged using the proposed technique
with informed consent.
Reconstruction algorithms were
implemented on dual NVIDIA GTX-1080 GPUs using C++ with OpenCL, and
supporting routines in python and MATLAB.Results and discussion
In Table 1 from Figure 4(a), the T1 measured by
IR-SE and the proposed technique on the NIST phantom are shown.
Percentage error along with a Bland-Altman plot is also presented. It can be seen that the proposed
method matches the reference very well.
In Figure 2 the first row shows
the 2 composite echo-images and B0 map, the second row shows the
fat/water separated images and the fat-signal-fraction map. The third
row of the image shows a comparison between the R2* maps from the
IDEAL reference and the proposed technique. Note on the R2* ROI
analysis, the proposed technique produced R2* values matching the
reference, but with much better image quality and resolution.
In Figure 3, representative
slices of the composite T1, water T1 and the R2* maps are shown.
Multi-slice data acquisition took a 15-second breath-hold. The
spatial resolution of the maps is 1.25mm×1.25mm×5.00mm. Note the
sharp details presented in the parameter maps. The high spatial
resolution in the T1 maps, allows the kidney’s cortex and medulla
to be conspicuously delineated. As for the liver and spleen
parenchyma, few artifacts are present and values are generally
uniform.
In Figure 4(b), ROI analysis of the
composite and water T1 maps are shown. With fat signals excluded, the
water T1 maps show higher estimates in the liver and pancreas. Other
organs with little fat, such as spleen, the estimates are unchanged.Conclusion
A fast multi-slice dual-echo
IR-SPGR technique is proposed for high-resolution abdominal
multi-parameter mapping. B0, fat-signal-fraction, composite T1, water
component T1 and R2* maps can be acquired around 3seconds/slice. The proposed
method is efficient in data acquisition and can provide high-quality
estimations of multiple parameters.Acknowledgements
The authors would like to thank NIH R01 EB009690, NIH R01 EB026136, GE HealthcareReferences
[1] Kim KA, Park MS, Kim IS, Kim
IS, et al. Quantitative evaluation of liver cirrhosis using T1
relaxation time with 3 Tesla MRI before and after oxygen inhalation.
JMR. 2012;36:405-410.
[2]
Goldberg MA, Hahn PF, Saini S, Cohen MS, Reimer P, Brady TJ, Mueller
PR. Value of T1 and T2 relaxation times from echoplanar MR imaging in
the characterization of focal hepatic lesions. AJR Am J Roentgenol.
1993;160(5):1011-1017.
[3]
Huang Y, Sadowski EA, Artz NS, et al. Measurement and comparison of
T1 relaxation times in native and transplanted kidney cortex and
medulla. J Magn Reson Imaging. 2011;33:1241-1247.
[4]
Messroghli DR, Radjenovic A, Kozerke S, Higgins DM, Sivananthan MU,
Ridgway JP. Modified Look-Locker inversion recovery (MOLLI) for
high-resolution T1 mapping of the heart. Magn Reson Med.
2004;52:141-146.
[5]
Ma D, Gulani V, Seiberlich N, Liu K, Sunshine JL, Duerk JL, Griswold
MA. Magnetic resonance fingerprinting. Nature. 2013;495:187-192.
[6]
Chen Y, Jiang Y, Pahwa S, Ma D, Lu L, Twieg MD, Wright KL, Seiberlich
N, Griswold MA, Gulani V. MR fingerprinting for rapid quantitative
abdominal imaging. Radiology. 2016;279(1):278-286.
[7]
Wang X, Roeloffs V, Klosowski J, Tan Z, Voit D, Uecker M, Frahm J.
Model-based T1 mapping with sparsity constraints using single-shot
inversion-recovery radial FLASH. Magn Reson Med. 2017;00:00-00.
[8]
Wundrak S, Paul J, Ulrici J, et al. Golden ratio sparse MRI using
tiny golden angles. Magn Reson Med 2016;75:2372–2378.
[9]
http://ksfoundationepic.org/
[10]
Fessler JA, Sutton BP. Nonuniform fast Fourier transforms using
min-max interpolation. IEEE Transactions on Signal Processing.
2003;51(2):560-574.
[11]
Berglund J, Ahlström H, Johansson L, Kullberg J, Two-point dixom
method with flexible echo times, Magn Reson Med. 2011;65:994-1004.
[12]
Li Z, Bilgin A, Johnson K, Galons JP, Vedantham S, Martin DR, Altbach
MI, Rapid high-resolution T1 mapping using a highly accelerated
radial steady-state free-precession technique, J Magn Reson Imaging
2019;49:239-252.
[13]
Barral JK, Gudmundson R, Stikov N, Etezadi-Amoli M, Stoica P,
Nishimura DG. A robust methodology for in vivo T1 mapping. Magn Reson
Med. 2010;64:1057-1067.
[14]
Russek SE, Boss M, Jackson EF, et al. Characterization of NIST/ISMRM
MRI System Phantom. In: Proceedings of the 20th Annual Meeting of
ISMRM, Melbourne Australia, 2012. (abstract 2456).