Jie Chen1,2, Jun Chen1, Jeremiah Heilman1, Kevin Glaser1, Richard Ehman1, and Meng Yin1
1Mayo Clinic, Rochester, MN, United States, 2West China Hospital, Sichuan University, Chengdu, China
Synopsis
Simultaneous stiffness assessment of multiple organs in patients with systemic
diseases may be beneficial. Here we used four passive MRE drivers to assess the
feasibility and repeatability of simultaneous stiffness measurements of the liver,
spleen and kidneys using 60-Hz vibrations. No significant bias was observed in
the stiffness measurements of the liver and kidneys with 1 versus multiple
drivers. Splenic stiffness was lower with four drivers. With four drivers, it is
possible to obtain simultaneous stiffness measurements of the liver, spleen and
kidneys with repeatability coefficients similar to those using a single driver
while also overcoming attenuation in the liver.
Introduction
Magnetic resonance elastography (MRE) is a phase-contrast-based MRI
technique used to visualize shear wave motion in biological tissues to provide quantitative
measurements of the mechanical properties of the tissue in vivo. To date, MRE
has primarily been used clinically as a quantitative tool for diagnosing and
staging liver fibrosis. Other applications in the upper abdomen include the
spleen, pancreas, and kidneys.1 Typically MRE is performed using a
single passive driver secured to the body (the source of the motion introduced
into the body) targeting a single organ.
However, there are emerging interests in the simultaneous stiffness
evaluation of multiple organs to assess systemic changes, such as simultaneous
hepatic and splenic stiffness measurements to evaluate portal pressure and the
hepatic venous pressure gradient (HVPG).2 Using a single driver may not produce the
necessary amount of motion in all of the organs of interest in a single
acquisition, so using multiple drivers simultaneously may be required for these
applications. Another potential benefit of
using multiple drivers is for liver MRE.
Wave attenuation,3,4 especially in patients with soft livers
with little or no fibrosis, can make it difficult to measure the stiffness of
deep liver tissue and tumors far away from the driver. Using multiple drivers around the abdomen may
produce more motion deeper in the liver to improve these measurements. The
purpose of this study was to explore the feasibility of simultaneous stiffness
evaluation of the liver, spleen and kidneys in healthy volunteers using
multiple acoustic drivers, and to compare the stiffness measurement of each organ
using multiple drivers to that from a single driver. Methods
Between July 2019 and October 2019, this IRB-approved study was
performed in 10 volunteers with no known significant liver diseases, as indicated
by their medical records. All volunteers provided written informed consent and
fasted at least four hours before the MRE scan. All examinations were performed
on a 3T scanner (GE, Signa HDxt, HD16.0_V03) at the end of expiration. Four soft,
flexible, acoustic passive drivers which was connected to independent active
driver (Resoundant,
Inc., Rochester, MN) were used and were placed as shown in Figure 1. The drivers
were activated individually and simultaneously in different combinations. A single-shot,
flow-compensated, spin-echo, EPI MRE sequence was used with the following
imaging parameters: 60-Hz vibrations, 32
slices, TR/TE = 1300/46.8 ms, slice thickness = 3.5 mm, ASSET factor = 2, 3 MRE
time points, and 6 motion-encoding directions (+- X, Y, Z). The examination was
repeated three times: twice on the same day, and the third time was conducted on
a different day. Stiffness images (elastograms) were calculated offline using a
customized 3D MRE algorithm. After visual evaluation of sufficient waves in
each organ for driver selection, stiffness measurement was conducted by the
same person for the liver, spleen, left and right kidney. For each driver
combination, the mean tissue stiffness was compared to that from using a single
driver using the Wilcoxon signed-rank test. The same-day, between-days and
overall repeatability of the stiffness measurement for the liver, spleen, left
and right kidney using 1 and multiple drivers were also calculated. A p-value
of 0.05 was considered statistically significant for all analyses. Results
All volunteers (male/female: 7/3; age:
26-43 years) had no significant liver diseases. The liver fat fraction, as
calculated by 2-point Dixon imaging performed in the same exams, ranged from
5.9% to 27.4%. The body mass index (BMI) ranged from 19.81 kg/m2 to
29.98 kg/m2. The mean time interval between the first and second
visit was 5.6 days (1-19 days). Assessed
by visual evaluation, when all four drivers were activated, the amplitude of the
shear waves was higher deeper in the liver than when only a single driver was
used (Figure 2). The mean stiffness of each organ using 1 and multiple drivers is
listed in Table 1. There was no significant difference in liver or kidney stiffness
using 1 or multiple drivers. Compared with a single driver, the splenic
stiffness was significantly lower when drivers 1+3 or all four drivers were
activated together. The overall, same-day and between-days repeatability
coefficient of the stiffness measurements using 1 and multiple drivers are
shown in Table 2.Discussion
As demonstrated in Figure 2, the deeper
portions of soft livers, which have high attenuation, can be better illuminated
with waves from different directions when all four drivers are activated and the
measurements can be reported with higher confidence. This could be beneficial for
characterizing deep focal liver lesions by using multiple-driver MRE to provide
a more uniform wave illumination of the lesion to address the shadowing effect.4
The calculated repeatability coefficient in the liver in this study group
agrees with our previous assessments of 3D MRE. The difference in spleen
stiffness using 1 and multiple drivers was speculated to be caused by the stronger
vibrations when the left posterior driver (driver 3) was activated, which was
placed right against the spleen. In conclusion, using 4 drivers allows
simultaneous stiffness evaluation of the liver, spleen and kidneys with
repeatability coefficients similar to those when using a single driver, and
provides a more complete assessment of the liver as well. Acknowledgements
No acknowledgement found.References
1. Wang,
J., et al., New and Emerging Applications
of Magnetic Resonance Elastography of Other Abdominal Organs. Top Magn
Reson Imaging, 2018. 27(5): p.
335-352.
2. Dyvorne, H.A., et al.,
Simultaneous measurement of hepatic and
splenic stiffness using MR elastography: preliminary experience. Abdom
Imaging, 2015. 40(4): p. 803-9.
3. Zheng, Y., et al., Magnetic resonance elastography with twin
pneumatic drivers for wave compensation. Conf Proc IEEE Eng Med Biol Soc,
2007. 2007: p. 2611-3.
4. Mariappan, Y.K., et
al., Magnetic resonance elastography with
a phased-array acoustic driver system. Magn Reson Med, 2009. 61(3): p. 678-85.