Mehrgan Shahryari1, Helge Herthum1, Gergely bertalan1, Tom Meyer1, Heiko Tzschätzsch1, Carsten Warmuth1, Jürgen Braun2, and Ingolf Sack1
1Department of Radiology, Charité - Universtitätsmedizin Berlin, Berlin, Germany, 2Institute of Medical Informatics, Charité - Universtitätsmedizin Berlin, Berlin, Germany
Synopsis
MR elastography can provide high-resolution stiffness maps of abdominal organs. However, MRE – in particular when applied with multiple drive frequencies – requires measure times which significantly exceed single breath holds. Therefore, reduction strategies for motion artifacts are required including breath-holds, navigators and image registration, which all were consistently applied and analyzed in this in-vivo study. Our results show that displacement of organs is smallest during breath-hold MRE while, remarkably, mean stiffness values are not significantly affected by breathing. Overall image quality is comparable between breath-hold
and free-breathing MRE when the latter is corrected by 2D-image registration
during post processing.
Introduction:
Abdominal magnetic
resonance elastography (MRE) is a reproducible quantitative MRI method that
provides imaging biomarkers reflecting the viscoelastic properties of tissues1. Since the
introduction of MRE in 1995 by Muthupillai
et al., various MRE methods have been developed which differ in image
acquisition sequences, mechanical vibration hardware or image processing
algorithms. Most abdominal MRE examinations are conducted during breath-hold,
whereas other methods, such as tomoelastography allow the patient to breathe
freely2,3. The impact of
breathing motion on abdominal MRE has been addressed by only few studies4,5. Therefore, we aimed
at (I) displacement quantification and (II) stiffness quantification of
abdominal organs acquired by MRE under free-breathing, breath-hold and respiratory
navigated tomoelastography, as well as (III) improvement of image quality by
2D-image registration before tomoelastography processing.Method:
11 healthy male
participants were examined on a 1.5 T MRI Scanner (Magnetom Sonata, Siemens,
Germany). MRE was performed using a single-shot spin-echo EPI sequence as
described in2 using externally
induced mechanical vibrations of 30, 40, 50 and 60 Hz. 9 coronal image slices with
2.7×2.7×5 mm³ voxel size were acquired at eight time steps equally spaced over
a full vibration cycle with three dimensional flow-compensated motion encoding
gradients (MEG). Full MRE examination was conducted in each participant four
times: (1) repeated breath-hold examinations (BH) in expiration, (2)
free-breathing (FB) during the whole scan, (3) respiratory navigator-triggered with
gating method (G) and (4) respiratory navigator-triggered with gating and following
(GF) by slice adjustment of the respective abdominal motion. Navigator MRE was
performed with a pencil beam technique capturing diaphragmatic motion. Total
MRE scan time was approximately 6min for BH, 3.5min for FB, 5min for G and 4.5
min for GF. 2D Displacement of the liver, spleen, pancreas and kidneys were quantified
along the cranio-caudal axis for each of the four experiments by 2D-image
registration, choosing one slice covering the majority of the respective organ.
Using elastix with (i) rigid body transformation
and (ii) advanced Mattes mutual information as a metric, all images are
registered to the first of all n=96 time points yielding the 2D displacement (xn, yn)6,7. Based on the
effective displacement un=√((x ̅-xn )2+(y ̅-yn )2) the standard deviation σu and the range Δu=max(un)-min(un) are calculated. MRE data were
postprocessed using k-MDEV providing
maps (elastograms) of shear wave speed c
in m/s which indicate tissue stiffness8.
Statistical analyses
were performed for comparing Δu and σu, as well as c in elastograms the liver, pancreas,
kidney and spleen acquired by the four MRE experiments (BH, FB, G, and GF).
Additionally, elastograms were generated after motion correction by image
registration.Results:
Organs were
differently affected by breathing motion (Figure 1A) with liver showing highest
displacement amplitudes (Δu=13.75 mm;
σu=3.04 mm) followed by spleen
(Δu=11.15 mm; σu=2.18 mm), pancreas (Δu=10.26 mm; σu=2.25
mm) and kidneys (Δu=7.45 mm; σu=1.71 mm). Kidneys
exhibited significantly lower displacement than liver (P<0.05).
Breath-hold MRE
(Figure 1B) resulted in relatively amotile organs with the following effective displacements:
liver (Δu=4.8 mm; σu=0.77 mm), pancreas (Δu=4.22 mm; σu=0.72 mm), spleen (Δu=3.43 mm; σu=0.53
mm) and kidneys (Δu=2.94 mm; σu=0.5 mm).
Navigator based MRE
allowed for motion compensation yielding good effective displacement reduction in
the liver (GF: Δu=8.43 mm; σu=1.82 mm; G: Δu=13.35 mm; σu=2.57 mm). GF resulted in significant lower effective displacement
than free-breathing MRE. Figure 1C shows effective displacement values of the
different measurement techniques.
Despite large
differences of effective displacement amplitudes among organs and techniques
mean shear wave speed c values were
not significantly different among the different techniques (Figure 2- 3 and
Table 1). In general, visual inspection indicates better detail resolution in
the stiffness maps was better in breath-hold than free-breathing MRE (Figure 2A
and B). Further improvement was observed by rigid body image registration of
free-breathing data (Figure 4). Specifically, in renal tissue and pancreas
sharper organ borders and less blurry images could be achieved by image
registration.Discussion and Conclusion:
Breath-hold MRE
results in smallest displacements of abdominal organs compared to
free-breathing MRE and navigator-triggered gating MRE techniques. Among
abdominal organs, the liver is most affected by breathing while the kidneys
show smallest displacement amplitudes. Other than indicated by blurred MRE
magnitude images in free-breathing examinations, mean stiffness values were not
significantly affected by the motion. 2D rigid body registration of images before
multifrequency inversion leads to an overall improved image quality and sharper
tissue interfaces especially for data acquired under free-breathing.
Among all tested
artifact reduction strategies, free-breathing multifrequency MRE in combination
with retrospective rigid body image registration is recommended for the
following reasons: i) shortest acquisition times, ii) no collaboration of the
patient is needed, and iii) overall stability of mean stiffness values.Acknowledgements
References
1. Muthupillai, R., et al. Magnetic resonance elastography by direct visualization of
propagating acoustic strain waves. Science
269, 1854-1857 (1995).
2. Dittmann, F., et al. Tomoelastography of the
abdomen: Tissue mechanical properties of the liver, spleen, kidney, and
pancreas from single MR elastography scans at different hydration states. Magnetic Resonance in Medicine 78, 976-983 (2017).
3. Shahryari, M., et al. Tomoelastography Distinguishes
Noninvasively between Benign and Malignant Liver Lesions. Cancer Res (2019).
4. Murphy, I.G., et al. Comparison of breath-hold,
respiratory navigated and free-breathing MR elastography of the liver. Magn Reson Imaging 37, 46-50 (2017).
5. Garcia, S.R.M., et al. Tomoelastography of the native
kidney: Regional variation and physiological effects on in vivo renal
stiffness. Magnetic Resonance in Medicine
79, 2126-2134 (2018).
6. Klein, S.,
Staring, M., Murphy, K., Viergever, M.A. & Pluim, J.P. elastix: a toolbox
for intensity-based medical image registration. IEEE Trans Med Imaging 29,
196-205 (2010).
7. Shamonin, D.P., et al. Fast parallel image
registration on CPU and GPU for diagnostic classification of Alzheimer's
disease. Front Neuroinform 7, 50 (2013).
8. Tzschatzsch, H., et al. Tomoelastography by
multifrequency wave number recovery from time-harmonic propagating shear waves.
Medical Image Analysis 30, 1-10 (2016).