Shan Pi1, Jonathan M. Scott2, Yin Li3, Hui Peng3, Huiquan Wen1, Matthew C. Murphy2, Jingbiao Chen1, Meng Yin2, Jun Chen2, Kevin J. Glaser2, Rchard L. Ehman2, and Jin Wang1
1Department of Radiology, the Third Affiliated Hospital, Sun Yat-sen University, Guang Zhou, China, 2Department of Radiology, Mayo Clinic, Rochester, Micronesia, 3Department of Nephrology, the Third Affiliated Hospital, Sun Yat-sen University, Guang Zhou, China
Synopsis
Chronic kidney disease (CKD)
is increasing in incidence and prevalence worldwide and early detection of CKD is
a major challenge. MR
elastography (MRE) is a noninvasive technique capable of quantifying the
mechanical properties of tissue that has shown potential for assessing kidney
diseases. MRE using 60-Hz and 90-Hz vibration frequencies can provide potential
quantitative biomarkers for evaluating kidney function and biopsy score in CKD patients.
Introduction
Chronic kidney disease (CKD)
related to hypoxia and fibrosis is increasing in incidence and prevalence
worldwide. Early detection of CKD is a major challenge and having noninvasive
medical imaging modalities sensitive to changes in renal function and
anatomical fibrosis would benefit the clinical practice. MR elastography (MRE)
has gained attention as a promising method to assess patients with diabetic
nephropathy, kidney transplants, lupus nephritis and renal tumors(1-4), and these studies showed
that the renal shear modulus depends on blood perfusion state and fibrosis of
the kidney. The aim of this study is to assess the diagnostic performance of 3D
multifrequency MRE for the evaluation of CKD.Methods
Following ethics committee approval with a
waived informed consent requirement, 38 patients with CKD underwent renal MRI
and 3D multifrequency MRE exams using a 3.0T MR system (Discovery MR750, GE
Healthcare) before a renal biopsy procedure. The MRE passive pelvic-wall drivers
were developed by Mayo Clinic, and multifrequency coronal MRE acquisitions with
60-Hz and 90-Hz vibration frequencies were performed. 10 patients were excluded
due to the MRE scan being performed after the renal biopsy (n=2) and unsatisfactory
MRE quality (n=8). Thus, 28 patients with CKD were enrolled in this study. Histology
analysis in the 23 of 28 remaining patients was based on a uniform, semiquantitative
approach to assessing change (including glomerulosclerosis, tubular atrophy,
interstitial fibrosis, and arteriosclerosis) followed by biopsy grading between
1 and 4(5). The criteria for scoring and grading the
chronic changes in renal tissue are described in more detail in Sethi et al(5). Renal function was determined using the
estimated glomerular filtration rate (eGFR), which was calculated from blood
creatinine levels using the Chronic Kidney Disease Epidemiology Collaboration equation(5). The total imaging time for MRE was about 64 seconds, performed in
six 11-second breathe-holds and at the end of expiration. The acquisition parameters for MRE were
as follows: single-shot, flow-compensated, spin-echo, echo-planar imaging pulse
sequence; TR/TE = 1300/55.8ms; FOV = 36×36 cm; acquisition matrix = 96×96;
number of excitations = 1; bandwidth = 250 kHz; number of slices = 32; slice
thickness = 3.5 mm with 0.5-mm gap. Stiffness maps
were obtained using a previously described direct inversion algorithm(6, 7) . The success of MRE was defined as the
presence of visually detectable wave propagation in the kidneys. Renal regions of
interest (ROIs) were drawn in the parenchyma of three consecutive slices
centered on the hilum of the kidney. The diagnostic performance for the differentiation
of CKD stage was assessed with the area under the receiver operating
characteristic curve (AUROC). The relationship between the stiffness and kidney
volume, CKD stage and biopsy scores were analyzed by Pearson correlations. Statistical
significance was defined as P<0.05.Results
Baseline variables, including age, sex, eGFR, biopsy score, stiffness and
kidney volume in the 28 CKD patients are shown in Table 1. The 60- and
90-Hz stiffness diagnostic values are shown in Table 2. The stiffness
cutoff value at 60 Hz and 90 Hz for discriminating CKD stage 1 from CKD stages
2-5 was 3.473 kPa and 5.500 kPa with 82.1% and 82.1% accuracy, 86.7% and 80%
sensitivity and 76.9% and 84.6% specificity, giving an AUROC of 0.841 and 0.856
(Figure 2). There was no
significant difference between the diagnostic efficacy at 90 Hz and 60 Hz (P=0.83).
MRE shear stiffness was
positively correlated with eGFR levels (60 Hz, r=0.696, P<0.001; 90 Hz, r=0.644,
P<0.001) (Figure 3), and was negatively correlated
with biopsy score (60 Hz, r =-0.557, P =0.006; 90 Hz, r=-0.569, P=0.005)
(Figure 3). Discussion
In this study, it was
shown that kidney stiffness using
60-Hz and 90-Hz vibration frequencies can be used to evaluate kidney function at
different CKD stages, and the stiffness was significantly correlated with eGFR and
biopsy score in patients with CKD. The stiffness also decreased along with the
progression of CKD. Possible explanations for these observations are that the stiffness
reflects the lower perfusion and fibrosis in the renal tissue, and that the
decreasing renal perfusion factor has more of a contribution to the observed
renal stiffness than renal fibrosis in advanced-stage CKD. Further research is
warranted to investigate the results. Conclusion
The results of our
study showed that 3D MRE is a promising and useful technique for predicting
kidney function and biopsy score. Acknowledgements
No acknowledgement found.References
1. Brown
RS, Sun MRM, Stillman IE, et al. The utility of magnetic resonance imaging for
noninvasive evaluation of diabetic nephropathy. Nephrol Dial Transplant 2019
2.
Marticorena Garcia SR, Fischer T, Durr M, et al. Multifrequency Magnetic
Resonance Elastography for the Assessment of Renal Allograft Function. Invest
Radiol 2016;51(9):591-595
3.
Marticorena Garcia SR, Grossmann M, Bruns A, et al. Tomoelastography Paired
With T2* Magnetic Resonance Imaging Detects Lupus Nephritis With Normal Renal
Function. Invest Radiol 2019;54(2):89-97
4.
Prezzi D, Neji R, Kelly-Morland C, et al. Characterization of Small Renal
Tumors With Magnetic Resonance Elastography: A Feasibility Study. Invest Radiol
2018;53(6):344-351
5. Sethi
S, D'Agati VD, Nast CC, et al. A proposal for standardized grading of chronic
changes in native kidney biopsy specimens. Kidney Int 2017;91(4):787-789
6. Shi
Y, Glaser KJ, Venkatesh SK, et al. Feasibility of using 3D MR elastography to
determine pancreatic stiffness in healthy volunteers. J Magn Reson Imaging
2015;41(2):369-375
7. Wang
J, Shan Q, Liu Y, et al. 3D MR Elastography of Hepatocellular Carcinomas as a
Potential Biomarker for Predicting Tumor Recurrence. J Magn Reson Imaging 2018