Korel Dursun Yildirim1,2, Christopher Bruce1, Rajiv Ramasawmy1, Kendall O'Brien1, Adrienne Campbell-Washburn1, Daniel Herzka1, Robert J. Lederman1, and Ozgur Kocaturk1,2,3
1Cardiovascular Branch, Division of Intramural Research, National Heart Lung and Blood Institute, National Institutes of Health, Bethesda, MD, United States, 2Institute of Biomedical Engineering, Bogazici University, Istanbul, Turkey, 3Transmural Systems, Andover, MA, United States
Synopsis
A clinical-grade active MRI 0.035” guidewire design
with a curved distal tip geometry and continuous shaft signal ensuring the
mechanical and electrical safety, was introduced. Proposed design was tested in-vitro and in-vivo for MRI visibility and mechanical
performance, and in-vitro for RF induced heating.
INTRODUCTION
A guidewire is a fundamental tool for image-guided minimally
invasive procedures. Previously, it was shown that an active guidewire design
based on a monopole loopless antenna is ideal for interventional MRI
applications thanks to its simplicity, miniaturization and high near-field SNR1-4.
The only drawback of the loopless antenna design is that the sensitivity
profile gradually decreases to zero towards to the tip, Therefore, the distal
end of the guidewire becomes invisible under MRI. This problem needs to be
fixed to ensure operational safety. Extending the antenna whip and packing the
extended conductor as a tight-pitched coil helps to overcome this problem3-5.
A clinical grade MRI guidewire must have sufficient mechanical properties to
navigate in tortuous vasculature ensuring the electrical and mechanical safety
in addition to the continuous conspicuity6. However, the literature
lacks an active guidewire design having a curved distal tip with a complete whip
conspicuity. In this study, we introduce a novel active MRI guidewire design
suitable for clinical applications, having a curved tip and comparable
mechanical characteristics to the commonly used commercial guidewires, and a
continuous sensitivity profile along the whip, providing a complete tip geometry
information under a prototype 0.55T MRI system.METHODS
The active guidewire was designed and manufactured
using medical grade MRI compatible materials. The length is 162,5 cm and the
diameter is 0.89 mm (0.035”) for the compatibility with the conventional
cardiovascular catheters. A nitinol wire with a changing profile (0.127mm min. and 0.508mm max. diameter) and a
curved tip geometry (Trutech, MI, USA) served as the inner conductor of
the monopole antenna. The inner conductor was coated with a Pebax tube (Microcatheter Components, NH,
USA) and placed in a nitinol hypotube (Memry, CT, USA) which is used as
the shield of the coaxial shaft. The inner conductor was extended 37.5 cm beyond
the nitinol hypotube to form the whip. A tightly wound insulated alloy solenoid
coil was electrically connected to the distal end of the whip. The subassembly
was finally insulated with a 0.8mm ID, 0.89mm OD Pebax tube. The coaxial shaft
was connected to a tune/match circuit via a detachable connector and then to
the MRI scanner, to enable the catheter exchange during an interventional
procedure which is expected from the conventional guidewires (Figure 1).
A 0.076mm diameter MP35N wire with 0.02mm insulation
thickness and a 0.05mm diameter 35N LT wire with 0.005mm insulation thickness were
used to form 0.71mm OD coils (Figure 2.a and 2.b) to assess the effects of the
number of turns and the coil length on the signal intensity. Coil length was optimized
for the continuous whip signal providing the curve information.
Active guidewire prototypes were tested in-vitro
for MRI visibility in a gel phantom prepared per ASTM 2182-11a7 using a bSSFP sequence (TE/TR=1.11/4ms, flip angle=45°, FOV=400mm, matrix=192×144,
slice thickness=8mm) at a prototype 0.55T MRI Scanner (Siemens,
Erlangen, Germany). After the visibility tests, the guidewire design with MP35N
coil was discarded and RF induced heating performance tests were performed using
the same bSSFP sequence with additional 65° and 75° flip angles. In-vitro
mechanical performance was assessed via push-ability, 3-point bending and
torquability tests and compared to a commercial clinical-grade guidewire
(Glidewire, Terumo, NJ, USA).
In-vivo MRI visibility and mechanical performance were tested in a swine
using the same bSSFP sequence used for in-vitro imaging tests, at 0.55T, complying
with the local animal care and use committee guidelines. the active guidewire
signal was artificially highlighted on the image using a custom real-time MRI
reconstruction tool.RESULTS
After the coil length optimization, 50mm long MP35N solenoid
coils provided a distinct tip signal, and 14mm long 35N LT coils provided a continuous
signal for the entire whip (figure 2.c and 2.d). RF induced temperature rise at
the guidewire tip (the hot spot) was 1.64C°, 3.41C° and 4.64C° for the 45°, 65° and 75° flip angles,
respectively (Figure 3). Mechanical properties of the guidewire prototype were
comparable to the commercial equivalent (Figure 4). Left ventricle of the swine
heart was accessed retrograde via femoral entry using the active guidewire
prototype, successfully (Figure 5).DISCUSSION
Extended whip design by connecting a solenoid coil
helps to balance and even reduce the signal intensity attenuation towards the distal
end of the whip. The number of turns and the length of the tip coil alter the
effective wavelength so that the sensitivity profile of the whip. The desired
signal signature can be obtained adjusting these parameters. The curved tip
geometry of a guidewire is found very helpful by the operators to navigate
through the vascular system. The distinct tip signal of the MP35N coils lacked
the tip geometry information while the 35N LT coils provided continuous tip curvature
signal which is crucial for operational safety and expected by the operators. The
detachable connector allows intra-operational device exchanges over the proposed
active guidewire design.CONCLUSION
In this study, we have successfully developed a
clinical-grade, low-field active MRI guidewire design with a curved tip
geometry and continuous whip signal ensuring the mechanical and electrical
safety which is essential to be used for clinical MRI guided minimally invasive
operations.Acknowledgements
We would like to acknowledge the
assistance of Siemens Healthcare in the modification of the MRI system for
operation at 0.55T under an existing cooperative research agreement (CRADA)
between NHLBI and Siemens Healthcare.This study was supported by the Division of Intramural Research, National Heart, Lung, and BloodInstitute, National Institutes of Health, USA (Z01-HL005062 and Z01-HL006061).References
1. Ocali O, Atalar E. Intravascular magnetic resonance imaging using a loopless catheter antenna. Magn Reson Med. 1997;37(1):112-8.
2. Kocaturk O, Kim AH, Saikus CE, Guttman MA, Faranesh AZ, Ozturk C, et al. Active two-channel 0.035'' guidewire for interventional cardiovascular MRI. J Magn Reson Imaging. 2009;30(2):461-5.
3. Sonmez M, Saikus CE, Bell JA, Franson DN, Halabi M, Faranesh AZ, et al. MRI active guidewire with an embedded temperature probe and providing a distinct tip signal to enhance clinical safety. J Cardiovasc Magn Reson. 2012;14:38.
4. Susil RC, Yeung CJ, Atalar E. Intravascular extended sensitivity (IVES) MRI antennas. Magn Reson Med. 2003;50(2):383-90.
5. Karmarkar PV, Kraitchman DL, Izbudak I, Hofmann LV, Amado LC, Fritzges D, et al. MR-trackable intramyocardial injection catheter. Magn Reson Med. 2004;51(6):1163-72.
6. Rogers T, Lederman RJ. Interventional CMR: Clinical applications and future directions. Curr Cardiol Rep. 2015;17(5):31.
7. International A. ASTM 2182-11a Standard Test Method for Measurement of Radio Frequency Induced Heating On or Near Passive Implants During Magnetic Resonance Imaging.