Yangzi Qiao1, Chao Zou1, Chuanli Cheng1, Changjun Tie1, Xin Liu1, and Hairong Zheng1
1Shenzhen Institutes of Advanced Technology, Chinese Academy of Sciences, Shenzhen, China
Synopsis
Simultaneous
MR acoustic radiation force imaging and MR thermometry (STARFI) based on
coherent echo-shifted sequence (cES) was proposed and comprehensively compared
to RF spoiled gradient echo (spGRE). The calculated displacement of cES STARFI was always
larger than the value of spGRE STARFI through both the simulation and
experiments, while the accuracy of the temperature monitoring of cES was
maintained. The temperature and displacement map acquired during HIFU heating
were in good accordance with each other. The cES STARFI can be an alternative for comprehensively monitoring of
HIFU treatment with increased displacement sensitivity and time efficiency
compared to spGRE STARFI.
Introduction
MR
acoustic radiation force imaging (MR-ARFI) and MR thermometry (MRT) represent
the mechanical and thermal effects, which are two of the most important effects
of high intensity focused ultrasound (HIFU) therapy. To comprehensively monitor
and evaluate HIFU treatment, simultaneous MR-ARFI and MRT imaging (STARFI) have
been proposed since 20121-5. Most of the STARFI researches were
based on RF-spoiled GRE (spGRE), which was the most frequently used MRT
sequence. In this study, STARFI based on coherent echo shifted (cES) sequence was
proposed and comprehensively compared to the spGRE STARFI. The echo shifting
gradients of cES makes the sequence motion sensitive. It was assumed that
multiple pathways could contribute to the displacement accumulated phase
without RF spoiling, making this sequence more sensitive to the HIFU pulse
induced displacement. Simulation and experiments validated our hypothesis. The
increased displacement sensitivity might lower the needed acoustic power for
focalization and improve the safety of HIFU treatment. Theory and methods
Fig 1. showed
the cES and the spGRE sequences in our study. The HIFU pulse was interleaved
ON/OFF in successive TRs. Based on the assumption that the temperature change
between Echo 1 and Echo 2 was negligible, displacement could be separated by
phase subtraction between the two echoes: $$D_{cES}=\frac{\triangle\phi_D}{2\pi\gamma M_D}=\frac{\phi_{echo2}-\phi_{echo1}}{2\pi\gamma M_D}$$ $$D_{spGRE}=\frac{\triangle\phi_D}{2\pi\gamma M_D}=\frac{\phi_{echo1}-\phi_{echo2}}{2\pi\gamma M_D}$$ ,where $$$\gamma$$$ is the
gyro-magnetic ratio, $$$M_D$$$ is the moment
of displacement encoding gradient (DEG). Four pairs of images before HIFU
sonication were acquire to ensure that the steady state of MR signal had been
reached. Image of pair #4 was used as the reference for temperature calculation,
according to the following formulas: $$\triangle T_{cES}=\frac{\triangle \phi_T}{\alpha2\pi\gamma B_0TE_{eff}}=\frac{\phi_{echo1}-\phi_{ref_-echo1}}{\alpha2\pi\gamma B_0(TE+TR)}$$ $$ \triangle T_{spGRE}=\frac{\triangle \phi_T}{\alpha2\pi\gamma B_0TE}=\frac{\phi_{echo2}-\phi_{ref_-echo2}}{\alpha2\pi\gamma B_0TE}$$, where $$$\alpha$$$ (-0.01 ppm/ºC) is the PRFS coefficient6,
B0 is the main magnetic field strength. Bloch equation was used to
simulate the influence of T1/T2 on the calculated displacement of cES and spGRE
STARFI. The instant tissue displacement at the focus was modeled based on
overdamped harmonic response7. All MRI scans were performed on a 3.0T
MR system (uMR790, Shanghai United Imaging Healthcare, Shanghai, China). The
system set-ups was shown in Fig 1(c). The influence of the toffset on the
displacement map was studied first to determine the optimal toffset used in the
following experiments. The acoustic power was increased linearly to analysis
the displacement sensitivity of these two sequences. Then the HIFU heating
experiment was carried out, and monitored by both the cES and spGRE. The
measurement differences between these two methods were compared. Results
The Bloch
simulation of displacement based on cES and spGRE STRAFI were given in Fig. 2 (a)
and (b). The results demonstrated that the displacement obtained by cES was
always larger than that obtained by spGRE STARFI. The displacement measured
from cES STARFI becomes larger with prolonged T1. Comparatively, the displacement
of spGRE STARFI shows little dependence on T1/T2.
As
illustrated in Fig.3, the toffset between the start time point of HIFU and the displacement
encoding gradient (DEG) would have a significant influence on the calculated
displacement of both cES and spGRE STARFI, while the input HIFU power and
duration were kept constant. Figure 3(a) and (b) are the simulated and
experimental displacements of cES and spGRE STARFI. The trends of
simulation and experiments were in good accordance.
Fig 4
compared the displacement detected by cES and spGRE STARFI by linearly
increasing the HIFU pressure. The measured apparent displacement was linearly
increased with acoustic pressure for both sequences. However, their regression
slopes were different. In the case of cES STARFI, the slope was 1.52 μm/MPa,
while the slope of spGRE STARFI was only 0.90 μm/MPa.
Fig. 5(a)
revealed that the displacements calculated by cES STARFI were always larger
than that of spGRE STARFI. Fig. 5(b) showed the temperature curves during the
heating and cooling process. The maximum temperature rise in the focus was 13.44
± 0.22 ºC when
calculated by cES STARFI and 12.34 ± 0.09 ºC when calculated by spGRE
STARFI at the end of HIFU sonication. The difference between the maximum
temperature change of these two sequences was 1.10 ºC. The temperature change and
the instantly acquired displacement curves showed great consistency between
both sequences.Discussion
Simultaneous MR-ARFI
and MRT imaging based on cES was comprehensively compared to spGRE. The cES is
more time efficient than spGRE STARFI, as cES allows TE to be greater than TR.
Besides, the echo shifting gradients of cES was intrinsically motion sensitive,
making the sequence suitable for MR-ARFI. In cES STARFI, gradient C and D were naturally
separate in time. This time interval can facilitate to accumulate sufficient
tissue displacement before the start of the DEG. Without RF spoiling, multiple
pathways can contribute to the displacement induced phase change. Both the
simulation and experiment results demonstrated that with the same HIFU input
power and DEG moment, the apparent displacement obtained by cES STARFI was
larger than spGRE. On the other hand, the HIFU heating experiment verified that
the accuracy of the temperature monitoring was maintained.Conclusion
The
sequence could be an alternative for comprehensively monitoring MR guided
focused ultrasound treatment. The increased displacement sensitivity allows for
lowering the needed acoustic power for focalization and improve the safety of
HIFU treatment. Acknowledgements
This work was supported by the Key Laboratory for Magnetic
Resonance and Multimodality Imaging of Guang-dong Province (No.
2014B030301013), the National Natural Science Foundation (Nos. 81327801,
81527901,11504401) References
1. Auboiroux
V, Viallon M, Roland J, Hyacinthe JN, Petrusca L, Morel DR, Goget T, Terraz S,
Gross P, Becker CD, Salomir R. ARFI-prepared MRgHIFU in liver: Simultaneous
mapping of ARFI-displacement and temperature elevation, using a fast GRE-EPI
sequence. Magnet Reson Med 2012;68(3):932-946.
2. Kaye
EA, Pauly KB. Adapting MRI acoustic radiation force imaging for in vivo human
brain focused ultrasound applications. Magnet Reson Med 2013;69(3):724-733.
3. Mougenot
C, Waspe A, Looi T, Drake JM. Variable ultrasound trigger delay for improved
magnetic resonance acoustic radiation force imaging. Phys Med Biol
2016;61(2):712-727.
4. Bour
P, Marquet F, Ozenne V, Toupin S, Dumont E, Aubry JF, Lepetit-Coiffe M, Quesson
B. Real-time monitoring of tissue displacement and temperature changes during
MR-guided high intensity focused ultrasound. Magnet Reson Med 2017, 78(5):
1911-1921.
5. de
Bever JT, Odeen H, Hofstetter LW, Parker DL. Simultaneous MR thermometry and
acoustic radiation force imaging using interleaved acquisition. Magnet Reson
Med 2018;79(3):1515-1524.
6. Poorter
JD, Wagter CD, Deene YD, Thomsen C, Ståhlberg F, Achten E. Noninvasive MRI
thermometry with the proton resonance frequency (PRF) method: in vivo results
in human muscle. Magnet Reson Med 1995;33(1):74-81.
7. Souchon
R, Salomir R, Beuf O, Milot L, Grenier D, Lyonnet D, Chapelon J-Y, Rouvière O.
Transient MR elastography (t-MRE) using ultrasound radiation force: Theory,
safety, and initial experiments in vitro. Magnet Reson Med 2008;60(4):871-881.