Peng Cao1, Di Cui1, Queenie Chan2, and Edward S. Hui1
1Diagnostic Radiology, The University of Hong Kong, Hong Kong, China, 2Philips Healthcare, Hong Kong, China
Synopsis
We proposed a new multi-inversion-recovery (mIR) myelin water mapping (MWF)-MRF sequence
that allows 24 s/slice scan speed for four-compartment (myelin water, cerebrospinal
fluid, gray matter and white matter)
brain
mapping on clinical 3T MRI.
Purpose
Myelin or myelin water mapping by MRI
is highly desirable for monitoring the demyelination and remyelination
processes. Such myelin measurement by MRI was conventionally based on the measurement
of 16 to 32 echoes to depict the T2 differences between tissue compartments,
namely free water and myelin water 1–3. The myelin water has a significantly shorter T1 and T2 than free
water, which allows the myelin water to be distinguished by multi-compartment
analysis on MRF data 1–3. The multi-compartment analysis is a recent key breakthrough in MRF
methodology 4. However, in the previous study, the measurement was based on the
combination of two separated MRF scans, one without delays after repetitions
and one with 5s-delays 4. It is desirable to further simplify and integrate the two scans into
one. This study aims to develop an MRF sequence sensitized for myelin water
fraction measurement (MWF-MRF). Methods
We used a 3T human MRI scanner (Achieva TX, Philips Healthcare) with an 8-channel
brain coil for signal reception in the brain. The MRF was based on an inversion recovery steady-state
precession (IR-FISP) sequence with variable flip angles (FA) and repetition
times (TR) 5. A constant speed spiral readout trajectory with an acquisition window
of 8.4 ms and acquisition factor = 58.4% was used. The trajectory was rotated
by 222.5° after each dynamic. A spiral-in/out trajectory was used for increasing
sensitivity while reducing the susceptibility effect 6. This
new multi-IR MRF sequence (in Figs. 1 and 2) is sensitive enough to probe myelin content in the
brain. The magnetization obtained after repeatedly applying the IR pulse is
more specific to the myelin water signals due to the inversion and prolonged
recovery of long T1 signal. Therefore, we propose to use this new MRF sequence
for rapid measurement of brain myelin water content in vivo. Our preliminary in
vivo experiment used the following imaging parameters: four IR pulses applied
at the 1, 501, 1001, and 1501 TR, TR = 12.1 to 14.1 ms, FA = 0 to 60°, field
of view = 300 × 300 mm2, acquisition matrix = 256 × 256, image
resolution = 1.17 × 1.17 mm2, and slice thickness = 5 mm. The
quantification was based on an in-house implementation of the non-negative least-square (NNLS)
algorithm with a reweighting iteration for updating the joint distribution of
the T1/T2 parameter across the slice/volume 4.Results
Figure
3 shows the multi-compartment myelin water mapping of a healthy volunteer using
our proposed multi-IR MRF. The four-compartment brain maps were obtained from a
pixel-wise separation. More importantly, the myelin water could be reliably and
rapidly measured using the proposed multi-IR MRF with a scan time of 24 s/slice.
The T1 and T2 ranges of the four compartments were 202 ms < T1 < 363 ms
and 14 ms < T2 < 37 ms for myelin water, 1200 ms < T1 < 2630 ms and
28 ms < T2 < 300 ms for gray matter, 628 ms < T1 < 1211 ms and 34 ms
< T2 < 73 ms for white matter, and 1200 ms < T1 < 2630 ms and 28 ms
< T2 < 300 ms for myelin water, and 557 ms < T1 < 3768 ms and 856 ms
< T2 < 1820 ms for cerebrospinal fluid. The method achieved maximal
spatial separation of compartments as determined empirically from the MRF data
in vivo. Figure 4 compared the MWF-MRF with or without mIR insertions. The mIR
obviously improved the detection of myelin water content.Discussion
The proposed multi-IR scheme can repeatedly
invert the signal of the long T1 compartment, which enhances the detection of
myelin water. In this
project, the mIR MWF-MRF and
multi-compartment analysis successfully estimated the volume fraction of myelin
water, gray matter, white matter, and cerebrospinal fluid by four-compartment
separation and quantification.Conclusion
We have successfully developed a new mIR MWF-MRF sequence.
The myelin water content could be reliably and rapidly measured using the
proposed multi-IR MRF with a scan time of 24 s/slice.Acknowledgements
This work is supported by HKU URC seed fund.References
1. Laule, C. et al. Water content
and myelin water fraction in multiple sclerosis: A T 2 relaxation study. J.
Neurol. (2004). doi:10.1007/s00415-004-0306-6
2. Du, Y. P. et al. Fast multislice
mapping of the myelin water fraction using multicompartment analysis of T2*
decay at 3T: A preliminary postmortem study. Magn. Reson. Med. (2007).
doi:10.1002/mrm.21409
3. Laule, C. et al. Myelin water
imaging of multiple sclerosis at 7 T: Correlations with histopathology. Neuroimage
(2008). doi:10.1016/j.neuroimage.2007.12.008
4. Nagtegaal, M., Koken, P., Amthor, T.
& Doneva, M. Fast multiācomponent analysis using a joint sparsity constraint
for MR fingerprinting. Magn. Reson. Med. 1–14 (2019).
doi:10.1002/mrm.27947
5. Jiang, Y., Ma, D., Seiberlich, N.,
Gulani, V. & Griswold, M. A. MR fingerprinting using fast imaging with
steady state precession (FISP) with spiral readout. Magn. Reson. Med. 74,
1621–1631 (2015).
6. Glover, G. H. & Law, C. S.
Spiral-in/out BOLD fMRI for increased SNR and reduced susceptibility artifacts.
Magn. Reson. Med. 46, 515–522 (2001).