Mareike Alicja Buck1 and Matthias Günther1,2
1Fraunhofer MEVIS, Bremen, Germany, 2MR-Imaging and Spectroscopy, Faculty 01 (Physics, Electrical Engineering), University Bremen, Bremen, Germany
Synopsis
This
abstract compares quantified perfusion values of the standard model and a new
dispersion model based on an AIF using the ASLIF-sequence. For different ATTs, the
voxel´s signal was simulated using the dispersion model. The simulations show
that the standard model overestimates the signal. This may result from lack of
dispersion effects especially in the inflow phase of the labeled bolus. Consequently,
the determined perfusion values vary for different ATTS. Thus, using an AIF
based on an acquired patient specific reference bolus instead could improve the
stability and robustness of quantified perfusion values.
Introduction
Getting reliable perfusion quantification
results is one of the biggest challenges in Arterial Spin Labeling (ASL). Hereby,
the underlying assumption about the shape of the Arterial Input Function (AIF)
of the labeled blood bolus is crucial for robustness and stability. The boxcar
shaped AIF generally used in the standard kinetic model1 is a
well-known over-simplification. In fact, often the label bolus is not exactly
rectangular, sometimes imperfect and most importantly, disperses during inflow.
To establish ASL in clinical applications it is essential to encounter these
effects in any quantification model to improve the reliability of quantified
perfusion values. By utilizing the ASLIF-sequence2, label boli at
two different positions were measured and were used to predict dispersion at a
later arterial transit time (ATT). This will ultimately allow for the
estimation of the AIF at voxel level.Methods
Imaging:
The perfusion phantom QASPER3 was
examined with an in-house developed ASLIF sequence2,4,5 on a 3T MR
scanner (MAGNETOM Skyra, SIEMENS Healthineers AG). The ASLIF images were acquired
with the following parameters: 4 Hadamard encoding cylces, subbolus duration
(SBD) = 1400ms, post labeling delay (PLD)=1000ms, TR=6.5s, TRpCASL=1420us,
phase shift = π/2 -23* π/2. The phantom had a flow rate of 350ml/min. The AIF
within the main inflow vessel was determined from 12 different positions resulting
in 12 different delay times (see Fig. 1). In addition, a so-called reference
bolus was measured directly behind the labeling slice to yield the basis for
the dispersion model.
Simulation:
A model is developed which simulates the ATT dependent
shape of the label bolus during the inflow. Using this model, the signal in a
voxel was simulated for ATTs between 500ms and 1600ms. The simulated
measurements at three inflow times (TI) (2000ms, 3000ms and 4000ms) were then
used to determine the quantified perfusion values with the standard model (MatLab,
MathWorks, Natick, MA, USA). For the calculation it is assumed that the ATT is
known and only the perfusion f needs to be determined. The following simulation
parameters were used: f=60 ml/100g/min, T1_b=1500ms, T1_tis=2000ms, lambda=90
ml/100g, M0=500, ATT=500-1600ms, BD=1400ms, alpha=0.9.Results
In Figure 2 the result
of the developed dispersion model compared
to the measurement signal is presented. A nearly identical reproducibility
of the measurement can be found.
The table
in Figure 3a shows the fitted quantified perfusion values of the standard model
for different ATTs. The determined perfusion values in the range of ATT=700-1300ms
are initially overestimated with the standard model but approach the adjusted
value of the dispersion model with increasing ATT (Fig. 3b). The observed
relative standard deviation is in the range of 12% with a mean value of 6% over
all ATTs. Figure 4 presents the fitted signal of the standard model with the quantified
perfusion values resulting from the standard model fit compared to the simulated
signal of the dispersion model. In all four cases, the comparison shows that
the signal is overestimated during the arrival phase of the labeled bolus in
the standard model compared to the dispersion model.Discussion & Conclusion
The
dispersion effects of the label bolus are not considered in the standard model.
This could explain the relative error of 6 % in the quantified perfusion
values. Furthermore, the phase in which the labeled bolus arrives in a voxel
cannot be reproduced very well with the standard model (Fig. 4).
For
ATT=1300ms the signal of the dispersion model and the standard model have
nearly identical parameters. However, at this point the signal of the inflow
phase of the standard model is overestimated. This shows that quantification
errors may occur by using the standard model. An AIF based on an acquired
reference bolus can take dispersion effects into account instead of a boxcar
shaped AIF.
The
comparison of the two models shows how important it is to consider dispersion
effects for a robust quantification.
Using the
ASLIF sequence, it is possible to measure a reference bolus directly behind the
labeling slice. This reference bolus represents the individual resulting AIF of
the labeling. By using this reference bolus with the dispersion model instead
of a boxcar shaped AIF, it can be possible to detect and extrapolate dispersion
effects, disturbances or deviations in the labeling process. The incorporation of
a reference bolus for an ATT dependent AIF could improve the robustness and
stability of perfusion quantification and may therefore be an important step towards
the clinical application of ASL.Acknowledgements
References
1. Buxton
RB, Frank LR, Wong EC, Siewert B, Warach S, Edelmann RR. A general kinetic
model for quantitative perfusion imaging with arterial spin labeling. Magnetic
resonance in medicine. 1998;40(3):383-396.
2. Guenther
M. Arterial Spin Labeled Input Function (ASLIF): signal acquisition during
pseudo-continuous arterial spin labeling in Proceeding of the 26th Annual
Meeting of ISMRM; 2018; Paris, France.
3. Oliver-Taylor A, Goncalves M, Hampshire T, Davis B, Daga P, Evans L, Bainbridge A, Wheeler-Kingshott C, Sokolska M, Thornten J, Vita ED, Golay X. A Calibrated Perfusion Phantom for Quality Assurance of Quantitative Arterial Spin Labelling in Proceeding of the 25th Annual Meeting of ISMRM; 2017; Honolulu, USA.
4. Günther
M, Highly efficient accelerated acquisition of perfusion inflow series by cycled
arterial spin labeling in Proceeding of the 15th Annual Meeting of ISMRM; 2007;
Berlin, Germany.
5. von
Samson-Himmelstjerna F, Madai VI, Sobesky J, Guenther M. Walsh-ordered hadamard
time-encoded pseudocontinuous ASL (WH pCASL). Magnetic resonance in medicine. 2016;76(6):1814-24.