Isabell K. Bones1, Suzanne L. Franklin1,2, Anita A. Harteveld1, Matthias J.P. van Osch2, Sophie Schmid2, Jeroen Hendrikse3, Chrit Moonen1, Marijn van Stralen1, and Clemens Bos1
1Center for Image Sciences, University Medical Center Utrecht, Utrecht, Netherlands, 2C.J.Gorter Center for High Field MRI, Department of Radiology, Leiden University Medical Center, Leiden, Netherlands, 3Department of Radiology, University Medical Center Utrecht, Utrecht, Netherlands
Synopsis
VSASL for renal application has constraints on the
cut-off velocity, as low Vc in the presence of respiratory motion
causes spurious labeling of moving tissue. With higher Vc, label
could be generated more upstream in the vascular tree, potentially introducing
ATT sensitivity. To study label dynamics of renal VSASL using a Vc
compatible with free-breathing (Vc of 10cm/s), data at multiple time
points were acquired. High ASL signal was already observed at early time
points, thus supporting that spatially
non-selective VSASL in the kidney generates label close to, or even inside the target tissue,
also using a free-breathing compatible Vc.
INTRODUCTION
Velocity selective arterial spin labeling (VSASL)
is a flow-based technique that magnetically labels blood with a velocity
exceeding a cut-off velocity Vc.1–3 Choosing a low Vc , label is thought even
to be generated in small vessels inside the tissue, in theory reducing the
dependence of ASL on arterial transit time (ATT).3,4 So far, for quantification of VSASL, Buxton’s pulsed
ASL model has been utilized assuming a negligible ATT. However, for ASL of the
kidneys, higher Vc are generally chosen than in e.g. the brain,
since low Vc in the presence of respiratory motion causes spurious
labeling of moving tissue.5 The label front could therefore shift more upstream
in the vascular tree, potentially introducing ATT sensitivity. In this study,
we acquired renal VSASL data at multiple time points (t) to assess its label dynamics and compared these to the label
dynamics of conventional pulsed (FAIR, flow alternating inversion recovery) and
pseudo-continuous ASL (pCASL).METHODS
Imaging: Kidneys of 6 healthy subjects (age 25-31, 2
men) were scanned at 1.5T (Ingenia, Philips, The Netherlands) using a
28-element phased-array receiver-coil during free-breathing. Five coronal-oblique
slices were acquired with a gradient echo EPI readout (Table 1) using single VSASL3, balanced pCASL6 and FAIR7. Sequence diagrams are illustrated in Figure 1.
Experiments: ASL data was acquired at six time points [t: 400, 800, 1200, 2000, 2600, 3200ms] for
single VSASL (Vc=10cm/s, gradient encoding=anterior-posterior), FAIR
and pCASL. For pCASL, the label duration varied for measurements at different t to allow for detection of short bolus
arrival times, resulting in short post labeling delays (PLDs) (Table 2). In
view of short PLDs we did not apply background suppression pulses for any ASL
technique. For all ASL techniques, imaging included vascular crushing gradients
to reduce venous
contributions and blood volume weighting from larger vessels: cut-off velocity
10cm/s in anterior-posterior direction. Per subject, a T1-map8 and M0-image were
acquired.
Analysis:
Images were aligned
using Elastix9 with a group based principal component approach10, for each kidney separately. Subtraction images
(ΔM) of all VSASL repetitions were checked for spurious labeling by visual
inspection. A renal cortex region-of-interest was obtained from
the T1-map. Dynamic behavior of
tagged blood was assessed by calculating the perfusion-weighted signal
(PWS=ΔM/M0×100%) averaged over the cortex region-of-interest for each t. Renal blood flow (RBF)
was calculated by fitting Buxton’s general kinetic model11 to the multiple time
point data. VSASL data was quantified using Buxton’s pulsed model including
correction for diffusion weighting.12 Group averages and inter-subject variation were
calculated. Statistical testing was done using Wilcoxon’s signed rank test with
α=0.05.RESULTS
Data of
all subjects were included in the analysis. With Vc=10cm/s and
gradient encoding in anterior-posterior direction no spurious labeling in VSASL
subtraction images was found. For VSASL the PWS measured at the earliest t=400ms
is already relatively high with 92% of the peak and rises slightly until it reaches
the peak at 800ms, with a steady decrease for increasing t (Figure 2A&D). In contrast, for FAIR
(Figure 2B&E) and pCASL (Figure 2C&F) the PWS starts
off lower with 28% and 12% of the peak, respectively (p<0.05). Also, the peak
appears later, around 1200ms. Overall, FAIR delivered the highest peak PWS of
4.31±0.60%, followed by pCASL 2.29± 1.18% and VSASL 1.90±0.21% (p<0.05).
Fitting of ASL data from multiple time points for all three techniques
yielded the following group average cortical RBF values for FAIR 385±39 mL/min/100g, VSASL 247±33 mL/min/100g and pCASL 228±61 mL/min/100g (Figure 3).DISCUSSION & CONLUSION
For VSASL, we found high PWS already
at early time points, quickly
reaching the peak, followed by constant signal decrease. These findings support
label generation close to, or even inside the target tissue with rapid inflow to
the cortex. This is in agreement with renal physiology, as even in interlobar
arteries velocities around 28cm/s were reported.13 Other factors
could contribute to high PWS at early time points. First, blood volume
weighting of ΔM,
originating from label generation in blood already in the tissue, instead of
inflow dependent perfusion weighting, even though vascular crushing was
applied. Second, eddy currents could result in subtraction artefacts at early
time points even with a minimum t of 400ms. Third, diffusion
contribution to the VSASL signal is stronger for short t, though
expected rather small with 0.05% of M0 for Vc=10cm/s.12
Our results for label dynamics of VSASL
in the kidney with a Vc of 10cm/s are similar to those of VSASL in
the brain with a low Vc of 2cm/s.4 Also here
VSASL label dynamics differ from conventional, spatially-selective ASL techniques
FAIR and pCASL.
We modified Buxton’s pulsed model
with diffusion weighting correction and found lower RBF for VSASL than FAIR.
However, as opposed to pCASL and FAIR, the PWS time curve for VSASL has few
features that anchor the model fit, using BD and ATT as free parameters,
challenging fit accuracy. Assuming a blood volume-weighted signal contribution,
the model should be extended for that. The results of this study support that spatially
non-selective VSASL brings the labeling location closer to, or even inside the
target tissue. This underlines its potential as an ATT insensitive technique,
also for renal ASL where respiratory motion limits the cut-off velocity to
higher values.Acknowledgements
This work is part of the research
program Applied and Engineering Sciences with project number 14951 which is
(partly) financed by the Netherlands Organization for Scientific Research
(NWO). We thank MeVis Medical Solutions AG (Bremen, Germany) for providing
MeVisLab medical image processing and visualization environment, which was used
for image analysis.References
1. Norris DG, Schwarzbauer
C. Velocity Selective Radiofrequency Pulse Trains. J Magn Reson. 1999;137:231-236.
2. Duhamel
G, De Bazelaire C, Alsop DC. Evaluation of
systematic quantification errors in velocity-selective arterial spin labeling
of the brain. Magn Reson Med. 2003;50:145-153.
3. Wong EC, Cronin M, Wu WC,
Inglis B, Frank LR, Liu TT. Velocity-selective arterial spin labeling. Magn
Reson Med. 2006;55:1334-1341.
4. Guo J, Wong EC. Increased
SNR efficiency in velocity selective arterial spin labeling using multiple
velocity selective saturation modules (mm-VSASL). Magn Reson Med. 2015;74(3):694-705.
5. Bones
IK, Franklin SL, Harteveld AA, et al. Influence
of labeling parameters of velocity selective arterial spin labeling for renal
perfusion imaging. In: Proceedings of the 27th Annual Meeting of ISMRM
2019:4948.
6. Dai W, Garcia D, De
Bazelaire C, Alsop DC. Continuous flow-driven inversion for arterial spin
labeling using pulsed radio frequency and gradient fields. Magn Reson Med.
2008;60(6):1488-1497.
7. Kim S-G. Quantification
of Relative Cerebral Blood Flow Change by Flow-Sesitive Alternating Inversion
Recovery (FAIR) Technique: Application to Functional Mapping. Magn Reson Med.
1995;34:293-301.
8. Clare S, Jezzard P. Rapid
T1 mapping using multislice echo planar imaging. Magn Reson Med.
2001;45:630-634.
9. Klein S, Staring M,
Murphy K, Viergever MA, Pluim J. elastix: A Toolbox for Intensity-Based Medical
Image Registration. IEEE
Trans Med Imaging.
2010;29(1):196-205.
10. Huizinga
W, Poot DHJ, Guyader JM, et al. PCA-based
groupwise image registration for quantitative MRI. Med Image Anal.
2016;29:65-78.
11. Buxton RB, Frank LR, Wong
EC, Siewert B, Warach S, Edelman RR. A General Kinetic Model for Quantitative
Perfusion Imaging with Arterial Spin Labeling. Magn Reson Med.
1998;40:383-396.
12. Schmid S, Ghariq E,
Teeuwisse WM, Webb A, Van Osch MJP. Acceleration-selective arterial spin
labeling. Magn
Reson Med.
2014;71(1):191-199.
13. Deeg
KH, Wörle K, Wolf A. Dopplersonographische Bestimmung von altersabhängigen
Normalwerten der Flussgeschwindigkeiten und des Resistance-Index in den
Nierenarterien gesunder Kinder. Ultraschall
der Medizin. 2003;24(5):312-322.