Benoît Bourassa-Moreau1, Réjean Lebel1, Guillaume Gilbert2, David Mathieu3, and Martin Lepage1
1Centre d’imagerie moléculaire de Sherbrooke, Département de médecine nucléaire et radiobiologie, Université de Sherbrooke, Sherbrooke, QC, Canada, 2MR Clinical Science, Philips Healthcare Canada, Markham, ON, Canada, 3Service de neurochirurgie, Département de chirurgie, Université de Sherbrooke, Sherbrooke, QC, Canada
Synopsis
The arterial input function measured for brain dynamic contrast-enhanced
MRI is contaminated by the signal contribution of surrounding tissues. This
work corrects these partial volume effects on signal level by using the
surrounding gray matter enhancement to discriminate pure arterial signal. The method
also accounts for the high contrast agent concentration reached in arteries and
veins that leads to signal non-linearity, saturation, and concurrent unwanted $$$T_2^*$$$ effects. This partial volume
correction method is compared to concentration scaling on a digital reference
object and on eight subjects. Better recovery of the arterial first pass and
recirculation are shown.
Introduction
Accurate pharmacokinetic modeling in DCE-MRI depends on the quality of
the arterial input function (AIF). In the brain, the AIF is typically measured
from arteries smaller than the voxel volume. The measured signal has
contributions from both arterial blood and surrounding tissue. These partial
volume effects (PVE) are minimal for the superior sagittal sinus (SSS) where the
venous output function (VOF) can be measured. This VOF can be taken as a
surrogate for the AIF, but it underestimates the transfer constants ($$$K^\mathrm{Trans}$$$), it overestimates distribution volumes ($$$v_e$$$), and it correlates poorly with plasma volumes ($$$v_p$$$).1 Alternatively, partial volume correction
methods (PVCM) scale the AIF with a blood volume fraction to match the VOF’s
area under the curve (AUC). However, scaling at the signal level is required to
account for signal-to-concentration non-linearity.2 We improved this method by using the nearby gray
matter (GM) signal as a reference for tissue contribution to extract the
arterial enhancement. This AIF pipeline will be shared in an open-source toolbox.Methods
Cohort: This study was approved by the institutional
ethics committee. Results are shown for eight participants.
Acquisition: Imaging
was performed on a Philips Ingenia 3.0 T scanner with a 32-channel head
receiver coil. A 3D T1-weighted spoiled gradient-echo (SPGR) sequence was
designed for DCE ($$$\mathrm{TE}=1.85\,\mathrm{ms}$$$; $$$\mathrm{TR}=4\,\mathrm{ms}$$$; flip angle$$$\,=9.2°$$$; $$$\mathrm{FOV}=220\times165\times90\,\mathrm{mm}^3$$$; acquired spatial resolution$$$\,=2.3\times2.3\times2.3\,\mathrm{mm}^3$$$; reconstructed images
matrix$$$\,=96\times96\times39$$$; SENSE factors$$$\,=2.3\times1.2$$$; time resolution$$$\,=2.0\,\mathrm{s}$$$; viewsharing keyhole with 10 s footprint). Dynamic acquisition duration was 4
min, with a half dose (0.05 mL/kg) injection of gadobutrol starting at 60 s.
Simulations: A digital reference object (DRO) of the
extended Tofts-Kety model was produced (QIBA_v4 specifications
3), which in our case included a parametrized Horsfield AIF
4 scaled to typical concentrations reached by
the half dose injection, tissue transverse relaxivity ($$$r_2^*$$$) of 87 mM
-1s
-1,
5 and added noise (SNR of 35). Three rows were
added to the DRO:
- A VOF, generated
by dispersing the AIF with a log-normal vascular transport function,6
- Two AIFs
with a blood volume fraction of 20% or 60% mixed with tissue signals.
All vascular rows incorporate a quadratic blood transverse relaxivity
7 and VOF includes phase variations (molar
susceptibility of $$$3.209\times{10}^{-7}\mathrm{mM}^{-1}$$$).
8 Processing: VOF is taken from SSS voxels and is considered
as a PVE-free reference. Our AIF-PVCM was applied at the measured voxel signal
($$$S_m(t)$$$) level correcting for surrounding tissues with nearby GM signal ($$$S_t(t)$$$)
from a voxel automatically selected from nearest neighbours with minimum AUC within a GM mask derived from
SPM. The corrected complex arterial signal ($$$S_a(t)$$$) was scaled with a range of blood volume
fraction ($$$k$$$) on a voxel-wise basis:
$$S_a(t)=\left[S_m(t)-(1-k)S_t(t)\right]/k$$ Repeated arterial concentration calculation was done to find $$$k$$$ for which the AUC matches that of
the VOF. Three different signal-to-concentration equations ($$$F_{S\to{C}}$$$) were considered:
- The standard
non-linear closed-form SPGR equation,
- The
non-linear open-form equation9 that includes $$$T_2^*$$$ relaxation with quadratic blood relaxivity,
fitted only for arterial-like voxels,
- The complex-form equation, fitted for the VOF of
the SSS as described by Simonis et al.,10 but modified to consider the
quadratic relaxivity of blood.
Figure 1 shows how these processing steps were integrated into a VOF/AIF
selection procedure with Horsfield model fitting
4,11 and recursive cluster analysis.
12Results and Discussion
Figure
2 shows the DRO image, reference and calculated vessel curves. Our semi-automated
voxel selection method successfully selected voxels from the DRO VOF and AIF rows
(Fig. 2A). The complex-form equation was chosen for the final VOF (Fig. 2B). It
recovers the peak concentration that is decreased with the closed-form, sensitive to $$$T_2^*$$$ effects, without the added noise of the
open-form. Uncorrected AIF vascular rows are degraded by PVE (Fig. 2C). AIF-PVCM
does not recover peak AIF non-linearity with concentration based scaling (Fig.
2D); signal based scaling with the closed-form
equation recovers the peak AIF (Fig. 2E) with reduced noise sensitivity compared
to the open-form equation (Fig. 2F). Figure 3 compares the calculated DRO parameters
globally against their true values for the clustered PVE-corrected AIF with
concentration scaling (Fig. 2D), with closed-form signal (Fig. 2E), or directly
with the VOF (Fig. 2B). Closed-form signal scaling results in an estimation of $$$K^\mathrm{Trans}$$$ closer to the line of equality (Fig. 3A) and provides
the highest coefficient of determination for $$$K^\mathrm{Trans}$$$ and $$$v_p$$$ (Fig. 3B,D). Figure 4 shows the detailed voxel selection,
AIF-PVCM and fitting for a representative patient. When comparing the AIF and
VOF curves for all patients, the closed-form signal-based PVCM resulted in an
average blood fraction of (23±6)%, a cardiac output of (6.9±3.0) L/min, a
(5.4±2.2) s shorter time to peak, and an AIF to VOF AUC ratio of (1.01±0.04)
and of (1.4±0.5) for the global acquisition and the first pass, respectively.
It was not clear, however, whether the AIF was less dispersed than the VOF with
a full width at half maximum ratio relative increase of (14±41)%.Conclusion
Robust PVCM is essential
to measure the AIF in small arteries that are free of inflow effect, bolus
delay and dispersion that typically affect the large arteries or veins in the
brain. Our PVCM with closed-form signal scaling recovers the PVE free concentration from small
arteries using surrounding tissue signal. Performance is increased for our DRO
pharmacokinetics against contrast based PVE or VOF surrogate.Acknowledgements
This work was supported by a grant from the Fonds de recherche du Québec (FRQ)–Nature et technologies (2018‐ PR‐206157). B.B.M. acknowledges a scholarship from the National Sciences and Engineering Research Council of Canada.References
1. Footit C, Cron GO, Mercier JF, et al. Optimizing perfusion imaging of brain tumors:
Validation of venous output function used as a surrogate AIF. In: Proc. Intl.
Soc. Mag Reson. Med. Montreal; 2011:2038.
2. Hansen AE, Pedersen
H, Rostrup E, Larsson HBW. Partial volume effect (PVE) on the arterial input
function (AIF) in T 1-weighted perfusion imaging and limitations of the
multiplicative rescaling approach. Magn Reson Med. 2009;62:1055-1059.
3. Barboriak DP.
QIBA_v4 Extended Tofts. https://sites.duke.edu/dblab/qibacontent/. Accessed
August 14, 2019.
4. Georgiou L, Wilson
DJ, Sharma N, Perren TJ, Buckley DL. A functional form for a representative
individual arterial input function measured from a population using high
temporal resolution DCE MRI. Magn Reson Med. 2019;81:1955-1963.
5. Kjølby BF,
Østergaard L, Kiselev VG. Theoretical model of intravascular paramagnetic
tracers effect on tissue relaxation. Magn Reson Med. 2006;56:187-197.
6. Willats L, Connelly
A, Calamante F. Modelling the Bolus Dispersion from DSC-MRI Data. In: Proc.
Intl. Soc. Mag Reson. Med. Berlin, Germany; 2007:1445.
7. Akbudak E, Kotys MS, Conturo TE. Quadraticity and hematocrit dependence of DR2*
AIF signals at 3T: a blood phantom study under physiologic conditions. In:
Syllabus of the ISMRM Workshop on Quantitative Cerebral Perfusion Imaging Using
MRI: A Technical Perspective. Venice, Italy; 2004:10-11.
8. van Osch MJP,
Vonken EPA, Viergever MA, van der Grond J, Bakker CJG. Measuring the arterial
input function with gradient echo sequences. Magn Reson Med. 2003;49:1067-1076.
9. Schabel MC, Parker
DL. Uncertainty and bias in contrast concentration measurements using spoiled
gradient echo pulse sequences. Phys Med Biol. 2008;53:2345-2373.
10. Simonis FFJ,
Sbrizzi A, Beld E, Lagendijk JJW, van den Berg CAT. Improving the arterial
input function in dynamic contrast enhanced MRI by fitting the signal in the
complex plane. Magn Reson Med. 2016;76:1236-1245.
11. Horsfield MA,
Thornton JS, Gill A, Jager HR, Priest AN, Morgan B. A functional form for
injected MRI Gd-chelate contrast agent concentration incorporating
recirculation, extravasation and excretion. Phys Med Biol. 2009;54:2933-2949.
12. Peruzzo D,
Bertoldo A, Zanderigo F, Cobelli C. Automatic selection of arterial input
function on dynamic contrast-enhanced MR images. Comput Methods Programs
Biomed. 2011;104:e148-e157.