Victor Han1 and Chunlei Liu1,2
1Electrical Engineering and Computer Sciences, University of California, Berkeley, Berkeley, CA, United States, 2Helen Wills Neuroscience Institute, University of California, Berkeley, Berkeley, CA, United States
Synopsis
We
present a fully geometric view of multiphoton excitation by taking a particular
rotating frame transformation. In this rotating frame, we find that multiphoton
excitations appear just like single-photon excitations again, and thus, we can
readily generalize concepts already explored in standard single-photon
excitation. With a homebuilt low-frequency (~ kHz) coil, we execute a standard
slice-selective pulse sequence with all of its excitations replaced by their
equivalent two-photon versions. With a multiphoton interpretation of
oscillating gradients, we present a novel way to transform a standard slice-selective
adiabatic pulse into a multiband one without modifying the RF pulse shape itself.
Introduction
Today’s MRI assumes single-photon excitation1. That is, for each nuclear
spin, a single photon accompanies the transition between energy states. This
photon must resonate near the Larmor frequency. Here, we show that, instead of
the usual single-photon resonance, we can excite multiphoton resonances to
generate signal for MRI by using multiple magnetic field frequencies, none of
which is near the Larmor frequency. Only the total energy absorbed by a spin
must correspond to the Larmor frequency. Multiphoton excitation has been
described for NMR2–5 and EPR6,7. For MRI, however,
multiphoton excitation has not yet been well explored. The idea that
oscillating z-direction magnetic fields creates frequency “sidebands” is well
known and has been used occasionally, both in excitation8 and reception9. For excitation, it was used
for simultaneous transmit and receive, and for reception, it was used to
accelerate spatial information acquisition by increasing the receive bandwidth.
However, as we will show, a multiphoton framework paints a more complete
picture than sideband excitation, and with it, we are able to better understand
some of the subtleties involved. Within this framework, it becomes clear that
these multiphoton excitations excite extra resonances, and concepts such slice
selective excitation and adiabatic RF pulses readily generalize.Theory
Standard single-photon excitation occurs when the RF field is
polarized in the xy-plane ($$$B_{1,xy}$$$), perpendicular to the main magnetic
field $$$B_0$$$. Multiphoton effects occur when we add more RF fields along the
z-axis ($$$B_{1,z}$$$), parallel to the main magnetic field $$$B_0$$$. The term
RF is used very loosely to include any oscillation frequency. Although it is
also possible to have multiphoton effects with RF only in the xy-plane
5, the effects are usually
orders of magnitude smaller.
To analyze multiphoton excitation, consider two $$$B_1$$$ fields
with frequencies $$$\omega_{xy}$$$ and $$$\omega_z$$$ along the xy-plane and
z-axis respectively. The total magnetic field in the laboratory frame is
$$B_z=B_0+B_{1,z}\cos{\left(\omega_zt\right)}\hspace{2em}[1]$$
$$B_x=B_{1,xy}\cos{\left(\omega_{xy}t\right)}\hspace{2em}[2]$$
$$B_y=-B_{1,xy}\sin{\left(\omega_{xy}t\right)}\hspace{2em}[3]$$
where the RF field in the xy-plane is clockwise circularly
polarized; $$$B_{1,xy}$$$ and $$$B_{1,z}$$$ are amplitudes of each RF field. In
a clockwise rotating frame with an angular velocity of $$$\omega_{rot}$$$, the
effective $$$B_z$$$ field is
$$B_{z,eff}=B_z-\frac{\omega_{rot}}{\gamma}\hspace{2em}[4]$$
To generate time-invariant $$$B_{z,eff}$$$, we choose
$$\omega_{rot}=\omega_{xy}+\gamma
B_{1,z}\cos{\left(\omega_zt\right)}\hspace{2em}[5]$$
We refer to this rotating frame as the phase-modulated rotating
frame following
Michal
2. This choice of $$$\omega_{rot}$$$
results in a non-stationary xy-plane RF field. To illustrate this effect, we
examine the phase accrual of the xy-RF field in the rotating frame. Let $$$\theta$$$
be the phase, then
$$B_{x,eff}=B_{1,xy}\cos{\left(\theta\right)}\hspace{2em}[6]$$
$$B_{y,eff}=B_{1,xy}\sin{\left(\theta\right)}\hspace{2em}[7]$$
By definition,
$$\theta=\int_{0}^{t}{(\omega_{rot}}-\omega_{xy})dt=\frac{\gamma
B_{1,z}}{\omega_z}\sin{\left(\omega_zt\right)}\hspace{2em}[8]$$
Combining Eqs. [4-8] gives
$$B_{z,eff}=B_0-\frac{\omega_{xy}}{\gamma}\hspace{2em}[9]$$
$$B_{x,eff}=B_{1,xy}\cos{\left(\frac{\gamma
B_{1,z}}{\omega_z}\sin{\left(\omega_zt\right)}\right)}\hspace{2em}[10]$$
$$B_{y,eff}=B_{1,xy}\sin{\left(\frac{\gamma B_{1,z}}{\omega_z}\sin{\left(\omega_zt\right)}\right)}\hspace{2em}[11]$$
Eqs. [10,11] can be Taylor expanded if $$$\frac{\gamma
B_{1,z}}{\omega_z}\ll1$$$ as
$$B_{x,eff}\approx B_{1,xy}\hspace{2em}[12]$$
$$B_{y,eff}\approx B_{1,xy}\frac{\gamma B_{1,z}}{\omega_z}\sin{\left(\omega_zt\right)}\hspace{2em}[13]$$
From Eqs. [9,12,13], three potential resonances exist
depending on the relative values of the frequencies (Fig. 1).
- Single-photon resonance: If $$$\omega_{xy}=\gamma
B_0$$$, then $$$B_{z,eff}=0$$$ and $$$B_{x,eff}$$$ tilts magnetization from the
z-axis. This is the normal on-resonance condition.
- Two-photon resonances: If $$$\omega_{xy}=\gamma
B_0\pm\omega_z$$$, then $$$B_{z,eff}=\mp\frac{\omega_z}{\gamma}$$$. The
linearly polarized RF field $$$B_{y,eff}$$$ oscillates with an angular
frequency of $$$\omega_z$$$, matching the amplitude of $$$B_{z,eff}$$$, which
induces resonances. These two new resonances correspond to state transitions
where a single xy-polarized photon is absorbed and a single z-polarized photon
is absorbed or emitted, depending on whether the xy-frequency is below or above
the Larmor frequency. From Eq. [13], consistent with past work, the effective
angular nutation frequency for two-photon excitation is
$$\omega_{nut}=\frac{\gamma B_{1,xy}\gamma B_{1,z}}{2\omega_z}\hspace{2em}[14]$$
If more terms are kept in the Taylor expansion of Eqs.
[10,11], we can analyze higher-order excitations with three or more photons.
However, Taylor expansion requires $$$\frac{\gamma B_{1,z}}{\omega_z}\ll1$$$.
Alternatively, we can exactly expand $$$B_{x,eff}$$$ and $$$B_{y,eff}$$$ using
Bessel functions, which reveals that for any integer $$$n$$$, resonance occurs
whenever
$$\omega_{xy}=\gamma B_0+n\omega_z\hspace{2em}[15]$$
and for each integer $$$n$$$, the corresponding effective
angular nutation frequency is
$$\omega_{nut}=\gamma B_{1,xy}J_n\left(\frac{\gamma B_{1,z}}{\omega_z}\right)\hspace{2em}[16]$$
Methods
Multiphoton resonances were verified in simulation. With an
extra solenoid coil in the $$$B_0$$$ direction, images of a lemon and pork rib
were acquired with all standard excitations replaced by effectively equivalent
two-photon versions (setup in Fig. 2). Without extra hardware, a vendor
provided adiabatic inversion pulse was played out with superimposed oscillating
and constant gradients to produce a multiband version. All experiments were on
an Aspect 1T wrist scanner (Aspect Imaging, Shoham, Israel).Results
Fig. 3 shows simulated lines depicting multiphoton
resonances, Fig. 4 shows single- and two-photon images of a lemon and pork rib,
and Fig. 5 shows a multiband inversion.Discussion and Conclusions
In the phase-modulated rotating frame, we see that
multiphoton excitation looks just like a scaled single-photon excitation with
some extra off-resonant fields. As shown here, with a constant amplitude $$$B_{1,z}$$$
field, frequency-offset slice selective $$$B_{1,xy}$$$ fields act as scaled
versions of their single photon counterparts, whether they are adiabatic or
not. Multiphoton excitation provides insight for designing excitation pulses,
with or without extra hardware, and can potentially be used for spin-locking
experiments to explore novel contrast for MRI with z-axis frequencies chosen to
probe biological or chemical processes of interest. As gradient10,11 and
traditional RF coils gain more channels and spatial-temporal flexibility, much
more can be achieved using multiphoton MRI to encode spatial, spectral and
temporal information.Acknowledgements
The
authors thank Michael Lustig, PhD, of UC Berkeley for discussions and
assistance with the Aspect scanner, Neelesh Ramachandran of UC Berkeley for
assistance with hardware construction, and Aspect Imaging for providing the
scanner.References
1. Lauterbur PC. Image
Formation by Induced Local Interactions: Examples Employing Nuclear Magnetic
Resonance. Nature. 1973;242(5394):190-191. doi:10.1038/242190a0
2. Michal
CA. Nuclear magnetic resonance noise spectroscopy using two-photon excitation. The
Journal of Chemical Physics. 2003;118(8):3451-3454. doi:10.1063/1.1553758
3. Eles
PT, Michal CA. Two-photon two-color nuclear magnetic resonance. The Journal
of Chemical Physics. 2004;121(20):10167-10173. doi:10.1063/1.1808697
4. Eles
PT, Michal CA. Two-photon excitation in nuclear magnetic and quadrupole
resonance. Progress in Nuclear Magnetic Resonance Spectroscopy.
2010;56(3):232-246. doi:10.1016/j.pnmrs.2009.12.002
5. Zur
Y, Levitt MH, Vega S. Multiphoton NMR spectroscopy on a spin system with I
=1/2. The Journal of Chemical Physics. 1983;78(9):5293-5310.
doi:10.1063/1.445483
6. Gromov,
Schweiger. Multiphoton resonances in pulse EPR. Journal of magnetic
resonance. 2000;146(1):110-121. doi:10.1006/jmre.2000.2143
7. Saiko
AP, Fedaruk R, Markevich SA. Multi-photon transitions and Rabi resonance in
continuous wave EPR. Journal of Magnetic Resonance. 2015;259:47-55.
doi:10.1016/j.jmr.2015.07.013
8. Brunner
D, Pavan M, Dietrich B, Rothmund D, Heller A, Pruessmann K. Sideband Excitation
for Concurrent RF Transmission and Reception. In: ISMRM Annual Scientific
Meeting & Exhibition. Montreal, Quebec, Canada; 2011.
9. Scheffler
K, Loktyushin A, Bause J, Aghaeifar A, Steffen T, Schölkopf B. Spread-spectrum
magnetic resonance imaging. Magnetic Resonance in Medicine.
2019;82(3):877-885. doi:10.1002/mrm.27766
10. Littin
S, Jia F, Layton KJ, et al. Development and implementation of an 84-channel
matrix gradient coil. Magnetic Resonance in Medicine. 2018;79(2):1181-1191.
doi:10.1002/mrm.26700
11. Ertan
K, Taraghinia S, Sadeghi A, Atalar E. A z-gradient array for simultaneous
multi-slice excitation with a single-band RF pulse. Magnetic Resonance in
Medicine. 2018;80(1):400-412. doi:10.1002/mrm.27031