Synopsis
Key hardware features and requirements of integrated MR-guided Linac-based radiotherapy systems and practical implementations are explained.
Introduction and scope
The
integration of an MRI scanner and a Linac based external beam radiotherapy
system, with the objective to image the irradiated part of the patient while
the radiation is being applied is a challenging task. This abstract describes
the major engineering issues of such an integrated MR-Linac system and presents
three different solutions that have proven to work. The main focus will be MRI
hardware. The clinical use of such systems is beyond the scope of this
presentation.General requirements
In all practical integrated MR-Linac systems the patient is lying on a horizontal table and the radiation source can be rotated around an axis in the head-feet direction of the patient. The radiation is collimated by means of a variable aperture diaphragm (the so-called Multi Leaf Collimator, MLC), to a field with a lateral extension of 200-300 mm in both directions. Dose is determined by the setting of the MLC and the dose rate of the radiation source. For MRI guidance of the radiation treatment, one needs the static, gradient and RF fields needed to perform MR imaging in the same volume as where the radiation is being applied. The MRI hardware needed to generate these fields has to allow the radiation to pass through with as little attenuation as possible and such that the dose at the tumor is predictable to within a few percent.Making the MRI system transparent to radiation
In the MR-Linac systems developed by
Elekta/Philips and by ViewRay the MRI magnet is cylindrical and the plane of
rotation of the linac is the z=0 plane of the MRI system. Thus the radiation
beam is perpendicular to the static field of the MRI system. In both systems
the field generating coils have a sufficiently large gap between them that the
beam can pass through. The cryostat of the ViewRay system has an air-gap in the
beam region whereas in the Elekta/Philips system the beam has to pass through
thin cryostat walls. In either case, there will be some radiation exposure of the cold mass. The power is so small that it will never quench the magnet but the exposure can lead to radiation damage of construction material and components. In the system developed at the Cross Cancer Institute
(MagnetTX) the radiation beam reaches the patient through a central hole in the
pole of a yoke-based vertical field MRI magnet. Thus the beam is parallel to the field. In order to irradiate from multiple angles, the entire
scanner and the Linac are made to rotate around the patient.
Avoiding magnetic field exposure of the Linac
The magnetic field at the Linac has to be kept
small and stable. If it is too high (typically above 0.5 millitesla), no
electrons can escape from the cathode and the tube stops working. If the field
is not stable, the position of the focal spot of the radiation will vary and
this affects dose predictability. In the Elekta/Philips system, which has a
central field of 1.5 tesla, an active shielding configuration with four shield
coils keeps the field at the Linac sufficiently small. The 0.35 tesla ViewRay
system, does not have such active shielding. Instead the Linac and at other
field-sensitive beam generating components are shielded with local
high-permeability shielding. The CCI system also relies on passive shielding of
the Linac. The Linac can also be disturbed by magnetic fields induced in
magnetic material in the walls of the radiation bunker when this is magnetized
by the MRI magnet. Avoiding such wall effects is another motivation for
optimizing the magnet's external field. Where it is inevitable that bunker wall
material can get magnetized (especially in the floor below the system), it
may be necessary to cut the magnetic material out and replace it by something
non-magnetic.Avoiding field distortion and modulation in the MR scanner
Reducing the external field of the magnet also avoids that magnetic material belonging to the beam generation
system gets magnetized and adds to the field in the MRI imaging volume. In
order to be able to make good quality MR images while the Linac is rotating the
field contribution from the Linac gantry has to be made rotationally symmetric
to within 100 nT. This may require local compensation of magnetic parts such as
MLC servo motors. Apart from the static field generated by magnetic material on
the Linac gantry, the beam generation system can also produce AC fields, for
example from large currents flowing in power cables. Careful design of cabling
and slip-rings has to keep the amplitude of such mains-frequency fields below a
few nanotesla in the imaging volume.Avoiding RF noise from the Linac
Linacs generate enough broad-band RF noise to spoil an MRI image. RF shielding has to prevent this. In the Elekta/Philips system, this is accomplished by making the MR magnet an integral part of the RF room shield and to place all of the beam generation components on the dirty side of this shield. The other MR-Linac implementations use local RF shielding around the noisy components. Some RF-noise will be generated by the free electrons that result from the interaction between the ionizing radiation and the patient and other materials. This leads to some reduction in image quality; at 1.5 tesla/64 MHz it is hardly visible.Gradient coil with radiation window
The gradient coil needs a sufficiently large
opening, free of conductors, that the radiation can pass through. A thin
radiation-transparent epoxy cylinder to keep the two coil halves together is
allowed. Due to the gap the efficiency of an MR-Linac gradient coil is
significantly lower than of a standard coil of the same diameter. The hole in
the stray-field shielding may also lead to some eddy-currents in conducting
parts outside the coil.RF transmit and receive coils
If the coils are made from very thin copper foil on thin support structures the radiation absorption is small. Lumped components such as capacitors, amplifiers and cables have to be kept out of the radiation field. With proper design, the performance of radiation-transparent RF coils is similar to that of standard RF coils. Most RF components are sufficiently radiation resistant to survive exposure to the treatment radiation.Acknowledgements
No acknowledgement found.References
BW
Raaymakers, J W Lagendijk, JA Overweg, J Kok, J E Raaijmakers, E Kerkhof, R van
der Put, I Meijsing, S Crijns, F Benedosso, M van Vulpen, H W de Graaff,
J Allen, K Brown (2009). Integrating a 1.5 T MRI scanner with a 6 MV
accelerator: Proof of concept. Physics in medicine and biology. 54. N229-37.
Mutic, Sasa
& F Dempsey, James. (2014). The ViewRay System: Magnetic Resonance-Guided
and Controlled Radiotherapy. Seminars in radiation oncology. 24. 196-199.
B.G.
Fallone, B. Murray, S. Rathee, T. Stanescu, S. Steciw, S. Vidakovic, E. Blosser
and D. Tymofichuk,"First MR images obtained during megavoltage photon
irradiation from a prototype linac-MR system."Med. Phys. 36,
2084-2088(2009).